SCANNING TUNNELING MICROSCOPE
MCG 5138 -Advanced Topics in Mechanical Engineering
Prepared by :
Jishang Tailor - 7462006
Roshan Teli - 7476661
Course Instructor - Prof. M. Yandouzi
Department of Mechanical Engineering,
University of Ottawa
Working Principle
• The principle of scanning tunneling microscopy is in quantum mechanics
which is different from classical mechanics.
• While classical mechanics deals up to macroscopic level, quantum
mechanics deals with microscopic level.
• Quantum mechanics explains the wave and particle like behavior of tiny
particles like photons and electrons.
• The quantum mechanics phenomenon which explains tunneling effect is
the working principal of scanning tunneling microscopy.
• Electron tunneling
Tunneling Effect
What is tunneling effect?
It is a phenomenon where a particle tunnels through a barrier that
it classically could not surmount.
Tunneling Effect
Tunneling Current
 In a metal, the energy levels of the
electrons are filled up to a particular
energy, known as the ‘Fermi energy’ 𝐸 𝐹. In
order for an electron to leave the metal, it
needs an additional amount of energy Φ,
the so-called ‘work function’.
 The electrons need to overcome a barrier
Φ to travel from tip to specimen or vice
versa.
 When a sharp metallic tip is brought very
close to the surface of a conductor, an
electric current can be detected due to the
tunneling of electrons through the air gap.
In this case, the air gap is considered as the
barrier and is only few 𝐴° thick.
 When an electrical voltage V is applied
between sample and tip, this tunneling
phenomenon results in a net electrical
current, the ‘tunneling current’.
Components of STM
The main components of STM include scanning tip, Piezoelectric scanner,
Distance control and scanning unit, Vibration isolation system and Data
processing unit (Computer).
1. Scanning Tip :
STM tips are usually made from tungsten metal or a platinum-iridium alloy
where at the very end of the tip (called apex) there is one atom of the
material. Scanning tip is the most important aspect of the STM as tunneling
current is carried by that particular atom.
STM tip under 103
x and 105
x magnificationSample and Tip at atomic level
2. Piezoelectric Scanner :
 The scanner tip is attached to a piezoelectric tube scanner. Piezoelectric
effect is a phenomenon under which the material changes its length
accordingly when put under an electrical voltage.
 By adjusting the voltage on the piezoelectric element, the distance
between the tip and the surface can be regulated.
 Piezoelectric crystals expand and contract very slightly depending on the
voltage applied to them and this principle is used to control the horizontal
position x, y, and the height z of the scanning tip.
Components of STM
3. Distance Control and Scanning unit :
 Position control using piezoelectric means is extremely fine, so a coarse
control is needed to position the tip close enough to the sample before the
piezoelectric control can take over.
4. Data Processing Unit (Computer) :
 The computer records the tunneling current and controls the voltage to the
piezoelectric tubes to produce a 3-dimensional map of the sample surface.
Components of STM
5. Vibration Isolation System :
 STM deals with extremely fine position measurements so the isolation of
any vibrations is very important.
 The tip and surface distance must be maintained in 𝐴°(0.1 𝑛𝑚)to get
desired atomic resolution.
 Due to extremely high sensitivity of tunneling current between tip and
sample surface height, it is absolutely necessary to reduce inner vibrations
and to isolate the system from external vibration.
 Damping can be achieved by : Pneumatic systems
Spring system
Eddy current system
Components of STM
Working of STM
• First, a voltage bias is applied and the tip is brought close to the sample by
coarse sample-to-tip control, which is turned off when the tip and sample
are sufficiently close. At close range, fine control of the tip in all three
dimensions near the sample is typically piezoelectric, maintaining tip-
sample separation W typically in the 4-7 𝐴° (0.4-0.7 nm) range.
• In this situation, the voltage bias will cause electrons to tunnel between the
tip and sample, creating a current that can be measured. Once tunneling is
established, the tip's bias and position with respect to the sample can be
varied and data are obtained from the resulting changes in current.
• If the tip is moved across the sample in the x-y plane, the changes in
surface height and density of states cause changes in current. These
changes are mapped in images. This change in current with respect to
position can be measured itself, or the height, z, of the tip corresponding to
a constant current can be measured.
• There two modes : Constant height mode
Constant current mode
Working of STM
Constant Height Mode :
 The voltage and height are both held constant while the current
changes to keep the voltage from changing; this leads to an image made
of current changes over the surface, which can be related to charge
density.
Working of STM
 The benefit to using a constant height
mode is that it is faster, as the piezoelectric
movements require more time to register
the height change in constant current
mode.
 However, it is applicable only when sample
surface is real flat, the corrugation more
than 6 to 7 𝐴°
will lead tip to crash.
 Generally less preferred due to the risk of
damaging the tip.
Constant Current Mode :
 Feedback electronics adjust the height by a voltage to the piezoelectric
height control mechanism.
 This leads to a height variation and thus the image comes from the tip
topography across the sample and gives a constant charge density
surface; this means contrast on the image is due to variations in charge
density
Working of STM
 It is a time consuming method compared to
the constant height mode as the feed back
control has to adjust the current constant
according height as the tip moves along the
specimen surface.
What STM measures?
 STM images are not direct surface images of the sample as in the case of
optical microscopy rather it is measure of the local density of states of a
material at it surface as a function of lateral (x-y) position on the sample
surface and energy.
 Within sample each electron has specific energy level and only certain
number of electrons can occupy that level at a time .
 The distribution that gives number of electrons allowed per energy level
as a function of certain energy level is called the density of states.
 So, the grey scale image generated is direct measurement atomic
corrugation of the surface.
Use of STM
Generation of images
 The amount of adjustment done
by the feedback loop is recorded
and defines the grid value which
can be displayed as the grey scale
image.
 Once the grid values are assigned
one can use it to deform it
perpendicular to the surface and
get 3D image.
 Now, different colors patterns are
used to color the image and
visualize it.
Images
Advantages:
 STMs are versatile. They can be used in ultra high vacuum, air, water and
other liquids and gasses.
 STMs give three dimensional profile of a surface, which allows researchers
to examine a multitude of characteristics, including roughness, surface
defects and molecule size.
 Lateral Resolution of 0.1 nm and 0.01 nm of resolution in depth can be
achieved.
Disadvantages:
 It is very expensive.
 It need specific training to operate effectively.
 STM need very clean surface, excellent vibration control while operation,
single atom tip.
Merits and Demerits
 If the tip has is not prepared as per standard which contains multipoint
and if the tip get contaminated during the operation, it creates multi
signal which are resulted as the artifacts in the image.
 Sometimes the frequency gained by amplifier is too high which makes the
tube to oscillate which gives noisy image.
 Sometimes rapid change in voltage are noted due to scanner encounters a
large step during scanning. The tube will continue to move even if the
voltage remains fixed which results in hazy image.
Artifacts in Image
Paper Review :
 The selected paper is research on Fabrication of nanoscale alumina on
NiAl(1 0 0) surface using scanning tunneling microscope under ultrathin
vacuum condition.
 In the research, two modes of oxidation were observed under different
conditions in formation of the alumina on the surface of NiAl(1 0 0).
 The first mode of oxidation observed under the tunneling current greater
than 0.4 nA and the power greater than 0.24 nW. Here, Al and O atoms
formed alumina on the surface in the proximity of the tip. The width and
thickness of the alumina strips growth are controlled by the current and
voltage bias.
Research Paper
 In the second mode oxidation smaller
power and a smaller bias (1.0V) were
used which resulted in growth of
crystalline alumina along direction [0 0
1] or [0 1 0] of NiAl(1 0 0) surface in
the tip scanned area of oxidized
surfaces, irrespective of the scanning
direction of tip.
Direction of crystalline alumina
Direction of scanning
Research Paper
 The combination of above two oxidation modes were used to fabricate
crystalline alumina on NaAl(1 0 0) surface in which bias was kept constant
at 0.7 V and current was raised from 0.3 nA to 1.6 nA which created
abrupt alumina on the surface.
 With same condition of oxidation on the surface keeping bias at 0.7 V and
lowering current to 0.3 nA just altered the topography of the surface and
it became more structured, indicating the gradual formation of the oxide
strips along direction [0 1 0] and [0 0 1] of NiAl (1 0 0)
 The STM tip induced oxidation has been widely studied and this is very
important break through in research as industrialization of such
technology at economical rate can use to fabricate nanosensors, bio
materials, nanomedicine.
 It helps to change the property of the structure at nano-scale which helps
to create strong structure.
Research Paper
Thank You . . .

Contenu connexe

Tendances

transmission electron microscopy
transmission electron microscopytransmission electron microscopy
transmission electron microscopyJessa Ariño
 
AFM and STM (Scanning probe microscopy)
AFM and STM (Scanning probe microscopy)AFM and STM (Scanning probe microscopy)
AFM and STM (Scanning probe microscopy)Preeti Choudhary
 
Scanning tunneling microscope (STM)
 Scanning tunneling microscope (STM) Scanning tunneling microscope (STM)
Scanning tunneling microscope (STM)Balsam Ata
 
Scanning tunneling microscopy
Scanning tunneling microscopyScanning tunneling microscopy
Scanning tunneling microscopyAzhar Ansari
 
Scanning electron microscope
Scanning electron microscopeScanning electron microscope
Scanning electron microscopeAkhtar Kamal
 
transmission Electron Microscopy (Tem)
transmission Electron Microscopy (Tem)transmission Electron Microscopy (Tem)
transmission Electron Microscopy (Tem)Sivasangari Shanmugam
 
Atomic force microscopy
Atomic force microscopyAtomic force microscopy
Atomic force microscopyRagavi Charm
 
X-ray Photoelecctron Spectroscopy (XPS)
X-ray Photoelecctron Spectroscopy (XPS)X-ray Photoelecctron Spectroscopy (XPS)
X-ray Photoelecctron Spectroscopy (XPS)faheem maqsood
 
Transmission electron microscope
Transmission electron microscopeTransmission electron microscope
Transmission electron microscopeRaj Mohan
 
Atomic force microscope
Atomic force microscopeAtomic force microscope
Atomic force microscopeSenthil Arasan
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopyJessa Ariño
 
Transmission electron microscope
Transmission electron microscopeTransmission electron microscope
Transmission electron microscopeSenthil Arasan
 
scanning electron microscope (SEM)
scanning electron microscope (SEM)scanning electron microscope (SEM)
scanning electron microscope (SEM)Himanshu Dixit
 

Tendances (20)

transmission electron microscopy
transmission electron microscopytransmission electron microscopy
transmission electron microscopy
 
AFM and STM (Scanning probe microscopy)
AFM and STM (Scanning probe microscopy)AFM and STM (Scanning probe microscopy)
AFM and STM (Scanning probe microscopy)
 
Raman spectroscopy
Raman spectroscopyRaman spectroscopy
Raman spectroscopy
 
Scanning tunneling microscope (STM)
 Scanning tunneling microscope (STM) Scanning tunneling microscope (STM)
Scanning tunneling microscope (STM)
 
Stem
StemStem
Stem
 
Scanning probe microscope
Scanning probe microscopeScanning probe microscope
Scanning probe microscope
 
Scanning tunneling microscopy
Scanning tunneling microscopyScanning tunneling microscopy
Scanning tunneling microscopy
 
Scanning electron microscope
Scanning electron microscopeScanning electron microscope
Scanning electron microscope
 
transmission Electron Microscopy (Tem)
transmission Electron Microscopy (Tem)transmission Electron Microscopy (Tem)
transmission Electron Microscopy (Tem)
 
Atomic force microscopy
Atomic force microscopyAtomic force microscopy
Atomic force microscopy
 
Sem n tem
Sem n temSem n tem
Sem n tem
 
X-ray Photoelecctron Spectroscopy (XPS)
X-ray Photoelecctron Spectroscopy (XPS)X-ray Photoelecctron Spectroscopy (XPS)
X-ray Photoelecctron Spectroscopy (XPS)
 
Atomic Force Microscopy
Atomic Force MicroscopyAtomic Force Microscopy
Atomic Force Microscopy
 
Transmission electron microscope
Transmission electron microscopeTransmission electron microscope
Transmission electron microscope
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopy
 
Atomic force microscope
Atomic force microscopeAtomic force microscope
Atomic force microscope
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopy
 
Transmission electron microscope
Transmission electron microscopeTransmission electron microscope
Transmission electron microscope
 
SEM,TEM & AFM
SEM,TEM & AFMSEM,TEM & AFM
SEM,TEM & AFM
 
scanning electron microscope (SEM)
scanning electron microscope (SEM)scanning electron microscope (SEM)
scanning electron microscope (SEM)
 

Similaire à STM ppt

Scanning tunneling microscope.pptx
Scanning tunneling microscope.pptxScanning tunneling microscope.pptx
Scanning tunneling microscope.pptxAliya Fathima Ilyas
 
ETE444-lec3.pdf
ETE444-lec3.pdfETE444-lec3.pdf
ETE444-lec3.pdfmashiur
 
ETE444-lec3.pdf
ETE444-lec3.pdfETE444-lec3.pdf
ETE444-lec3.pdfmashiur
 
Scanning Tunneling Microscopy and UHV Scanning Tunneling Microscopy
Scanning Tunneling Microscopy and UHV Scanning Tunneling MicroscopyScanning Tunneling Microscopy and UHV Scanning Tunneling Microscopy
Scanning Tunneling Microscopy and UHV Scanning Tunneling MicroscopyRamkumar Niluroutu
 
ETE444-lec3.pptx
ETE444-lec3.pptxETE444-lec3.pptx
ETE444-lec3.pptxmashiur
 
electron scattering,SEM,TEM,tunnel effect and lenses
electron scattering,SEM,TEM,tunnel effect and lenseselectron scattering,SEM,TEM,tunnel effect and lenses
electron scattering,SEM,TEM,tunnel effect and lensesKASHISHMANGAL2
 
Scanning Electron Microscopy
Scanning Electron MicroscopyScanning Electron Microscopy
Scanning Electron MicroscopySejal9787
 
Scanning Tunnelling Microscope (STM).pdf
Scanning Tunnelling Microscope (STM).pdfScanning Tunnelling Microscope (STM).pdf
Scanning Tunnelling Microscope (STM).pdfAadeshRajguru
 
Lecture_10_Atomic force microscopy(1).pdf
Lecture_10_Atomic  force microscopy(1).pdfLecture_10_Atomic  force microscopy(1).pdf
Lecture_10_Atomic force microscopy(1).pdfkeerthiraja8998
 
Scanning tunneling microscope by joshna.pptx
Scanning tunneling microscope by joshna.pptxScanning tunneling microscope by joshna.pptx
Scanning tunneling microscope by joshna.pptxamalajanet
 
Charecterization techniques
Charecterization techniquesCharecterization techniques
Charecterization techniquesMayur Gaikwad
 
Spin valve transistor
Spin valve transistorSpin valve transistor
Spin valve transistorEeshan Mishra
 
Afm electrical mode - parametric testing(i-v ,c-v)
Afm   electrical mode - parametric testing(i-v ,c-v)Afm   electrical mode - parametric testing(i-v ,c-v)
Afm electrical mode - parametric testing(i-v ,c-v)Abhishek Kumar
 
ELECTRON MICROSCOPY (TEM & SEM)
ELECTRON MICROSCOPY (TEM & SEM)ELECTRON MICROSCOPY (TEM & SEM)
ELECTRON MICROSCOPY (TEM & SEM)ADITYA ARYA
 

Similaire à STM ppt (20)

TEM.pptx
TEM.pptxTEM.pptx
TEM.pptx
 
TEM.pptx
TEM.pptxTEM.pptx
TEM.pptx
 
Scanning tunneling microscope.pptx
Scanning tunneling microscope.pptxScanning tunneling microscope.pptx
Scanning tunneling microscope.pptx
 
ETE444-lec3.pdf
ETE444-lec3.pdfETE444-lec3.pdf
ETE444-lec3.pdf
 
ETE444-lec3.pdf
ETE444-lec3.pdfETE444-lec3.pdf
ETE444-lec3.pdf
 
Stm 07.08.13
Stm 07.08.13Stm 07.08.13
Stm 07.08.13
 
Scanning Tunneling Microscopy and UHV Scanning Tunneling Microscopy
Scanning Tunneling Microscopy and UHV Scanning Tunneling MicroscopyScanning Tunneling Microscopy and UHV Scanning Tunneling Microscopy
Scanning Tunneling Microscopy and UHV Scanning Tunneling Microscopy
 
ETE444-lec3.pptx
ETE444-lec3.pptxETE444-lec3.pptx
ETE444-lec3.pptx
 
Stm
StmStm
Stm
 
electron scattering,SEM,TEM,tunnel effect and lenses
electron scattering,SEM,TEM,tunnel effect and lenseselectron scattering,SEM,TEM,tunnel effect and lenses
electron scattering,SEM,TEM,tunnel effect and lenses
 
Scanning Electron Microscopy
Scanning Electron MicroscopyScanning Electron Microscopy
Scanning Electron Microscopy
 
Scanning Tunnelling Microscope (STM).pdf
Scanning Tunnelling Microscope (STM).pdfScanning Tunnelling Microscope (STM).pdf
Scanning Tunnelling Microscope (STM).pdf
 
Lecture_10_Atomic force microscopy(1).pdf
Lecture_10_Atomic  force microscopy(1).pdfLecture_10_Atomic  force microscopy(1).pdf
Lecture_10_Atomic force microscopy(1).pdf
 
AFMppt.ppt
AFMppt.pptAFMppt.ppt
AFMppt.ppt
 
Scanning tunneling microscope by joshna.pptx
Scanning tunneling microscope by joshna.pptxScanning tunneling microscope by joshna.pptx
Scanning tunneling microscope by joshna.pptx
 
IAS Report Final
IAS Report FinalIAS Report Final
IAS Report Final
 
Charecterization techniques
Charecterization techniquesCharecterization techniques
Charecterization techniques
 
Spin valve transistor
Spin valve transistorSpin valve transistor
Spin valve transistor
 
Afm electrical mode - parametric testing(i-v ,c-v)
Afm   electrical mode - parametric testing(i-v ,c-v)Afm   electrical mode - parametric testing(i-v ,c-v)
Afm electrical mode - parametric testing(i-v ,c-v)
 
ELECTRON MICROSCOPY (TEM & SEM)
ELECTRON MICROSCOPY (TEM & SEM)ELECTRON MICROSCOPY (TEM & SEM)
ELECTRON MICROSCOPY (TEM & SEM)
 

STM ppt

  • 1. SCANNING TUNNELING MICROSCOPE MCG 5138 -Advanced Topics in Mechanical Engineering Prepared by : Jishang Tailor - 7462006 Roshan Teli - 7476661 Course Instructor - Prof. M. Yandouzi Department of Mechanical Engineering, University of Ottawa
  • 2. Working Principle • The principle of scanning tunneling microscopy is in quantum mechanics which is different from classical mechanics. • While classical mechanics deals up to macroscopic level, quantum mechanics deals with microscopic level. • Quantum mechanics explains the wave and particle like behavior of tiny particles like photons and electrons. • The quantum mechanics phenomenon which explains tunneling effect is the working principal of scanning tunneling microscopy.
  • 4. What is tunneling effect? It is a phenomenon where a particle tunnels through a barrier that it classically could not surmount. Tunneling Effect
  • 5. Tunneling Current  In a metal, the energy levels of the electrons are filled up to a particular energy, known as the ‘Fermi energy’ 𝐸 𝐹. In order for an electron to leave the metal, it needs an additional amount of energy Φ, the so-called ‘work function’.  The electrons need to overcome a barrier Φ to travel from tip to specimen or vice versa.  When a sharp metallic tip is brought very close to the surface of a conductor, an electric current can be detected due to the tunneling of electrons through the air gap. In this case, the air gap is considered as the barrier and is only few 𝐴° thick.  When an electrical voltage V is applied between sample and tip, this tunneling phenomenon results in a net electrical current, the ‘tunneling current’.
  • 6. Components of STM The main components of STM include scanning tip, Piezoelectric scanner, Distance control and scanning unit, Vibration isolation system and Data processing unit (Computer). 1. Scanning Tip : STM tips are usually made from tungsten metal or a platinum-iridium alloy where at the very end of the tip (called apex) there is one atom of the material. Scanning tip is the most important aspect of the STM as tunneling current is carried by that particular atom. STM tip under 103 x and 105 x magnificationSample and Tip at atomic level
  • 7. 2. Piezoelectric Scanner :  The scanner tip is attached to a piezoelectric tube scanner. Piezoelectric effect is a phenomenon under which the material changes its length accordingly when put under an electrical voltage.  By adjusting the voltage on the piezoelectric element, the distance between the tip and the surface can be regulated.  Piezoelectric crystals expand and contract very slightly depending on the voltage applied to them and this principle is used to control the horizontal position x, y, and the height z of the scanning tip. Components of STM
  • 8. 3. Distance Control and Scanning unit :  Position control using piezoelectric means is extremely fine, so a coarse control is needed to position the tip close enough to the sample before the piezoelectric control can take over. 4. Data Processing Unit (Computer) :  The computer records the tunneling current and controls the voltage to the piezoelectric tubes to produce a 3-dimensional map of the sample surface. Components of STM
  • 9. 5. Vibration Isolation System :  STM deals with extremely fine position measurements so the isolation of any vibrations is very important.  The tip and surface distance must be maintained in 𝐴°(0.1 𝑛𝑚)to get desired atomic resolution.  Due to extremely high sensitivity of tunneling current between tip and sample surface height, it is absolutely necessary to reduce inner vibrations and to isolate the system from external vibration.  Damping can be achieved by : Pneumatic systems Spring system Eddy current system Components of STM
  • 11. • First, a voltage bias is applied and the tip is brought close to the sample by coarse sample-to-tip control, which is turned off when the tip and sample are sufficiently close. At close range, fine control of the tip in all three dimensions near the sample is typically piezoelectric, maintaining tip- sample separation W typically in the 4-7 𝐴° (0.4-0.7 nm) range. • In this situation, the voltage bias will cause electrons to tunnel between the tip and sample, creating a current that can be measured. Once tunneling is established, the tip's bias and position with respect to the sample can be varied and data are obtained from the resulting changes in current. • If the tip is moved across the sample in the x-y plane, the changes in surface height and density of states cause changes in current. These changes are mapped in images. This change in current with respect to position can be measured itself, or the height, z, of the tip corresponding to a constant current can be measured. • There two modes : Constant height mode Constant current mode Working of STM
  • 12. Constant Height Mode :  The voltage and height are both held constant while the current changes to keep the voltage from changing; this leads to an image made of current changes over the surface, which can be related to charge density. Working of STM  The benefit to using a constant height mode is that it is faster, as the piezoelectric movements require more time to register the height change in constant current mode.  However, it is applicable only when sample surface is real flat, the corrugation more than 6 to 7 𝐴° will lead tip to crash.  Generally less preferred due to the risk of damaging the tip.
  • 13. Constant Current Mode :  Feedback electronics adjust the height by a voltage to the piezoelectric height control mechanism.  This leads to a height variation and thus the image comes from the tip topography across the sample and gives a constant charge density surface; this means contrast on the image is due to variations in charge density Working of STM  It is a time consuming method compared to the constant height mode as the feed back control has to adjust the current constant according height as the tip moves along the specimen surface.
  • 14. What STM measures?  STM images are not direct surface images of the sample as in the case of optical microscopy rather it is measure of the local density of states of a material at it surface as a function of lateral (x-y) position on the sample surface and energy.  Within sample each electron has specific energy level and only certain number of electrons can occupy that level at a time .  The distribution that gives number of electrons allowed per energy level as a function of certain energy level is called the density of states.  So, the grey scale image generated is direct measurement atomic corrugation of the surface. Use of STM
  • 15. Generation of images  The amount of adjustment done by the feedback loop is recorded and defines the grid value which can be displayed as the grey scale image.  Once the grid values are assigned one can use it to deform it perpendicular to the surface and get 3D image.  Now, different colors patterns are used to color the image and visualize it. Images
  • 16. Advantages:  STMs are versatile. They can be used in ultra high vacuum, air, water and other liquids and gasses.  STMs give three dimensional profile of a surface, which allows researchers to examine a multitude of characteristics, including roughness, surface defects and molecule size.  Lateral Resolution of 0.1 nm and 0.01 nm of resolution in depth can be achieved. Disadvantages:  It is very expensive.  It need specific training to operate effectively.  STM need very clean surface, excellent vibration control while operation, single atom tip. Merits and Demerits
  • 17.  If the tip has is not prepared as per standard which contains multipoint and if the tip get contaminated during the operation, it creates multi signal which are resulted as the artifacts in the image.  Sometimes the frequency gained by amplifier is too high which makes the tube to oscillate which gives noisy image.  Sometimes rapid change in voltage are noted due to scanner encounters a large step during scanning. The tube will continue to move even if the voltage remains fixed which results in hazy image. Artifacts in Image
  • 18. Paper Review :  The selected paper is research on Fabrication of nanoscale alumina on NiAl(1 0 0) surface using scanning tunneling microscope under ultrathin vacuum condition.  In the research, two modes of oxidation were observed under different conditions in formation of the alumina on the surface of NiAl(1 0 0).  The first mode of oxidation observed under the tunneling current greater than 0.4 nA and the power greater than 0.24 nW. Here, Al and O atoms formed alumina on the surface in the proximity of the tip. The width and thickness of the alumina strips growth are controlled by the current and voltage bias. Research Paper
  • 19.  In the second mode oxidation smaller power and a smaller bias (1.0V) were used which resulted in growth of crystalline alumina along direction [0 0 1] or [0 1 0] of NiAl(1 0 0) surface in the tip scanned area of oxidized surfaces, irrespective of the scanning direction of tip. Direction of crystalline alumina Direction of scanning Research Paper
  • 20.  The combination of above two oxidation modes were used to fabricate crystalline alumina on NaAl(1 0 0) surface in which bias was kept constant at 0.7 V and current was raised from 0.3 nA to 1.6 nA which created abrupt alumina on the surface.  With same condition of oxidation on the surface keeping bias at 0.7 V and lowering current to 0.3 nA just altered the topography of the surface and it became more structured, indicating the gradual formation of the oxide strips along direction [0 1 0] and [0 0 1] of NiAl (1 0 0)  The STM tip induced oxidation has been widely studied and this is very important break through in research as industrialization of such technology at economical rate can use to fabricate nanosensors, bio materials, nanomedicine.  It helps to change the property of the structure at nano-scale which helps to create strong structure. Research Paper
  • 21. Thank You . . .