2. AtomicAbsorption
2
Thanks
To everyone who has helped us with support,
new books, hard/soft ware And over the internet
Special thanks for Thermo
http://www.thermofisher.com
3. AtomicAbsorption
3
Contents
Introduction
Theory
AAS Setup
Validity
Accessories
Techniques and facilities
Software
Application
o Air
o Water & Soil
o Foods
o Clinical
o Petrochemicals
o Pharmaceutical
4.
5. AtomicAbsorption
5
Metals
Metals account for about two thirds of
all the elements and about 24% of the
mass of the planet.
Metals have useful properties including
strength, ductility, high melting points,
thermal and electrical conductivity, and
toughness.
From the periodic table, it can be seen
that a large number of the elements
are classified as being a metal.
7. AtomicAbsorption
7
Abundance of Metals in the Human Body
Calcium (1.4%)
Magnesium (0.50%)
Potassium (0.34%)
Sodium (0.14%)
Iron (0.004%)
Zinc (0.003%)
8. AtomicAbsorption
8
Toxic and nutrition elements
Toxic metals
Pb, Cd, As, Hg, Al, Cr, Cu
Nutrition elements
Ca, Mg, Na, K, P
Micro-nutrients
Zn , Se, Mo, etc…
9. AtomicAbsorption
9
Heavy metals
Heavy metals are natural components of
the Earth's crust. They cannot be degraded
or destroyed.
To a small extent they enter our bodies via
food, drinking water and air.
As trace elements, some heavy metals (e.g.
copper, selenium, zinc) are essential to
maintain the metabolism of the human
body.
However, at higher concentrations they
can be toxic.
10.
11. AtomicAbsorption
11
Absorption
The process whereby the intensity of a
beam of electromagnetic radiation is
attenuated in passing through a material
medium by conversion of the energy of the
radiation to an equivalent amount of
energy appearing within the medium;
The radiant energy is converted into heat
or some other form of molecular energy.
12. AtomicAbsorption
12
The beer – Lambert Law
The absorption that takes place in an atomic
absorption system follows beer law.
A beam of light with intensity I0 is aimed at the
tested solution placed in a cuvette.
The intensities of the entering beam I0 and the
emerging beam I1 are measured, and the absorbance
A - is calculated from the ratio of the two
A = - log (I1/I0)
Different molecules absorb radiation of different
wavelengths. An absorption spectrum will show a
number of absorption bands corresponding to
structural groups within the molecule
16. AtomicAbsorption
16
Calibration
Optical absorption spectrometry is a
comparative technique in which the
signals by solutions of known
concentrations used to generate a
calibration curve is compared to the
signals of unknown samples to generate
results.
17.
18. AtomicAbsorption
18
AA Steps
The sample is weighed and then dissolved.
The resulting solution is sprayed into the
flame and atomized.
Light of a suitable wavelength for a particular
element is shone through the flame.
Some of this light is absorbed by the atoms of
the sample.
The amount of light absorbed is proportional
to the concentration of the element .
20. AtomicAbsorption
20
1. Lamps
Hollow Cathode Lamps (HCLs) are high
intensity, stable light sources that emit the
element specific spectral lines required for
Atomic Absorption spectrometry.
Provide a constant intense beam of
analytical light.
There are Coded or uncoded lamps.
21. AtomicAbsorption
21
Deuterium Lamp
The deuterium lamp emits radiation extending from 112
nm to 900 nm, although its continuous spectrum is only
from 180 nm to 300 nm.
The Deuterium lamp emits a blue-white light.
However, these lamps are used to produce Ultra-Violet
(UV) emissions which we can't see.
The outer lamp envelope is made form quartz rather than
glass. because glass does not transmit short wave UV light.
What makes Deuterium lamps so special, as a UV source,
is its continuous spectrum in the range from 180nm - 300
nm.
22. AtomicAbsorption
22
Background
Background interference is caused by either, non-specific absorption arising from light
scattering caused by solid particles or liquid droplets in the atomizing cell or, by light
absorption caused by molecules or radicals originating in the sample matrix.
It is usually measured by separate experiment and subtracted from the absorption of the
sample solution.
23. AtomicAbsorption
23
Background Correction
The cathode lamp and the deuterium lamp are
sequentially pulsed with a chopper or electronically
with delay of about 2ms.
When hollow cathode lamp is on and deuterium
lamp off total absorbance (AA + BG) is measured.
When the HCL is off and the deuterium lamp on the
continuum energy recorded is (BG).
The atomic signal is automatically calculated by
subtracting background from total absorbance.
24. AtomicAbsorption
24
Zeeman Background
Zeeman Background Correction is used
mainly in graphite furnace atomic
absorption systems. When an atom is
placed in a magnetic field and its
absorption of observed in polarized
light, the normal single line is split into
three components – б-, π and б
+displaced symmetrically about the
normal position
25. AtomicAbsorption
25
Zeeman Background Correction
Free atoms show Zeeman splitting in a magnetic field but molecules, liquid droplets
or solid particles show no Zeeman splitting and so advantage can be taken of
polarized light.
The π component is linearly polarized parallel to the magnetic field while the б
components are circularly polarized perpendicular to the magnetic field.
A polarizer is positioned in the optical system to remove the π components of the
transmitted radiation.
This affords background measurement at the exact analyte wavelength when
magnetic field is applied. Since the background is measured at the analyte
wavelength and not averaged as in D2 system structural molecular background and
spectral interferences are easily corrected.
26. AtomicAbsorption
26
2. Atomizers
Metal in the sample must undergo
desolation and vaporization in a
high-temperature source such as a
flame or graphite furnace to be free
atoms.
Destroy any analyte ions and
breakdown complexes
Create atoms (the elemental form)
of the element of interest
28. AtomicAbsorption
28
3. Monochromator
A monochromator is an optical device that
transmits a mechanically selectable narrow band
of wavelengths of light or other radiation chosen
from a wider range of wavelengths available at the
input.
Echelle monochromator provides automatic
wavelength and band pass set-up.
The high energy Quad Line background correction
system corrects for up to 2A of background with
less than 2 % error, and is fitted as standard to all
instruments.
29. AtomicAbsorption
29
The Grating
Grating An optical device within the
spectrometer used to separate the emitted light
into its component wavelengths.
The grating has a dual feature: it diffracts the
light and focuses it on the slits.
The grating is the main optic part of the
spectrometer;
It separates the light into all the wavelength
that composes it.
It Has 1800 grooves/mm.
30. AtomicAbsorption
30
4. Photo multiplier tube PMT
The PMT change the incidence
photons into electrical signal
As the detector the PMT determines
the intensity of photons of the
analytical line exiting the
monochromator
32. AtomicAbsorption
32
Auto-aligning Optics
Perfect setup every time
Auto-alignment and memory
Lamp carousel for 6 lamps
Dedicated power supply for each
position
Data coding of both element and
lamp current
High optical modulation frequency
33. AtomicAbsorption
33
6. Automatic Gas Control
Full safety monitoring facilities and safe
shutdown.
Completely enclosed “kitchen” area.
Automatic binary flow gas control system for
superb reliability and reproducibility.
36. AtomicAbsorption
36
A- Flame Atomic Absorption
Atomization through flame
Air- Acetylene (9 psi) .
Air- Acetylene -Nitrous oxide
Nitrous oxide (N2O) needs to be used with air (78%
N2 + 21% O2)
The level of measuring is about mg/l (ppm)
Support gases
Air 2.07 bar (30psi)
Nitrous oxide 2.75 bar (40psi)
37. AtomicAbsorption
37
Flame System
Universal Finned Titanium 50mm burner suitable for air/acetylene and nitrous
oxide/acetylene flame types.
An inert fluoroplastic spray chamber incorporating an externally adjustable inert
impact bead and flow spoiler.
An inert over-pressure membrane should be housed in the rear of the spray
chamber for maximum operator safety.
Automatic gas system using binary flow control and programmable array state
logic for reliability
Full safety interlocks, including pressure sensors on both lines, power failure
protection, burner interlock and flame sensor
Fuel and oxidant flow rates software controllable
Automatic flame ignition and optimization
38. AtomicAbsorption
38
Flame Safety
If carbon deposits appear, the flame
must be extinguished immediately
and the deposits removed.
Aspiration of solutions of perchloric
acid and metal per chlorates into a
nitrous oxide supported flame can
increase the risk of explosion or
flashback
Certain elements, notably Ag, Au
and Cu, can form unstable
acetylides, increasing the risk of
explosion or flashback.
The use of organic solvents in flame
AAS is an inherently hazardous
procedure.
The door must be closed when
lighting a flame, and during normal
operation.
All flames produce large quantities
of heat and toxic combustion
products. These must be removed
by a suitable fume extraction
system.
39. AtomicAbsorption
39
Burners
Two types of burner are available for
spectrometer:
5 cm slot Universal Titanium Burner
suitable for general purpose use with all
flame types
10 cm slot Titanium Burner
suitable for air/acetylene flames only.
Sensitivity for elements measured with this
flame will be improved compared to the
Universal Burner.
40. AtomicAbsorption
40
B- Furnace atomic Absorption
Electrothermal atomization (ETA) is a
technique for improving the sensitivity and
limit-of-detection (1000 times) for atomic
absorption measurements.
A small amount of sample or standard
solution is placed inside a hollow graphite
tube.
This is resistively-heated in a temperature
program to remove liquid, burn off organics,
atomize the residuals to form a plume of free
metal vapour, detect the metals and finally
clean the tube.
41. AtomicAbsorption
41
Furnace main parts
The furnace main Technique parts
1. Electrical Thermal heating "power supply”.
2. Furnace head.
3. Auto sampler.
4. Shield and cleaning gas.
5. Cooling system.
42. AtomicAbsorption
42
Furnace System
Choice of Deuterium or Zeeman background correction furnaces
Mount directly in dedicated compartment
Binary flow controlled internal gas system
Choice of alternate or inert gases
Furnace cycle to allow up to 20 phases to be programmed
Cuvette firings counter
Furnace auto-sampler to be included with furnace head and power supply
Slow injection and uptake options
Wash and waste vessels to be part of the auto-sampler system and not occupy
extra space on the floor or bench of the laboratory
43. AtomicAbsorption
43
Furnace Program
The main four Furnace programs
1. Drying phase, where the sample is
warmed to remove the solvent.
2. Ashing phase, where as much of the
sample matrix as possible is removed.
3. Atomization or measurement phase.
4. Cleaning phase, where the cuvette is
heated to a high temperature to
remove any previous sample.
44. AtomicAbsorption
44
1. Furnace power supply
200/220/240V at 50/60 Hz, 30A Single phase.
Power consumption 7.2kVA. GF95Z - additional 1.5kVA
Description
All cuvettes mount directly in an all-graphite containment
with end loaded contacts.
Cuvettes are self aligning, and can be rapidly exchanged
with a single lever movement.
The binary flow controlled internal gas system, with gas
stop, offers a choice of the inert gas or an alternate gas,
and the fixed external inert gas flow protects the cuvette
and purges the optical temperature sensor.
45. AtomicAbsorption
45
2. Furnace Head
Dynamic, optical cuvette temperature control, pre-heated
cuvette injection and coolant water temperature
compensation optimize analyses
All graphite containment reduces contamination risk
Very wide range of alternative cuvettes: Extended Lifetime
Cuvettes (ELC) provide uninterrupted overnight analysis and
lower cost of ownership
Ash/Atomize self-optimization with SOLAAR software
Maximum furnace sensitivity and the widest furnace dynamic
working range Correction up to 2A of background, with <2%
residual error for even the fastest transient signals
The optional Zeeman Background Correction is performed at
the exact analyte wavelength
46. AtomicAbsorption
46
The Zeeman Background
The Zeeman background is the
splitting of spectral lines into several
polarized components as a result of
the effect of an applied magnetic field.
On the application of the magnetic
field a central line appears at the same
wavelength as the original line (the π
line) having half the intensity of the
original line.
On either side of the π line appears
two other lines (the σ± lines) having
one quarter of the intensity of the
original line.
The π line is linearly polarized with the
electric vector parallel to the magnetic
field and the σ± lines are circularly
polarized at right angles to the
direction of the magnetic field.
the π line is absorbed by both sample
and back ground whereas the σ±
components are only absorbed by the
background.
47. AtomicAbsorption
47
Zeeman Background
The magnet fitted to the GF95Z
Zeeman Furnace Head produces a
variable magnetic field up to 0.85
Tesla at mains frequency during the
atomization and auto zero phases.
This can affect other electronic
systems in the vicinity.
48. AtomicAbsorption
48
Cuvettes
Normal Cuvettes (Electro graphite)
Volatile elements
Coated Electro graphite (Pyrolytically coated)
Carbide forming elements
Medium volatile
Refractory elements
Extended Life-time Cuvettes (ELC’s)
More stable
Omega Platform ELC’s
Volatile elements in ‘heavy’ matrices
49. AtomicAbsorption
49
3. Autosampler
Features of the Furnace Autosampler
1. Automatic matrix modification - wet and dry
mixing options .
2. Automatic standard preparation - fixed and
variable volume may be used
3. Automatic re-concentration of samples, using
multiple injections .
4. Automatic, intelligent dilution of samples
5. Automatic standards addition preparation
6. Automatic re-scale and re-calibration
50. AtomicAbsorption
50
4. Argon Gas
This protects the hot cuvette from atmospheric
oxygen, and flushes sample vapours from the
cuvette interior.
Argon is recommended; nitrogen can be used
with some loss of performance for some
elements.
Connect the inert gas supply to the inlet port
labeled ARGON 2 at the rear of the Furnace
Power Supply unit.
The inert gas supply must be regulated to
1.1±0.14 bar (15±2 psi).
51. AtomicAbsorption
51
5. Cooling System
A supply of reasonably clean (e.g. drinking) water, at a temperature of less than
30ºC and a pressure of 1.4 to 6.9 bar (20 - 100psi), capable of providing a
minimum flow rate of 0.7l/min is required.
Do not allow the pressure to exceed 6.9 bar (100 psi).
Connect the cooling water inlet and outlet hoses to the water inlet and outlet
connections on the Furnace Power Supply unit.
Recirculators
The Furnace can be cooled by a temperature controlled Recirculators/chiller unit
instead of mains water.
Set the recirculating water temperature to about 5ºC above ambient
temperature, providing that this is less than 30ºC.
52. AtomicAbsorption
52
Chiller
An air-cooled re-circulating water chiller
shall be provided to cool.
suitable for operation with an ambient
temperature range +15 C to + 35 C
Highly efficient cooling
Accurate temperature control
Environmental friendly (CFC - free)
Quiet operation
53. AtomicAbsorption
53
C- Vapor Atomic Absorption
Parts per billion sensitivities for a
number of environmentally
important elements are not
attainable by conventional flame
atomic absorption spectrometry
and alternative techniques have to
be used.
Hydride generation AAS is
applicable to mercury and the
arsenic group elements, and
provides cost effective analysis with
sub-ppb detection limits.
The elements that can be
determined with the VP100 are
those that can form gaseous
hydrides, or in the case of mercury,
a mono-atomic vapour.
These include As, Se, Bi, Te, Sb, Sn,
and Hg.
54. AtomicAbsorption
54
Vapor AA
Detection limits typically 1000x better than
those achieved by conventional flame analysis
Hydrochloric acid + sodium borohydride
unstable hydride of 8 elements give free atoms
Hg - As - Se – Sb
Ge - Bi - Sn – Te
For a volatile elements in the range of ug/l
(ppb)
55. AtomicAbsorption
55
Vapor Reaction
As3+ NaBH4 AsH3 (gas) + B2H6
AsH3 (gas) As + 3/2 H2
The arsenic ions are reduced by the strong
reducing agent sodium borohydride (NaBH4)
and the arsenic hydride is formed.
56. AtomicAbsorption
56
4 Channel peristaltic pump
Full automation
All functions controlled through
software
Simple Installation and Plumbing
4 Colour coded channels
Single RS232C connection to
spectrometer
Compatible with all SOLAAR
supported auto-samplers
Channel 1 – Reductant
Channel 2 – Acid Reagent
Channel 3 – Sample
Channel 4 – Drain
57. AtomicAbsorption
57
Peristaltic pump
Stepper motor driven
Precise and accurate
Software controlled
Pump speed is now a Method parameter
Flexible operation
Optimise reagent
consumption/sensitivity trade off
Repeatable results
60. AtomicAbsorption
60
Lab Requirements
Ensure avoidance of:
Direct sunlight.
Proximity to heat sources.
Draughts, particularly from such items
as air conditioning vents and fans.
Excessive vibration.
temperature is maintained between
+ 5 oC and + 40 oC with a maximum
temperature variation of less than 2oC
per hour.
Relative humidity should be
maintained between 20% and 80%.
These instruments are designed for
operation in clean air conditions.
The laboratory must be free of all
contaminants that could have a
degrading effect on the instrument
components.
Dust, acid and organic vapours must
be excluded from the work area..
61. AtomicAbsorption
61
Fume Extraction
All flames produce large quantities of heat and toxic
combustion products.
These must be removed by a suitable fume extraction
system. Specifications of a suitable extraction system
are provided in the Pre-Installation Manual.
The fume extraction hood must not be attached to the
chimney, and an air-gap of between 150 and 230mm
must be made.
62.
63. AtomicAbsorption
63
Slotted Tube Trap
This accessory enhances the flame
sensitivity for certain elements by 2-5
times.
It consists of a slotted tube held in the
flame .
NEVER ATTEMPT TO LIGHT OR
EXTINGUISH A FLAME WITH The stat in
the operational position.
64. AtomicAbsorption
64
ID 100 Auto-dilutor
This is an accessory that will provide automatic sample
dilution and standard preparation for flame AAS
ID 100 is not compatible with VP90 AND VP 100.
The ID100 Auto dilutor System can automatically prepare
working calibration standards from a single master standard
as they are needed, so that no manual dilution steps are
required.
It also simplifies the task of handling over-range samples by
intelligently diluting them into the calibration range, thus
extending the effective working range.
Full automation is possible by combining the ID100 with any
of the Thermo Elemental range of AA Flame autosampler.
65. AtomicAbsorption
65
Auto-sampler ASX520-ASXR8
ASX520 Auto sampler
A workhorse auto sampler for
unattended analysis
with a maximum load of 360
samples for busier laboratories
EXR8 Auto sampler
A large sample load of up to 720
samples with automatic,
unattended analysis for high
throughput laboratories
66. AtomicAbsorption
66
EC 90 Electric heating
It replaces the Flame Heated Measurement
Cell with an Electrically Heated Atomization
Cell, so that Vapor measurements can be made
without a flame.
The accessory consists of two parts:
1. The EC90 Furnace Head.
2. The EC90 Power Supply
67. AtomicAbsorption
67
Furnace Vision System GF TV
CCD camera provides high definition images
Uses spectrometer optical path for a clear, direct,
on-axis view of the cuvet
Simplified method development
Accurate adjustment of capillary possible
Perfect sample injection depth
Drying and ashing phases can be easily optimised
High quality and reproducible results
Images can be captured and stored
Correct depth
Capillary too low
Capillary too high
Correct depth
Capillary too low
Capillary too high
68. AtomicAbsorption
68
ISQ – Intelligent Spectrometer Qualification
ISQ is an automatic, software driven process to demonstrate that your spectrometer
hardware is performing consistently to specification over the intended operating ranges.
Intelligent
Identifies instrument in use
Selects appropriate tests
Software automatically controls tests
Clear result display
Spectrometer
Tests the hardware of the spectrometer
Helps to diagnose the source of any hardware problems at an early stage
Qualification
Verifies that the instrument is operating entirely within the designed specifications
Confirms instrument is capable of producing sound analytical data.
69.
70. AtomicAbsorption
70
Validation of the Machine
Validation kit helps
1. Monitor the regulatory compliance status of
spectrometer.
2. Determine the conformity of an AA
spectrometer to internally established standards.
3. Provide confidence for managerial and regulatory
personnel that the system is under control
provide all the documentation, hardware and
standard solutions necessary for the
validation process.
77. AtomicAbsorption
77
Essential and Toxic elements
Essential major elements
C N O P S Cl Na K Ca Mg
Essential trace elements
F I Se V Cr Mn Fe Co Ni Cu ZnMo Si Sn As
Toxic elements
Li Be Ba F Cl Br As Sb Bi Pb Sn Tl V Cr Mn Fe Co Ni Cu Zn Cd Hg
78. AtomicAbsorption
78
2. Clinical Applications
The majority of samples analysed are taken
from the main group of biological fluids, such
as whole blood, plasma, serum and urine.
Hard and soft tissues, such as bone, finger
nails and hair
Flame based analysis for the major and minor
essential elements, graphite furnace analysis
for the trace elements and vapour analysis for
the group of toxic
80. AtomicAbsorption
80
Toxic elements
Toxic elements are often defined as those
that interfere with metabolic processes.
The elements usually included in this group
are as follows:
Lead, Mercury, Arsenic, Thallium, Cadmium,
Aluminum, Boron, Antimony.
81. AtomicAbsorption
81
3. Pharmaceutical Applications
Samples of diclofenac sodium for
Na, K, Ca and Al analysis
Magnesium Stearate sample to be
analyzed for Cd, Ni and Pb.
Samples of vitamin tablets Se
analysis
82. AtomicAbsorption
82
Drug discovery and testing
Most Pharmaceutical Companies these days
develop drugs which are targeted at specific
cells in the body.
These drugs must be tested for correct
activity but more importantly for the absence
of any adverse side reactions.
83. AtomicAbsorption
83
Toxic and nutrition elements
Toxic metals
Pb, Cd, As, Hg, Al, Cr, Cu
Nutrition elements
Ca, Mg, Na, K, P
Micro-nutrients
Zn , Se, Mo, etc…
Process monitoring and control
Fe, other transition elements in frying oil
Ni in fat hydrogenation
Levels vary with process, typically 1000 – 10PPM
84. AtomicAbsorption
84
4. Agriculture Applications
soil analysis provides a measure of a
soil‘s potential to supply the
necessary nutrients to plants.
Plants may be sampled to monitor
nutrient uptake efficiency and also
to check for toxic metal
accumulation for health reasons.
85. AtomicAbsorption
85
Soil nutrient elements
P, K, Ca, Mg Macro nutrients at %
level
Cu, Fe, Mn, Zn Micro nutrients at
ppm level
Al, B, Na, Mo, Se Other nutrient
elements
As, Cd, Co, Cr, Ni, Pb Elements of
toxic interest
86. AtomicAbsorption
86
5. Petrochemical Applications
Measure refinery contaminant elements
Na, V, Fe, Ni by Flame-PPM
Measure fuel elements
Pb, Mn, usually low ppm-typically flame or furnace
Measure lube oil elements
Fresh - Ca, Ba, Mg, Zn, Mo, Na (flame)
Used - Ag, Al, Cr, Fe, Mn, Ni, Pb, Sn, Ti, Zn (often furnace)
87. AtomicAbsorption
87
lubricating oils
Every rotating mechanism in machinery of
all types depends on their use for smooth
operation.
Like engines and gearboxes used in modern
transportation, such as aircraft, ships, cars
and lorries and heavy construction
equipment.
Oils in use so that oil changes can be carried
out in time to prevent excessive wear
occurring in the components concerned.
88. AtomicAbsorption
88
Wear Metals
Wear metal Indicated condition
1 Silicon
Indicates dust intrusion, usually from improper air cleaner service.
Causes rapid engine wear and early failure.
2 Iron
Indicates wear originating from engine block, cylinder, gears, wrist
pins, rings (case iron), camshaft, oil pump, or crankshaft.
3 Copper
Usually indicates wear in bushings, injector shields, valve guides
,connecting rods, or piston pins.
4 Nickel
l Wear of plating on gears and certain
types of bearings.
5 Tin
Wear of certain types of bearings and coatings of connecting rods and
iron pistons
89. AtomicAbsorption
89
Wear Metals
Wear metal Indicated condition
6 Lead
In diesel engines indicates wear of bearings. In petrol engines fuel
blow by is indicated.
7 Chromium
Indicates ring wear or cooling system leakage if chromates are
used as inhibitors.
8 Aluminum
Indicates wear of pistons and certain
types of bearings.
9 Molybdenum
Indicates wear in certain types of
bearing alloys and in oil coolers.