Effluent Testing: Testing of BOD,
COD, TOC and interpretation of
results
for more information visit:-
http://www.textilerun.blogspot.com/
BOD (biochemical oxygen demand)
• BOD is the traditional, most widely used test to establish concentration of
organic matter in wastewater samples (i.e., relative strength).
• BOD is based on the principle that if sufficient oxygen is available, aerobic
biological decomposition (i.e., stabilization of organic waste) by
microorganisms will continue until all waste is consumed.
• The BOD test is also known as "BOD5" since it is based on the accurate
measure of DO (dissolved oxygen) at the beginning and end of a five-day
period in which the sample is held in dark, incubated conditions (i.e., 20°C or
68°F).
for more information visit:-
http://www.textilerun.blogspot.com/
•The change in DO concentration over five days represents the "oxygen
demand" for respiration by the aerobic biological microorganisms in the
sample.
•The five-day completion window is an inherent disadvantage of the test
because wastewater treatment system personnel cannot use it to make real-
time operational adjustments.
•An extended UBOD (ultimate BOD) test that measures oxygen consumption
after 60 days or more is sometimes required in wastewater permits.
for more information visit:-
http://www.textilerun.blogspot.com/
BOD Test Procedures
• To ensure proper biological activity during the BOD test, a
wastewater sample:
– Must be free of chlorine. If chlorine is present in the sample, a
dechlorination chemical (e.g, sodium sulfite) must be added prior to
testing.
– Needs to be in the pH range of 6.5-7.5 S.U. If the sample is outside this
range, then acid or base must be added as needed.
– Needs to have an existing adequate microbiological population. If the
microbial population is inadequate or unknown, a "seed" solution of
bacteria is added along with an essential nutrient buffer solution that
ensures bacteria population vitality.
for more information visit:-
http://www.textilerun.blogspot.com/
• Specialized 300 mL BOD bottles designed to allow full filling with no air space
and provide an airtight seal are used.
• The bottles are filled with the sample to be tested or dilution (distilled or
deionized) water and various amounts of the wastewater sample are added to
reflect different dilutions.
• At least one bottle is filled only with dilution water as a control or "blank."
• A DO meter is used to measure the initial dissolved oxygen concentration
(mg/L) in each bottle, which should be a least 8.0 mg/L. Each bottle in then
placed into a dark incubator at 20°C for five days.
for more information visit:-
http://www.textilerun.blogspot.com/
• After five days (± 3 hours) the DO meter is used again to measure a final
dissolved oxygen concentration (mg/L), which ideally will be a reduction
of at least 4.0 mg/L.
• The final DO reading is then subtracted from the initial DO reading and
the result is the BOD concentration (mg/L).
• If the wastewater sample required dilution, the BOD concentration
reading is multiplied by the dilution factor.
for more information visit:-
http://www.textilerun.blogspot.com/
What is DO (dissolved oxygen)?
• As the name implies, a DO test measures the concentration of oxygen
dissolved in a water or wastewater sample.
• DO measurement most often takes place using an electronic meter fitted with
a specialized DO probe.
• The concentration of DO in a water sample is significantly influenced by:
– Temperature: As water temperature increases, DO decreases (i.e., as
water gets warmer, it holds less oxygen) (Table 2).
– Salinity: As water salinity increases, DO decreases (i.e., as water gets
saltier, it holds less oxygen).
– Atmospheric Pressure: As pressure increases, DO also increases (i.e.,
water holds less oxygen as you increase altitude).
for more information visit:-
http://www.textilerun.blogspot.com/
COD (chemical oxygen demand)
• COD is the most popular alternative test to BOD for establishing the
concentration of organic matter in wastewater samples.
• The COD test only takes a few hours to complete, giving it a major
advantage over the 5-day BOD test.
• COD can test wastewater that is too toxic for the BOD test.
• The COD test should be considered an independent measure of the organic
matter in a wastewater sample rather than a substitute for the BOD test.
for more information visit:-
http://www.textilerun.blogspot.com/
• The COD test uses a chemical (potassium dichromate in a 50% sulfuric acid
solution) that “oxidizes” both organic (predominate) and inorganic substances
in a wastewater sample, which results in a higher COD concentration than
BOD concentration for the same wastewater sample since only organic
compounds are consumed during BOD testing.
• The most popular current testing method for COD involves using sealed and
heated (i.e., closed reflux) low-range (3-150 ppm) or high-range (20-1500
ppm) pre-prepared vials that change color from orange to green based on
the amount of oxidation and that are read using a laboratory colorimeter
for more information visit:-
http://www.textilerun.blogspot.com/
COD Test Procedures
• Prior to completing the COD test, a series of known standards are prepared using
KHP (potassium hydrogen phthalate).
• Most wastewater samples will fall in the high range, so standards of 100, 250, 500
and 1000 mg/L are typically prepared. COD standards can also be purchased.
• A COD reactor/heating (150°C) block and a colorimeter are TURNED ON so that
both instruments are allowed to stabilize.
• Pre-prepared low-range (3-50 ppm) or high-range (20-1500 ppm) vials are
selected for the COD test based on expected results. Both ranges can be used if
expected results are unknown.
for more information visit:-
http://www.textilerun.blogspot.com/
• One vial is marked as a “blank,” and three or four vials are marked with known
standard levels.
• Two vials are then marked for the wastewater sample to make a duplicate run.
Note: If multiple wastewater samples are being run, at least 10% of samples are
duplicated.
• 2 mL of liquid are added to each vial.
• In the case of the “blank,” 2 mL of DI water are added. 2 mL of each standard are
added to the corresponding vials.
• If the wastewater sample is tested at full strength, then 2 mL is added to the
corresponding vial.
• If dilution is required, then serial dilutions are performed and 2 mL of the diluted
sample are added to the corresponding vial.
• Each vial is mixed well and placed into the reactor block for two hours. After two
hours, the vials are removed from the block to a cooling rack for about 15
minutes.
• The colorimeter is set and calibrated per the specific instructions for that
unit (i.e., proper wavelength, blank and standards) and each vial is placed
in the unit and the COD concentration read.
• If the sample was diluted, the corresponding multiplication is made
for more information visit:-
http://www.textilerun.blogspot.com/
Can we use my COD results to predict my
BOD?
• YES. Although COD should be considered an independent test from BOD,
and will generate a higher concentration reading than BOD for a particular
wastewater sample, it is generally accepted that COD and BOD share an
empirical relationship. Extensive observation of the COD and BOD levels on
the same wastewater has shown that the COD to BOD ratio of a particular
wastewater will remain constant over time.
• For example, food processing wastewater will generally have a COD:BOD
ratio of ~2:1, while textile wastewater that can contain dyes will often have
a much higher COD:BOD ratio of ~5:1.
• To establish the COD:BOD ratio for your wastewater, simply have both COD
and BOD run on several wastewater samples. Divide the COD concentration
by the BOD concentration for each sample and average the results
for more information visit:-
http://www.textilerun.blogspot.com/
WARNING! COD…Hazardous Waste
• Along with the potassium dichromate in 50% sulfuric acid solution, pre-
prepared COD vials also contain silver sulfate as a catalyst and mercuric
sulfate to eliminate chloride interference.
• Thus, COD vials are considered HAZARDOUS WASTE and must be handled and
disposed of in an approved manner.
Do Not Dispose of COD Vial Contents Down the Drain!
• Most pre-prepared COD vial vendors will have a return policy for used COD vials
so that used vials can be returned to the vendor sealed in the original
containers for proper disposal.
for more information visit:-
http://www.textilerun.blogspot.com/
TOC (TOTAL ORGANIC CARBON)
• Total organic carbon (TOC) is the amount of carbon found in an organic
compound and is often used as a non-specific indicator of water
quality or cleanliness of pharmaceutical manufacturing equipment.
• The TOC test is gaining popularity because it only takes 5-10 minutes to
complete.
• Like COD, the TOC test can be used to rapidly estimate BOD concentration
once a consistent TOC to BOD ratio is established on a particular
wastewater stream.
• At the heart of the TOC test is a carbon analyzing instrument that
measures the total organic carbon in a wastewater sample.
for more information visit:-
http://www.textilerun.blogspot.com/
• Various heat and oxygen, ultraviolet radiation and
chemical oxidant-based methods are available to
measure TOC.
• In the TOC test, organic carbon is converted to carbon
dioxide (CO2) and typically measured with an infrared
analyzer.
for more information visit:-
http://www.textilerun.blogspot.com/
TOC Test Procedures
• TOC test procedures are relatively simple and straight-forward, but are specific to
the type of carbon-analyzing instrument utilized in the laboratory.
• Thus, no “typical” TOC procedure exists.
formoreinformationvisit:-
http://www.textilerun.blogspot.com/
How is Total Organic Carbon Measured?
• TOC is measured at very different concentrations in a very wide range of
systems. Levels vary widely within each type, but, broadly, they range from
sub-ppb levels in ultra-pure water for laboratory and microelectronic
applications up to hundreds of ppm in effluents and process streams.
• For many of these systems, the TOC alone does not provide enough
information.
• The carbon-containing compounds can be present in different forms and the
proportions of each can be critical.
• A breakdown of some of these divisions is shown in the diagram.
for more information visit:-
http://www.textilerun.blogspot.com/
for more information visit:-
http://www.textilerun.blogspot.com/
• Dissolved organic carbon (DOC) is generally taken as that which will pass
through a 0.45um filter.
• Large particle size TOC is classified as particulate or non-dissolved
(NDOC). About 50 to 75% of DOC in natural waters is in the form of
polymeric organic acids - fulvic and humic acids.
• About 10% of the TOC is in colloids, mainly humic acids and various
minerals.
• A further 10 to 20% are small molecules from the decomposition of
organic matter.
for more information visit:-
http://www.textilerun.blogspot.com/

Contenu connexe

Tendances

Validation of mixing, granulation, lubrication, compression and coating
Validation of mixing, granulation, lubrication, compression and coatingValidation of mixing, granulation, lubrication, compression and coating
Validation of mixing, granulation, lubrication, compression and coatingMalla Reddy College of Pharmacy
 
Process validation of Tablets and Lquids
Process validation of Tablets and LquidsProcess validation of Tablets and Lquids
Process validation of Tablets and LquidsRaheem Kurikkal
 
PILOT PLANT DESIGN FOR TABLETS
PILOT PLANT DESIGN FOR TABLETSPILOT PLANT DESIGN FOR TABLETS
PILOT PLANT DESIGN FOR TABLETSSuneal Saini
 
Introduction and scope of validation
Introduction and scope of validationIntroduction and scope of validation
Introduction and scope of validationJahnabi Sarmah
 
Qualification of FBD.pptx
Qualification of FBD.pptxQualification of FBD.pptx
Qualification of FBD.pptxDhruvi50
 
Qualification of tablet compression machine
Qualification of tablet compression machineQualification of tablet compression machine
Qualification of tablet compression machinePritam Kolge
 
Validation regulatory review
Validation regulatory reviewValidation regulatory review
Validation regulatory reviewArpitSuralkar
 
Process validation of tablet compression
Process validation of tablet compressionProcess validation of tablet compression
Process validation of tablet compressionSanket Shinde
 
PROBLEMS ENCOUNTERED COATING TECHNOLOGY
PROBLEMS ENCOUNTERED COATING TECHNOLOGYPROBLEMS ENCOUNTERED COATING TECHNOLOGY
PROBLEMS ENCOUNTERED COATING TECHNOLOGYPRATIKKUMARMORE
 
Role of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalRole of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalganpat420
 
Auditing of microbiological lab
Auditing of microbiological lab Auditing of microbiological lab
Auditing of microbiological lab KhushbooKunkulol
 

Tendances (20)

Pilot plant design for tablets and capsules
Pilot plant design for tablets and capsulesPilot plant design for tablets and capsules
Pilot plant design for tablets and capsules
 
Validation of mixing, granulation, lubrication, compression and coating
Validation of mixing, granulation, lubrication, compression and coatingValidation of mixing, granulation, lubrication, compression and coating
Validation of mixing, granulation, lubrication, compression and coating
 
Process validation of Tablets and Lquids
Process validation of Tablets and LquidsProcess validation of Tablets and Lquids
Process validation of Tablets and Lquids
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
PILOT PLANT DESIGN FOR TABLETS
PILOT PLANT DESIGN FOR TABLETSPILOT PLANT DESIGN FOR TABLETS
PILOT PLANT DESIGN FOR TABLETS
 
IQ-OQ-PQ-for-FBD.pptx
IQ-OQ-PQ-for-FBD.pptxIQ-OQ-PQ-for-FBD.pptx
IQ-OQ-PQ-for-FBD.pptx
 
Introduction and scope of validation
Introduction and scope of validationIntroduction and scope of validation
Introduction and scope of validation
 
Water system validation
Water system validationWater system validation
Water system validation
 
Ampule filling and_sealing_machine.ppt1
Ampule filling and_sealing_machine.ppt1Ampule filling and_sealing_machine.ppt1
Ampule filling and_sealing_machine.ppt1
 
Qualification of FBD.pptx
Qualification of FBD.pptxQualification of FBD.pptx
Qualification of FBD.pptx
 
Clean in-place (cip) systems
Clean  in-place (cip) systemsClean  in-place (cip) systems
Clean in-place (cip) systems
 
Pharmaceutical validation of water system
Pharmaceutical validation of  water system Pharmaceutical validation of  water system
Pharmaceutical validation of water system
 
Equipment qualification
Equipment qualificationEquipment qualification
Equipment qualification
 
Qualification of tablet compression machine
Qualification of tablet compression machineQualification of tablet compression machine
Qualification of tablet compression machine
 
Validation regulatory review
Validation regulatory reviewValidation regulatory review
Validation regulatory review
 
Process validation of tablet compression
Process validation of tablet compressionProcess validation of tablet compression
Process validation of tablet compression
 
PROBLEMS ENCOUNTERED COATING TECHNOLOGY
PROBLEMS ENCOUNTERED COATING TECHNOLOGYPROBLEMS ENCOUNTERED COATING TECHNOLOGY
PROBLEMS ENCOUNTERED COATING TECHNOLOGY
 
Role of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalRole of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceutical
 
Vendor qualification
Vendor qualification Vendor qualification
Vendor qualification
 
Auditing of microbiological lab
Auditing of microbiological lab Auditing of microbiological lab
Auditing of microbiological lab
 

Similaire à Effluent testing

BIOLOGICAL OXYGEN DEMAND
BIOLOGICAL OXYGEN DEMANDBIOLOGICAL OXYGEN DEMAND
BIOLOGICAL OXYGEN DEMANDHanu Pratap
 
Exp. 4 & 5 COD and BOD..pptx
Exp. 4 & 5 COD and BOD..pptxExp. 4 & 5 COD and BOD..pptx
Exp. 4 & 5 COD and BOD..pptxssuserb2c0be
 
ABG_water testing lab requirements_1-1-1.pptx
ABG_water testing  lab requirements_1-1-1.pptxABG_water testing  lab requirements_1-1-1.pptx
ABG_water testing lab requirements_1-1-1.pptxssuser77047f
 
Chemical oxygen demand
Chemical oxygen demandChemical oxygen demand
Chemical oxygen demandUsman Khawaja
 
Types of Tests Used to Determine BOD.pptx
Types of Tests Used to Determine BOD.pptxTypes of Tests Used to Determine BOD.pptx
Types of Tests Used to Determine BOD.pptxUet Peshawar
 
Chemical Parameters in Basic Environmental Technology-1.ppt
Chemical Parameters in Basic Environmental Technology-1.pptChemical Parameters in Basic Environmental Technology-1.ppt
Chemical Parameters in Basic Environmental Technology-1.pptssuser325d67
 
BOD Analysis as per APHA Manual
BOD Analysis as per APHA ManualBOD Analysis as per APHA Manual
BOD Analysis as per APHA ManualGayathri Mohan
 
Effluent Treatment Plant-Project-02
Effluent Treatment Plant-Project-02Effluent Treatment Plant-Project-02
Effluent Treatment Plant-Project-02Nilusha Rajashilpa
 
02 Characterization and Primary Treatment of Sewage
02 Characterization and Primary Treatment of Sewage02 Characterization and Primary Treatment of Sewage
02 Characterization and Primary Treatment of Sewageakashpadole
 
L8- ESTIMATION OF ORGANIC CONTENT.pptx
L8- ESTIMATION OF ORGANIC CONTENT.pptxL8- ESTIMATION OF ORGANIC CONTENT.pptx
L8- ESTIMATION OF ORGANIC CONTENT.pptxPRACHI DESSAI
 
Routine analysis of wastewaters quality parameters
Routine analysis of wastewaters quality parametersRoutine analysis of wastewaters quality parameters
Routine analysis of wastewaters quality parametersArvind Kumar
 

Similaire à Effluent testing (20)

BIOLOGICAL OXYGEN DEMAND
BIOLOGICAL OXYGEN DEMANDBIOLOGICAL OXYGEN DEMAND
BIOLOGICAL OXYGEN DEMAND
 
Exp. 4 & 5 COD and BOD..pptx
Exp. 4 & 5 COD and BOD..pptxExp. 4 & 5 COD and BOD..pptx
Exp. 4 & 5 COD and BOD..pptx
 
L 13 bod
L 13 bodL 13 bod
L 13 bod
 
ABG_water testing lab requirements_1-1-1.pptx
ABG_water testing  lab requirements_1-1-1.pptxABG_water testing  lab requirements_1-1-1.pptx
ABG_water testing lab requirements_1-1-1.pptx
 
Chemical oxygen demand
Chemical oxygen demandChemical oxygen demand
Chemical oxygen demand
 
L 16 and 17 bod
L 16 and 17 bodL 16 and 17 bod
L 16 and 17 bod
 
Lecture-3.pptx
Lecture-3.pptxLecture-3.pptx
Lecture-3.pptx
 
Types of Tests Used to Determine BOD.pptx
Types of Tests Used to Determine BOD.pptxTypes of Tests Used to Determine BOD.pptx
Types of Tests Used to Determine BOD.pptx
 
sewage determination by bod method
sewage determination by bod methodsewage determination by bod method
sewage determination by bod method
 
Chemical Parameters in Basic Environmental Technology-1.ppt
Chemical Parameters in Basic Environmental Technology-1.pptChemical Parameters in Basic Environmental Technology-1.ppt
Chemical Parameters in Basic Environmental Technology-1.ppt
 
BOD Analysis as per APHA Manual
BOD Analysis as per APHA ManualBOD Analysis as per APHA Manual
BOD Analysis as per APHA Manual
 
Effluent Treatment Plant-Project-02
Effluent Treatment Plant-Project-02Effluent Treatment Plant-Project-02
Effluent Treatment Plant-Project-02
 
02 Characterization and Primary Treatment of Sewage
02 Characterization and Primary Treatment of Sewage02 Characterization and Primary Treatment of Sewage
02 Characterization and Primary Treatment of Sewage
 
Bod
BodBod
Bod
 
Waste_Water_testing[1].pdf
Waste_Water_testing[1].pdfWaste_Water_testing[1].pdf
Waste_Water_testing[1].pdf
 
L8- ESTIMATION OF ORGANIC CONTENT.pptx
L8- ESTIMATION OF ORGANIC CONTENT.pptxL8- ESTIMATION OF ORGANIC CONTENT.pptx
L8- ESTIMATION OF ORGANIC CONTENT.pptx
 
PH, COD AND TSS
PH, COD AND TSSPH, COD AND TSS
PH, COD AND TSS
 
Biological oxygen demand
Biological oxygen demandBiological oxygen demand
Biological oxygen demand
 
Routine analysis of wastewaters quality parameters
Routine analysis of wastewaters quality parametersRoutine analysis of wastewaters quality parameters
Routine analysis of wastewaters quality parameters
 
sewage determination by COD method
sewage determination by COD methodsewage determination by COD method
sewage determination by COD method
 

Plus de Nitish Sharma

Chemical Processing Printing
Chemical Processing  PrintingChemical Processing  Printing
Chemical Processing PrintingNitish Sharma
 
Primary, secondary and tertiary treatments in ETP (Effluent Treatment Plant)
Primary, secondary and tertiary treatments in ETP (Effluent Treatment Plant)Primary, secondary and tertiary treatments in ETP (Effluent Treatment Plant)
Primary, secondary and tertiary treatments in ETP (Effluent Treatment Plant)Nitish Sharma
 
Chemical Proessing Finishing
Chemical Proessing FinishingChemical Proessing Finishing
Chemical Proessing FinishingNitish Sharma
 
Chemical Processing Dyeing
Chemical Processing DyeingChemical Processing Dyeing
Chemical Processing DyeingNitish Sharma
 
Chemical processing pretreatment
Chemical processing pretreatmentChemical processing pretreatment
Chemical processing pretreatmentNitish Sharma
 
Secondary motions in LOOM
Secondary motions in LOOMSecondary motions in LOOM
Secondary motions in LOOMNitish Sharma
 
FIBER LENGTH TESTING
FIBER LENGTH TESTINGFIBER LENGTH TESTING
FIBER LENGTH TESTINGNitish Sharma
 

Plus de Nitish Sharma (11)

Chemical Processing Printing
Chemical Processing  PrintingChemical Processing  Printing
Chemical Processing Printing
 
Primary, secondary and tertiary treatments in ETP (Effluent Treatment Plant)
Primary, secondary and tertiary treatments in ETP (Effluent Treatment Plant)Primary, secondary and tertiary treatments in ETP (Effluent Treatment Plant)
Primary, secondary and tertiary treatments in ETP (Effluent Treatment Plant)
 
Chemical Proessing Finishing
Chemical Proessing FinishingChemical Proessing Finishing
Chemical Proessing Finishing
 
Chemical Processing Dyeing
Chemical Processing DyeingChemical Processing Dyeing
Chemical Processing Dyeing
 
Chemical processing pretreatment
Chemical processing pretreatmentChemical processing pretreatment
Chemical processing pretreatment
 
Textile Testing
Textile TestingTextile Testing
Textile Testing
 
Fibrefiness
FibrefinessFibrefiness
Fibrefiness
 
FIBER LENGTH
FIBER LENGTHFIBER LENGTH
FIBER LENGTH
 
Secondary motions in LOOM
Secondary motions in LOOMSecondary motions in LOOM
Secondary motions in LOOM
 
FIBER LENGTH TESTING
FIBER LENGTH TESTINGFIBER LENGTH TESTING
FIBER LENGTH TESTING
 
FIBRE Sampling
 FIBRE Sampling FIBRE Sampling
FIBRE Sampling
 

Dernier

IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024Mark Billinghurst
 
Comparative Analysis of Text Summarization Techniques
Comparative Analysis of Text Summarization TechniquesComparative Analysis of Text Summarization Techniques
Comparative Analysis of Text Summarization Techniquesugginaramesh
 
Application of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptxApplication of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptx959SahilShah
 
Concrete Mix Design - IS 10262-2019 - .pptx
Concrete Mix Design - IS 10262-2019 - .pptxConcrete Mix Design - IS 10262-2019 - .pptx
Concrete Mix Design - IS 10262-2019 - .pptxKartikeyaDwivedi3
 
Risk Assessment For Installation of Drainage Pipes.pdf
Risk Assessment For Installation of Drainage Pipes.pdfRisk Assessment For Installation of Drainage Pipes.pdf
Risk Assessment For Installation of Drainage Pipes.pdfROCENODodongVILLACER
 
Introduction-To-Agricultural-Surveillance-Rover.pptx
Introduction-To-Agricultural-Surveillance-Rover.pptxIntroduction-To-Agricultural-Surveillance-Rover.pptx
Introduction-To-Agricultural-Surveillance-Rover.pptxk795866
 
An experimental study in using natural admixture as an alternative for chemic...
An experimental study in using natural admixture as an alternative for chemic...An experimental study in using natural admixture as an alternative for chemic...
An experimental study in using natural admixture as an alternative for chemic...Chandu841456
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AIabhishek36461
 
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...srsj9000
 
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)dollysharma2066
 
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)Dr SOUNDIRARAJ N
 
Introduction to Machine Learning Unit-3 for II MECH
Introduction to Machine Learning Unit-3 for II MECHIntroduction to Machine Learning Unit-3 for II MECH
Introduction to Machine Learning Unit-3 for II MECHC Sai Kiran
 
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfCCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfAsst.prof M.Gokilavani
 
Work Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvWork Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvLewisJB
 
Churning of Butter, Factors affecting .
Churning of Butter, Factors affecting  .Churning of Butter, Factors affecting  .
Churning of Butter, Factors affecting .Satyam Kumar
 
complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...asadnawaz62
 
8251 universal synchronous asynchronous receiver transmitter
8251 universal synchronous asynchronous receiver transmitter8251 universal synchronous asynchronous receiver transmitter
8251 universal synchronous asynchronous receiver transmitterShivangiSharma879191
 
INFLUENCE OF NANOSILICA ON THE PROPERTIES OF CONCRETE
INFLUENCE OF NANOSILICA ON THE PROPERTIES OF CONCRETEINFLUENCE OF NANOSILICA ON THE PROPERTIES OF CONCRETE
INFLUENCE OF NANOSILICA ON THE PROPERTIES OF CONCRETEroselinkalist12
 

Dernier (20)

IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024
 
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCRCall Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
 
Comparative Analysis of Text Summarization Techniques
Comparative Analysis of Text Summarization TechniquesComparative Analysis of Text Summarization Techniques
Comparative Analysis of Text Summarization Techniques
 
Application of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptxApplication of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptx
 
Concrete Mix Design - IS 10262-2019 - .pptx
Concrete Mix Design - IS 10262-2019 - .pptxConcrete Mix Design - IS 10262-2019 - .pptx
Concrete Mix Design - IS 10262-2019 - .pptx
 
Risk Assessment For Installation of Drainage Pipes.pdf
Risk Assessment For Installation of Drainage Pipes.pdfRisk Assessment For Installation of Drainage Pipes.pdf
Risk Assessment For Installation of Drainage Pipes.pdf
 
Introduction-To-Agricultural-Surveillance-Rover.pptx
Introduction-To-Agricultural-Surveillance-Rover.pptxIntroduction-To-Agricultural-Surveillance-Rover.pptx
Introduction-To-Agricultural-Surveillance-Rover.pptx
 
An experimental study in using natural admixture as an alternative for chemic...
An experimental study in using natural admixture as an alternative for chemic...An experimental study in using natural admixture as an alternative for chemic...
An experimental study in using natural admixture as an alternative for chemic...
 
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AI
 
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
 
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
 
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
 
Introduction to Machine Learning Unit-3 for II MECH
Introduction to Machine Learning Unit-3 for II MECHIntroduction to Machine Learning Unit-3 for II MECH
Introduction to Machine Learning Unit-3 for II MECH
 
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfCCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
 
Work Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvWork Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvv
 
Churning of Butter, Factors affecting .
Churning of Butter, Factors affecting  .Churning of Butter, Factors affecting  .
Churning of Butter, Factors affecting .
 
complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...
 
8251 universal synchronous asynchronous receiver transmitter
8251 universal synchronous asynchronous receiver transmitter8251 universal synchronous asynchronous receiver transmitter
8251 universal synchronous asynchronous receiver transmitter
 
INFLUENCE OF NANOSILICA ON THE PROPERTIES OF CONCRETE
INFLUENCE OF NANOSILICA ON THE PROPERTIES OF CONCRETEINFLUENCE OF NANOSILICA ON THE PROPERTIES OF CONCRETE
INFLUENCE OF NANOSILICA ON THE PROPERTIES OF CONCRETE
 

Effluent testing

  • 1. Effluent Testing: Testing of BOD, COD, TOC and interpretation of results for more information visit:- http://www.textilerun.blogspot.com/
  • 2. BOD (biochemical oxygen demand) • BOD is the traditional, most widely used test to establish concentration of organic matter in wastewater samples (i.e., relative strength). • BOD is based on the principle that if sufficient oxygen is available, aerobic biological decomposition (i.e., stabilization of organic waste) by microorganisms will continue until all waste is consumed. • The BOD test is also known as "BOD5" since it is based on the accurate measure of DO (dissolved oxygen) at the beginning and end of a five-day period in which the sample is held in dark, incubated conditions (i.e., 20°C or 68°F). for more information visit:- http://www.textilerun.blogspot.com/
  • 3. •The change in DO concentration over five days represents the "oxygen demand" for respiration by the aerobic biological microorganisms in the sample. •The five-day completion window is an inherent disadvantage of the test because wastewater treatment system personnel cannot use it to make real- time operational adjustments. •An extended UBOD (ultimate BOD) test that measures oxygen consumption after 60 days or more is sometimes required in wastewater permits. for more information visit:- http://www.textilerun.blogspot.com/
  • 4. BOD Test Procedures • To ensure proper biological activity during the BOD test, a wastewater sample: – Must be free of chlorine. If chlorine is present in the sample, a dechlorination chemical (e.g, sodium sulfite) must be added prior to testing. – Needs to be in the pH range of 6.5-7.5 S.U. If the sample is outside this range, then acid or base must be added as needed. – Needs to have an existing adequate microbiological population. If the microbial population is inadequate or unknown, a "seed" solution of bacteria is added along with an essential nutrient buffer solution that ensures bacteria population vitality. for more information visit:- http://www.textilerun.blogspot.com/
  • 5. • Specialized 300 mL BOD bottles designed to allow full filling with no air space and provide an airtight seal are used. • The bottles are filled with the sample to be tested or dilution (distilled or deionized) water and various amounts of the wastewater sample are added to reflect different dilutions. • At least one bottle is filled only with dilution water as a control or "blank." • A DO meter is used to measure the initial dissolved oxygen concentration (mg/L) in each bottle, which should be a least 8.0 mg/L. Each bottle in then placed into a dark incubator at 20°C for five days. for more information visit:- http://www.textilerun.blogspot.com/
  • 6. • After five days (± 3 hours) the DO meter is used again to measure a final dissolved oxygen concentration (mg/L), which ideally will be a reduction of at least 4.0 mg/L. • The final DO reading is then subtracted from the initial DO reading and the result is the BOD concentration (mg/L). • If the wastewater sample required dilution, the BOD concentration reading is multiplied by the dilution factor. for more information visit:- http://www.textilerun.blogspot.com/
  • 7. What is DO (dissolved oxygen)? • As the name implies, a DO test measures the concentration of oxygen dissolved in a water or wastewater sample. • DO measurement most often takes place using an electronic meter fitted with a specialized DO probe. • The concentration of DO in a water sample is significantly influenced by: – Temperature: As water temperature increases, DO decreases (i.e., as water gets warmer, it holds less oxygen) (Table 2). – Salinity: As water salinity increases, DO decreases (i.e., as water gets saltier, it holds less oxygen). – Atmospheric Pressure: As pressure increases, DO also increases (i.e., water holds less oxygen as you increase altitude). for more information visit:- http://www.textilerun.blogspot.com/
  • 8. COD (chemical oxygen demand) • COD is the most popular alternative test to BOD for establishing the concentration of organic matter in wastewater samples. • The COD test only takes a few hours to complete, giving it a major advantage over the 5-day BOD test. • COD can test wastewater that is too toxic for the BOD test. • The COD test should be considered an independent measure of the organic matter in a wastewater sample rather than a substitute for the BOD test. for more information visit:- http://www.textilerun.blogspot.com/
  • 9. • The COD test uses a chemical (potassium dichromate in a 50% sulfuric acid solution) that “oxidizes” both organic (predominate) and inorganic substances in a wastewater sample, which results in a higher COD concentration than BOD concentration for the same wastewater sample since only organic compounds are consumed during BOD testing. • The most popular current testing method for COD involves using sealed and heated (i.e., closed reflux) low-range (3-150 ppm) or high-range (20-1500 ppm) pre-prepared vials that change color from orange to green based on the amount of oxidation and that are read using a laboratory colorimeter for more information visit:- http://www.textilerun.blogspot.com/
  • 10. COD Test Procedures • Prior to completing the COD test, a series of known standards are prepared using KHP (potassium hydrogen phthalate). • Most wastewater samples will fall in the high range, so standards of 100, 250, 500 and 1000 mg/L are typically prepared. COD standards can also be purchased. • A COD reactor/heating (150°C) block and a colorimeter are TURNED ON so that both instruments are allowed to stabilize. • Pre-prepared low-range (3-50 ppm) or high-range (20-1500 ppm) vials are selected for the COD test based on expected results. Both ranges can be used if expected results are unknown. for more information visit:- http://www.textilerun.blogspot.com/
  • 11. • One vial is marked as a “blank,” and three or four vials are marked with known standard levels. • Two vials are then marked for the wastewater sample to make a duplicate run. Note: If multiple wastewater samples are being run, at least 10% of samples are duplicated. • 2 mL of liquid are added to each vial. • In the case of the “blank,” 2 mL of DI water are added. 2 mL of each standard are added to the corresponding vials. • If the wastewater sample is tested at full strength, then 2 mL is added to the corresponding vial. • If dilution is required, then serial dilutions are performed and 2 mL of the diluted sample are added to the corresponding vial. • Each vial is mixed well and placed into the reactor block for two hours. After two hours, the vials are removed from the block to a cooling rack for about 15 minutes.
  • 12. • The colorimeter is set and calibrated per the specific instructions for that unit (i.e., proper wavelength, blank and standards) and each vial is placed in the unit and the COD concentration read. • If the sample was diluted, the corresponding multiplication is made for more information visit:- http://www.textilerun.blogspot.com/
  • 13. Can we use my COD results to predict my BOD? • YES. Although COD should be considered an independent test from BOD, and will generate a higher concentration reading than BOD for a particular wastewater sample, it is generally accepted that COD and BOD share an empirical relationship. Extensive observation of the COD and BOD levels on the same wastewater has shown that the COD to BOD ratio of a particular wastewater will remain constant over time. • For example, food processing wastewater will generally have a COD:BOD ratio of ~2:1, while textile wastewater that can contain dyes will often have a much higher COD:BOD ratio of ~5:1. • To establish the COD:BOD ratio for your wastewater, simply have both COD and BOD run on several wastewater samples. Divide the COD concentration by the BOD concentration for each sample and average the results for more information visit:- http://www.textilerun.blogspot.com/
  • 14. WARNING! COD…Hazardous Waste • Along with the potassium dichromate in 50% sulfuric acid solution, pre- prepared COD vials also contain silver sulfate as a catalyst and mercuric sulfate to eliminate chloride interference. • Thus, COD vials are considered HAZARDOUS WASTE and must be handled and disposed of in an approved manner. Do Not Dispose of COD Vial Contents Down the Drain! • Most pre-prepared COD vial vendors will have a return policy for used COD vials so that used vials can be returned to the vendor sealed in the original containers for proper disposal. for more information visit:- http://www.textilerun.blogspot.com/
  • 15. TOC (TOTAL ORGANIC CARBON) • Total organic carbon (TOC) is the amount of carbon found in an organic compound and is often used as a non-specific indicator of water quality or cleanliness of pharmaceutical manufacturing equipment. • The TOC test is gaining popularity because it only takes 5-10 minutes to complete. • Like COD, the TOC test can be used to rapidly estimate BOD concentration once a consistent TOC to BOD ratio is established on a particular wastewater stream. • At the heart of the TOC test is a carbon analyzing instrument that measures the total organic carbon in a wastewater sample. for more information visit:- http://www.textilerun.blogspot.com/
  • 16. • Various heat and oxygen, ultraviolet radiation and chemical oxidant-based methods are available to measure TOC. • In the TOC test, organic carbon is converted to carbon dioxide (CO2) and typically measured with an infrared analyzer. for more information visit:- http://www.textilerun.blogspot.com/
  • 17. TOC Test Procedures • TOC test procedures are relatively simple and straight-forward, but are specific to the type of carbon-analyzing instrument utilized in the laboratory. • Thus, no “typical” TOC procedure exists. formoreinformationvisit:- http://www.textilerun.blogspot.com/
  • 18. How is Total Organic Carbon Measured? • TOC is measured at very different concentrations in a very wide range of systems. Levels vary widely within each type, but, broadly, they range from sub-ppb levels in ultra-pure water for laboratory and microelectronic applications up to hundreds of ppm in effluents and process streams. • For many of these systems, the TOC alone does not provide enough information. • The carbon-containing compounds can be present in different forms and the proportions of each can be critical. • A breakdown of some of these divisions is shown in the diagram. for more information visit:- http://www.textilerun.blogspot.com/
  • 19. for more information visit:- http://www.textilerun.blogspot.com/
  • 20. • Dissolved organic carbon (DOC) is generally taken as that which will pass through a 0.45um filter. • Large particle size TOC is classified as particulate or non-dissolved (NDOC). About 50 to 75% of DOC in natural waters is in the form of polymeric organic acids - fulvic and humic acids. • About 10% of the TOC is in colloids, mainly humic acids and various minerals. • A further 10 to 20% are small molecules from the decomposition of organic matter. for more information visit:- http://www.textilerun.blogspot.com/