Lecture 14: Emission Tomography III
Shahid Younas
NUCLEAR IMAGING
Emission Tomography III
Single Photon Emission Computed Tomography (SPECT)
Attenuation correction
Lecture 14: Emission Tomography III
X- or gamma rays that must traverse long paths through the patient produce
fewer counts, due to attenuation, than those from activity closer to the near
surface of the patient.
Introduction-Attenuation correction
Lecture 14: Emission Tomography III
 Images acquired with SPECT has,
 Poor spatial resolution
 Apparent decrease in activity
Introduction-Attenuation correction
Lecture 14: Emission Tomography III
 Transverse image slices of a phantom with a
uniform activity distribution will show a
gradual decrease in activity toward the center.
Introduction-Attenuation correction
Lecture 14: Emission Tomography III
 The primary mechanism for attenuation in tissue is Compton
Scattering.
 This changes photon direction with loss of energy.
 The change of direction results in missed count.
Introduction-Attenuation correction
Lecture 14: Emission Tomography III
The effects of attenuation
are more intense at lower
energies but are still
significant at the highest
energy value.
Introduction-Attenuation correction
Lecture 14: Emission Tomography III
Summing two planar
projection images
separated by 180.
Introduction-Attenuation correction
Lecture 14: Emission Tomography III
The magnitude of
attenuation effect depends
on the tissue type.
Attenuation correction
Lecture 14: Emission Tomography III
Thus, to accurately represent the activity distribution measured with SPECT,
it is necessary to accurately correct for the effects of attenuation.
Attenuation correction Techniques
Lecture 14: Emission Tomography III
 Approximate methods are available for attenuation correction.
 Change Method, assumes a constant attenuation coefficient
throughout the patient.
 Over-undercompensate-as attenuation is not uniform
Attenuation correction Techniques
Lecture 14: Emission Tomography III
Constant Attenuation Coefficient
A1 A1 A1
A1 A1 A1
A1 A1 A1
Attenuation correction Techniques
Lecture 14: Emission Tomography III
 Some SPECT cameras have radioactive sources to measure the
attenuation through the patient,
 After acquisition, the transmission projection data are reconstructed
to provide maps of tissue attenuation characteristics across transverse
sections of the patient, similar to x-ray CT images.
Attenuation correction Techniques
Lecture 14: Emission Tomography III
 Some SPECT cameras have radioactive sources to measure the
attenuation through the patient,
 Finally these attenuation maps are used during SPECT image
reconstruction to provide attenuation-corrected SPECT images.
Attenuation correction Techniques
Lecture 14: Emission Tomography III
 Transmission sources are available in several configurations,
 Scanning Collimated Line Sources
 Fixed Line Sources
Attenuation correction Techniques
Lecture 14: Emission Tomography III
 Transmission data usually acquired simultaneously with the
acquisition of the emission projection data,
 Performing the two separately poses significant problems in
the spatial alignment of the two data sets.
Attenuation correction Techniques
Lecture 14: Emission Tomography III
 Radionuclide used for transmission measurements is chosen to have
primary gamma-ray emissions that differ significantly in energy from
those of the radiopharmaceuticals.
 Separate energy windows are used
Attenuation correction Techniques
Lecture 14: Emission Tomography III
 Scattering of the higher energy photons in the patient and in the
detector causes some cross-talk in the lower energy window.
 AC using transmission sources is used Myocardial perfusion
imaging.
 AC using transmission sources is promising but it is still under
development.
SPECT Collimator
Lecture 14: Emission Tomography III
Most commonly used is the high-resolution parallel-hole collimator
 Fan-beam collimators mainly used for brain SPECT
 FOV decreases with distance from collimator
Multihead SPECT Cameras
Lecture 14: Emission Tomography III
 Two or three scintillation camera heads reduce limitations imposed
by collimation and limited time per view.
 Y-offsets and X- and Y-magnification factors of all heads must be
precisely matched throughout rotation.
SPECT Performance
Lecture 14: Emission Tomography III
 Spatial resolution
 X- and Y-magnification factors and multi-energy spatial registration
 Alignment of projection images to axis-of-rotation
 Uniformity
 Camera head tilt
SPECT Spatial resolution
Lecture 14: Emission Tomography III
 Can be measured by acquiring a SPECT study of a line source
(capillary tube filled with a solution of Tc-99m, placed parallel to
axis of rotation).
 FWHM of the line sources are determined from the reconstructed
transverse images (ramp filter).
SPECT Spatial resolution
Lecture 14: Emission Tomography III
 National Electrical
Manufacturers
Association (NEMA)
specifies a cylindrical
plastic water-filled
phantom, 22 cm in
diameter, containing 3
line sources
SPECT Spatial resolution
Lecture 14: Emission Tomography III
 NEMA spatial resolution measurements are primarily determined by
the collimator used.
 Tangential resolution 7 to 8 mm FWHM for LEHR
 central resolution 9.5 to 12 mm
 radial resolution 9.4 to 12 mm
SPECT Spatial resolution
Lecture 14: Emission Tomography III
 NEMA measurements not necessarily representative of clinical
performance
 Studies can be acquired using longer imaging times and closer orbits
than would be possible in a patient.
SPECT Spatial resolution
Lecture 14: Emission Tomography III
 NEMA measurements not necessarily representative of clinical
performance
 Studies can be acquired using longer imaging times and closer orbits
than would be possible in a patient.
SPECT Spatial resolution
Lecture 14: Emission Tomography III
 NEMA measurements not necessarily representative of clinical
performance
 Studies can be acquired using longer imaging times and closer orbits
than would be possible in a patient.
 Filters used for clinical studies have lower spatial frequency cutoffs
than the ramp filters used in NEMA measurements.
Comparison with conventional planar scintillation camera imaging
Lecture 14: Emission Tomography III
 In theory, SPECT should produce spatial resolution similar to that of
planar scintillation camera imaging.
 In clinical imaging, its resolution is usually slightly worse.
 Camera head is closer to patient in conventional planar imaging than
in SPECT.
Comparison with conventional planar scintillation camera imaging
Lecture 14: Emission Tomography III
 Short time per view of SPECT may mandate use of lower resolution
collimator to obtain adequate number of counts.
 In planar imaging, radioactivity in tissues in front of and behind an
organ of interest causes a reduction in contrast.
Comparison with conventional planar scintillation camera imaging
Lecture 14: Emission Tomography III
 Main advantage of SPECT is markedly improved contrast and
reduced structural noise produced by eliminating the activity in
overlapping structures.
 SPECT also offers promise of partial correction for effects of
attenuation and scattering of photons in the patient
Magnification factors
Lecture 14: Emission Tomography III
The X- and Y-magnification factors, often called X and Y gains,
related distances in the object being imaged, in the x and y directions, to
the numbers of pixels between the corresponding points in the resultant
image.
Magnification factors
Lecture 14: Emission Tomography III
 Magnification factors determined from a digital image of two point
sources placed against the camera’s collimator
If X- and Y-magnification factors are unequal, the projection images will
be distorted in shape, as will coronal, sagittal, and oblique images.
COR calibration
Lecture 14: Emission Tomography III
 The axis of rotation (AOR) is an imaginary reference line about
which the head or heads of a SPECT camera rotate.
 If a radioactive line source were placed on the AOR, each projection
image would depict a vertical straight line near the center of the
image.
COR calibration
Lecture 14: Emission Tomography III
 This projection of the AOR into the image is called the center of
rotation (COR).
 Ideally, the COR is aligned with the center, in the x-direction, of each
projection image.
COR calibration
Lecture 14: Emission Tomography III
 Misalignment may be mechanical or electronic.
 Camera head may not be exactly centered in the gantry.
COR calibration
Lecture 14: Emission Tomography III
COR Degradation and Sinogram
COR calibration
Lecture 14: Emission Tomography III
 COR misalignment causes a loss of spatial resolution in the resultant
transverse images.
 Large misalignment cause a point source to appear as “doughnut”.
 Doughnut are not centered in the image so can be distinguished from
“ring” artifacts produced by non-uniformities.
COR calibration
Lecture 14: Emission Tomography III
 COR alignment is assessed by placing a point source or line source in
the camera field of view.
 Projected imaged and or sinogram is analyzed by the camera’s
computer.
COR calibration
Lecture 14: Emission Tomography III
 Misalignment may be corrected by shifting each image in the x-
direction by the proper number of pixels prior to filtered back-
projection
 If COR misalignment varies with camera head angle, it can only be
corrected if computer permits angle-by-angle corrections.
Uniformity
Lecture 14: Emission Tomography III
 Nonuniformities that are not apparent in low-count daily uniformity
studies can cause significant artifacts in SPECT.
 Artifact appears in transverse images as a ring centered about the
AOR.
Uniformity
Lecture 14: Emission Tomography III
Cylinder filled with a uniform
radionuclide solution showing
a ring artifact due to non-
uniformity.
Uniformity
Lecture 14: Emission Tomography III
 Primary intrinsic causes of non-uniformity are,
a. Spatial non-linearities
 stretch the image in some areas
 reducing the local count density
 compress other areas of the images
 Increasing the count density
a. Local variation in the light collection efficiency
Uniformity
Lecture 14: Emission Tomography III
 Lookup table can not correct,
 Local variations in detection efficiency such as dents or
manufacturing defects in the collimators.
Uniformity
Lecture 14: Emission Tomography III
 High-count uniformity images used to determine pixel correction
factors,
 At least 30 million counts for 64 x 64 images
 At least 120 million counts for 128 x 128 images
 Collected every 1 or 2 weeks; separate images for each camera head
Camera head tilt
Lecture 14: Emission Tomography III
 Camera head or heads must be exactly parallel to the AOR.

Contenu connexe

Tendances

Nuclear Medicine - PET/CT
Nuclear Medicine - PET/CTNuclear Medicine - PET/CT
Nuclear Medicine - PET/CT@Saudi_nmc
 
Single photon emission computed tomography (spect)
Single photon emission computed tomography (spect)Single photon emission computed tomography (spect)
Single photon emission computed tomography (spect)Syed Hammad .
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct Muntaser S.Ahmad
 
Magnetic resonance (mr) spectroscopy
Magnetic resonance (mr) spectroscopyMagnetic resonance (mr) spectroscopy
Magnetic resonance (mr) spectroscopyMaajid Mohi ud din
 
importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapymakhhi
 
PET/MRI Current & Future Status
PET/MRI Current & Future StatusPET/MRI Current & Future Status
PET/MRI Current & Future Status@Saudi_nmc
 
Fluoroscopy for Residents in Radiology
Fluoroscopy for Residents in RadiologyFluoroscopy for Residents in Radiology
Fluoroscopy for Residents in RadiologyGirendra Shankar
 
Clinical Radiotherapy Planning basics for beginners
Clinical Radiotherapy Planning basics for beginnersClinical Radiotherapy Planning basics for beginners
Clinical Radiotherapy Planning basics for beginnersDina Barakat
 
Electronic portal imaging by rose wekesa
Electronic portal imaging by rose wekesaElectronic portal imaging by rose wekesa
Electronic portal imaging by rose wekesaKesho Conference
 
Basic principle of ct and ct generations
Basic principle of ct and ct generationsBasic principle of ct and ct generations
Basic principle of ct and ct generationsTarunGoyal66
 
Digital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemDigital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemMuhammad Arif Afridi
 
Image guided radiation therapy (2011)
Image guided radiation therapy (2011)Image guided radiation therapy (2011)
Image guided radiation therapy (2011)Parminder S. Basran
 

Tendances (20)

Nuclear Medicine - PET/CT
Nuclear Medicine - PET/CTNuclear Medicine - PET/CT
Nuclear Medicine - PET/CT
 
Gamma Camera
Gamma CameraGamma Camera
Gamma Camera
 
Single photon emission computed tomography (spect)
Single photon emission computed tomography (spect)Single photon emission computed tomography (spect)
Single photon emission computed tomography (spect)
 
Mri artifacts
Mri artifactsMri artifacts
Mri artifacts
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct
 
Magnetic resonance (mr) spectroscopy
Magnetic resonance (mr) spectroscopyMagnetic resonance (mr) spectroscopy
Magnetic resonance (mr) spectroscopy
 
importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapy
 
PET/MRI Current & Future Status
PET/MRI Current & Future StatusPET/MRI Current & Future Status
PET/MRI Current & Future Status
 
Fluoroscopy for Residents in Radiology
Fluoroscopy for Residents in RadiologyFluoroscopy for Residents in Radiology
Fluoroscopy for Residents in Radiology
 
Fluoroscopy
FluoroscopyFluoroscopy
Fluoroscopy
 
Gamma Camera
Gamma CameraGamma Camera
Gamma Camera
 
Qc
QcQc
Qc
 
Clinical Radiotherapy Planning basics for beginners
Clinical Radiotherapy Planning basics for beginnersClinical Radiotherapy Planning basics for beginners
Clinical Radiotherapy Planning basics for beginners
 
CT artifact
CT artifact CT artifact
CT artifact
 
Electronic portal imaging by rose wekesa
Electronic portal imaging by rose wekesaElectronic portal imaging by rose wekesa
Electronic portal imaging by rose wekesa
 
Ct Basics
Ct BasicsCt Basics
Ct Basics
 
Basic principle of ct and ct generations
Basic principle of ct and ct generationsBasic principle of ct and ct generations
Basic principle of ct and ct generations
 
Digital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemDigital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging System
 
Image guided radiation therapy (2011)
Image guided radiation therapy (2011)Image guided radiation therapy (2011)
Image guided radiation therapy (2011)
 
IMRT and 3DCRT
IMRT and 3DCRT IMRT and 3DCRT
IMRT and 3DCRT
 

En vedette

Chang’s attenuation correction method hossein aslian
Chang’s attenuation correction method  hossein aslianChang’s attenuation correction method  hossein aslian
Chang’s attenuation correction method hossein aslianHossein Aslian
 
Nuclear imaging, PET CT MEDICAL PHYSICS
Nuclear imaging, PET CT MEDICAL PHYSICSNuclear imaging, PET CT MEDICAL PHYSICS
Nuclear imaging, PET CT MEDICAL PHYSICSShahid Younas
 
L 3 ct physics, Shahid Younas, Sr. Medical Physicist
L 3  ct physics, Shahid Younas, Sr. Medical PhysicistL 3  ct physics, Shahid Younas, Sr. Medical Physicist
L 3 ct physics, Shahid Younas, Sr. Medical PhysicistShahid Younas
 
Medical physics
Medical physicsMedical physics
Medical physicssrujan7777
 
Digital Techniques For Myocardial Perfusion Spect
Digital Techniques For Myocardial Perfusion SpectDigital Techniques For Myocardial Perfusion Spect
Digital Techniques For Myocardial Perfusion SpectMuhammad Ayub
 
Medical Physics Club Presentation Sept 2015
Medical Physics Club Presentation Sept 2015Medical Physics Club Presentation Sept 2015
Medical Physics Club Presentation Sept 2015Devrhoid Davis
 
2 radioactivity and nuclear transformation
2 radioactivity and nuclear transformation2 radioactivity and nuclear transformation
2 radioactivity and nuclear transformationShahid Younas
 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteShahid Younas
 
Radioactivity and nuclear transformation
Radioactivity and nuclear transformationRadioactivity and nuclear transformation
Radioactivity and nuclear transformationShahid Younas
 
Glimpse of clinical radiobiology course
Glimpse of clinical radiobiology courseGlimpse of clinical radiobiology course
Glimpse of clinical radiobiology courseManoj Gupta
 
Two Dimensional Image Reconstruction Algorithms
Two Dimensional Image Reconstruction AlgorithmsTwo Dimensional Image Reconstruction Algorithms
Two Dimensional Image Reconstruction Algorithmsmastersrihari
 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement Shahid Younas
 

En vedette (20)

SPECT
SPECTSPECT
SPECT
 
Chang’s attenuation correction method hossein aslian
Chang’s attenuation correction method  hossein aslianChang’s attenuation correction method  hossein aslian
Chang’s attenuation correction method hossein aslian
 
Nuclear imaging, PET CT MEDICAL PHYSICS
Nuclear imaging, PET CT MEDICAL PHYSICSNuclear imaging, PET CT MEDICAL PHYSICS
Nuclear imaging, PET CT MEDICAL PHYSICS
 
L 3 ct physics, Shahid Younas, Sr. Medical Physicist
L 3  ct physics, Shahid Younas, Sr. Medical PhysicistL 3  ct physics, Shahid Younas, Sr. Medical Physicist
L 3 ct physics, Shahid Younas, Sr. Medical Physicist
 
Linear Accelerator
Linear AcceleratorLinear Accelerator
Linear Accelerator
 
SPECT/PET Scans
SPECT/PET ScansSPECT/PET Scans
SPECT/PET Scans
 
Medical physics
Medical physicsMedical physics
Medical physics
 
Digital Techniques For Myocardial Perfusion Spect
Digital Techniques For Myocardial Perfusion SpectDigital Techniques For Myocardial Perfusion Spect
Digital Techniques For Myocardial Perfusion Spect
 
Medical Physics Club Presentation Sept 2015
Medical Physics Club Presentation Sept 2015Medical Physics Club Presentation Sept 2015
Medical Physics Club Presentation Sept 2015
 
spect and pet
spect and petspect and pet
spect and pet
 
Lecture 12
Lecture 12Lecture 12
Lecture 12
 
2 radioactivity and nuclear transformation
2 radioactivity and nuclear transformation2 radioactivity and nuclear transformation
2 radioactivity and nuclear transformation
 
Ct physics/ dental implant courses
Ct physics/ dental implant coursesCt physics/ dental implant courses
Ct physics/ dental implant courses
 
Nuclear imaging
Nuclear imagingNuclear imaging
Nuclear imaging
 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT Constitute
 
Radioactivity and nuclear transformation
Radioactivity and nuclear transformationRadioactivity and nuclear transformation
Radioactivity and nuclear transformation
 
Glimpse of clinical radiobiology course
Glimpse of clinical radiobiology courseGlimpse of clinical radiobiology course
Glimpse of clinical radiobiology course
 
Two Dimensional Image Reconstruction Algorithms
Two Dimensional Image Reconstruction AlgorithmsTwo Dimensional Image Reconstruction Algorithms
Two Dimensional Image Reconstruction Algorithms
 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement
 
physics lecture
physics lecturephysics lecture
physics lecture
 

Similaire à Medical Physics Imaging PET CT SPECT CT Lecture

Alternatives to Point-Scan Confocal Microscopy
Alternatives to Point-Scan Confocal MicroscopyAlternatives to Point-Scan Confocal Microscopy
Alternatives to Point-Scan Confocal Microscopymchelen
 
CT Image reconstruction
CT Image reconstructionCT Image reconstruction
CT Image reconstructionSantosh Ojha
 
Modeling of Optical Scattering in Advanced LIGO
Modeling of Optical Scattering in Advanced LIGOModeling of Optical Scattering in Advanced LIGO
Modeling of Optical Scattering in Advanced LIGOHunter Rew
 
Qualitative analysis of Fruits and Vegetables using Earth’s Field Nuclear Mag...
Qualitative analysis of Fruits and Vegetables using Earth’s Field Nuclear Mag...Qualitative analysis of Fruits and Vegetables using Earth’s Field Nuclear Mag...
Qualitative analysis of Fruits and Vegetables using Earth’s Field Nuclear Mag...IJERA Editor
 
System testing for the fresnel lens-based optical
System testing for the fresnel lens-based opticalSystem testing for the fresnel lens-based optical
System testing for the fresnel lens-based opticalJuan Milton Garduno Rubio
 
Total Variation-Based Reduction of Streak Artifacts, Ring Artifacts and Noise...
Total Variation-Based Reduction of Streak Artifacts, Ring Artifacts and Noise...Total Variation-Based Reduction of Streak Artifacts, Ring Artifacts and Noise...
Total Variation-Based Reduction of Streak Artifacts, Ring Artifacts and Noise...Jan Michálek
 
Iisrt zz saravana kumar
Iisrt zz saravana kumarIisrt zz saravana kumar
Iisrt zz saravana kumarIISRT
 
computedtomography-170122041110.pdf
computedtomography-170122041110.pdfcomputedtomography-170122041110.pdf
computedtomography-170122041110.pdfVanshikaGarg76
 
CT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QACT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QASambasivaselli R
 
Gold presentation lastri
Gold presentation lastriGold presentation lastri
Gold presentation lastrigrssieee
 
Structured Illumination Microscopy - Fundamentals and Applications
Structured Illumination Microscopy - Fundamentals and ApplicationsStructured Illumination Microscopy - Fundamentals and Applications
Structured Illumination Microscopy - Fundamentals and ApplicationsRafael Camacho Dejay
 
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...Prem Murti
 

Similaire à Medical Physics Imaging PET CT SPECT CT Lecture (20)

X ray tomography
X ray tomographyX ray tomography
X ray tomography
 
Alternatives to Point-Scan Confocal Microscopy
Alternatives to Point-Scan Confocal MicroscopyAlternatives to Point-Scan Confocal Microscopy
Alternatives to Point-Scan Confocal Microscopy
 
CT Image reconstruction
CT Image reconstructionCT Image reconstruction
CT Image reconstruction
 
Modeling of Optical Scattering in Advanced LIGO
Modeling of Optical Scattering in Advanced LIGOModeling of Optical Scattering in Advanced LIGO
Modeling of Optical Scattering in Advanced LIGO
 
Qualitative analysis of Fruits and Vegetables using Earth’s Field Nuclear Mag...
Qualitative analysis of Fruits and Vegetables using Earth’s Field Nuclear Mag...Qualitative analysis of Fruits and Vegetables using Earth’s Field Nuclear Mag...
Qualitative analysis of Fruits and Vegetables using Earth’s Field Nuclear Mag...
 
System testing for the fresnel lens-based optical
System testing for the fresnel lens-based opticalSystem testing for the fresnel lens-based optical
System testing for the fresnel lens-based optical
 
CT History
CT HistoryCT History
CT History
 
Minor_project
Minor_projectMinor_project
Minor_project
 
Optical coherence tomography
Optical coherence tomographyOptical coherence tomography
Optical coherence tomography
 
Paper 1 (2019)
Paper 1 (2019)Paper 1 (2019)
Paper 1 (2019)
 
Total Variation-Based Reduction of Streak Artifacts, Ring Artifacts and Noise...
Total Variation-Based Reduction of Streak Artifacts, Ring Artifacts and Noise...Total Variation-Based Reduction of Streak Artifacts, Ring Artifacts and Noise...
Total Variation-Based Reduction of Streak Artifacts, Ring Artifacts and Noise...
 
Iisrt zz saravana kumar
Iisrt zz saravana kumarIisrt zz saravana kumar
Iisrt zz saravana kumar
 
computedtomography-170122041110.pdf
computedtomography-170122041110.pdfcomputedtomography-170122041110.pdf
computedtomography-170122041110.pdf
 
Computed tomography
Computed tomographyComputed tomography
Computed tomography
 
CT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QACT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QA
 
Gold presentation lastri
Gold presentation lastriGold presentation lastri
Gold presentation lastri
 
Study of the Mesh Field and Rotation in Various Planes of Optical Spherical N...
Study of the Mesh Field and Rotation in Various Planes of Optical Spherical N...Study of the Mesh Field and Rotation in Various Planes of Optical Spherical N...
Study of the Mesh Field and Rotation in Various Planes of Optical Spherical N...
 
Structured Illumination Microscopy - Fundamentals and Applications
Structured Illumination Microscopy - Fundamentals and ApplicationsStructured Illumination Microscopy - Fundamentals and Applications
Structured Illumination Microscopy - Fundamentals and Applications
 
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
 
CT dose reduction
CT dose reductionCT dose reduction
CT dose reduction
 

Plus de Shahid Younas

Skm symps poster 2015 spect ct
Skm symps poster 2015 spect ctSkm symps poster 2015 spect ct
Skm symps poster 2015 spect ctShahid Younas
 
Nuclear imaging PET CT Imaging Medical Physics Nuclear Medicine
Nuclear imaging PET CT Imaging Medical Physics Nuclear MedicineNuclear imaging PET CT Imaging Medical Physics Nuclear Medicine
Nuclear imaging PET CT Imaging Medical Physics Nuclear MedicineShahid Younas
 
1 interaction of radiation with matter
1 interaction of radiation with matter1 interaction of radiation with matter
1 interaction of radiation with matterShahid Younas
 
1 interaction of radiation with matter
1 interaction of radiation with matter1 interaction of radiation with matter
1 interaction of radiation with matterShahid Younas
 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement Shahid Younas
 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement Shahid Younas
 
1 radiation detection and measurement
1 radiation detection and measurement 1 radiation detection and measurement
1 radiation detection and measurement Shahid Younas
 

Plus de Shahid Younas (13)

Skm symps poster 2015 spect ct
Skm symps poster 2015 spect ctSkm symps poster 2015 spect ct
Skm symps poster 2015 spect ct
 
Nuclear imaging PET CT Imaging Medical Physics Nuclear Medicine
Nuclear imaging PET CT Imaging Medical Physics Nuclear MedicineNuclear imaging PET CT Imaging Medical Physics Nuclear Medicine
Nuclear imaging PET CT Imaging Medical Physics Nuclear Medicine
 
1 interaction of radiation with matter
1 interaction of radiation with matter1 interaction of radiation with matter
1 interaction of radiation with matter
 
1 interaction of radiation with matter
1 interaction of radiation with matter1 interaction of radiation with matter
1 interaction of radiation with matter
 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement
 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement
 
1 radiation detection and measurement
1 radiation detection and measurement 1 radiation detection and measurement
1 radiation detection and measurement
 
L 7 ct physics
L 7 ct physics L 7 ct physics
L 7 ct physics
 
L 6 ct physics
L 6 ct physicsL 6 ct physics
L 6 ct physics
 
L 5 ct physics
L 5 ct physicsL 5 ct physics
L 5 ct physics
 
L 4 ct physics
L 4  ct physicsL 4  ct physics
L 4 ct physics
 
L 2 ct physics
L 2  ct physics L 2  ct physics
L 2 ct physics
 
L 1 ct physics
L 1  ct physicsL 1  ct physics
L 1 ct physics
 

Dernier

Lipid Profile test & Cardiac Markers for MBBS, Lab. Med. and Nursing.pptx
Lipid Profile test & Cardiac Markers for MBBS, Lab. Med. and Nursing.pptxLipid Profile test & Cardiac Markers for MBBS, Lab. Med. and Nursing.pptx
Lipid Profile test & Cardiac Markers for MBBS, Lab. Med. and Nursing.pptxRajendra Dev Bhatt
 
Medisep insurance policy , new kerala government insurance policy for govrnm...
Medisep insurance policy , new  kerala government insurance policy for govrnm...Medisep insurance policy , new  kerala government insurance policy for govrnm...
Medisep insurance policy , new kerala government insurance policy for govrnm...LinshaLichu1
 
Mark-Klimek-Lectures-1-To-12 NCLEX EXAMINATION
Mark-Klimek-Lectures-1-To-12 NCLEX EXAMINATIONMark-Klimek-Lectures-1-To-12 NCLEX EXAMINATION
Mark-Klimek-Lectures-1-To-12 NCLEX EXAMINATIONes5735583
 
Unlocking the Mysteries of the Lymphatic System
Unlocking the Mysteries of the Lymphatic SystemUnlocking the Mysteries of the Lymphatic System
Unlocking the Mysteries of the Lymphatic SystemSasikiranMarri
 
Subconjunctival Haemorrhage,causes,treatment..pptx
Subconjunctival Haemorrhage,causes,treatment..pptxSubconjunctival Haemorrhage,causes,treatment..pptx
Subconjunctival Haemorrhage,causes,treatment..pptxvideosfildr
 
Discover the Art Deco Style at Spa Dental
Discover the Art Deco Style at Spa DentalDiscover the Art Deco Style at Spa Dental
Discover the Art Deco Style at Spa DentalA-dec Australia
 
Your Radiotherapy Destination Gokuldas Hospital.
Your Radiotherapy Destination Gokuldas Hospital.Your Radiotherapy Destination Gokuldas Hospital.
Your Radiotherapy Destination Gokuldas Hospital.Gokuldas Hospital
 
Disseminated Intravascular Coagulation.ppt
Disseminated Intravascular Coagulation.pptDisseminated Intravascular Coagulation.ppt
Disseminated Intravascular Coagulation.pptSameer Jain
 
LABORATORY PROCEDURES-ALTERATION IN THE ENDOCRINE SYSTEM
LABORATORY PROCEDURES-ALTERATION IN THE ENDOCRINE SYSTEMLABORATORY PROCEDURES-ALTERATION IN THE ENDOCRINE SYSTEM
LABORATORY PROCEDURES-ALTERATION IN THE ENDOCRINE SYSTEMRommel Luis III Israel
 
EHR Market Growth is The Boom Over - Jasper Colin
EHR Market Growth is The Boom Over - Jasper ColinEHR Market Growth is The Boom Over - Jasper Colin
EHR Market Growth is The Boom Over - Jasper ColinJasper Colin
 
Advance Directives and Advance Care Planning: Ensuring Patient Voices Are Heard
Advance Directives and Advance Care Planning: Ensuring Patient Voices Are HeardAdvance Directives and Advance Care Planning: Ensuring Patient Voices Are Heard
Advance Directives and Advance Care Planning: Ensuring Patient Voices Are HeardVITASAuthor
 
Family Adoption Programme first year mbbs.pptx
Family Adoption Programme first year mbbs.pptxFamily Adoption Programme first year mbbs.pptx
Family Adoption Programme first year mbbs.pptxKritikaMishra43
 
Text Neck Syndrome and its probable way out.pptx
Text Neck Syndrome and its probable way out.pptxText Neck Syndrome and its probable way out.pptx
Text Neck Syndrome and its probable way out.pptxProf. Satyen Bhattacharyya
 
AI in Pediatrics: Taking Baby Steps in the Big World of Data
AI in Pediatrics: Taking Baby Steps in the Big World of DataAI in Pediatrics: Taking Baby Steps in the Big World of Data
AI in Pediatrics: Taking Baby Steps in the Big World of DataIris Thiele Isip-Tan
 
Learn Tips for Managing Chemobrain or Mental Fogginess
Learn Tips for Managing Chemobrain or Mental FogginessLearn Tips for Managing Chemobrain or Mental Fogginess
Learn Tips for Managing Chemobrain or Mental Fogginessbkling
 
Back care and back massage. powerpoint presentation
Back care and back massage. powerpoint presentationBack care and back massage. powerpoint presentation
Back care and back massage. powerpoint presentationpratiksha ghimire
 
Incentive spirometry powerpoint presentation
Incentive spirometry powerpoint presentationIncentive spirometry powerpoint presentation
Incentive spirometry powerpoint presentationpratiksha ghimire
 
Understanding Cholera: Epidemiology, Prevention, and Control.pdf
Understanding Cholera: Epidemiology, Prevention, and Control.pdfUnderstanding Cholera: Epidemiology, Prevention, and Control.pdf
Understanding Cholera: Epidemiology, Prevention, and Control.pdfSasikiranMarri
 

Dernier (20)

Lipid Profile test & Cardiac Markers for MBBS, Lab. Med. and Nursing.pptx
Lipid Profile test & Cardiac Markers for MBBS, Lab. Med. and Nursing.pptxLipid Profile test & Cardiac Markers for MBBS, Lab. Med. and Nursing.pptx
Lipid Profile test & Cardiac Markers for MBBS, Lab. Med. and Nursing.pptx
 
Medisep insurance policy , new kerala government insurance policy for govrnm...
Medisep insurance policy , new  kerala government insurance policy for govrnm...Medisep insurance policy , new  kerala government insurance policy for govrnm...
Medisep insurance policy , new kerala government insurance policy for govrnm...
 
Mark-Klimek-Lectures-1-To-12 NCLEX EXAMINATION
Mark-Klimek-Lectures-1-To-12 NCLEX EXAMINATIONMark-Klimek-Lectures-1-To-12 NCLEX EXAMINATION
Mark-Klimek-Lectures-1-To-12 NCLEX EXAMINATION
 
Unlocking the Mysteries of the Lymphatic System
Unlocking the Mysteries of the Lymphatic SystemUnlocking the Mysteries of the Lymphatic System
Unlocking the Mysteries of the Lymphatic System
 
Subconjunctival Haemorrhage,causes,treatment..pptx
Subconjunctival Haemorrhage,causes,treatment..pptxSubconjunctival Haemorrhage,causes,treatment..pptx
Subconjunctival Haemorrhage,causes,treatment..pptx
 
Discover the Art Deco Style at Spa Dental
Discover the Art Deco Style at Spa DentalDiscover the Art Deco Style at Spa Dental
Discover the Art Deco Style at Spa Dental
 
Your Radiotherapy Destination Gokuldas Hospital.
Your Radiotherapy Destination Gokuldas Hospital.Your Radiotherapy Destination Gokuldas Hospital.
Your Radiotherapy Destination Gokuldas Hospital.
 
Disseminated Intravascular Coagulation.ppt
Disseminated Intravascular Coagulation.pptDisseminated Intravascular Coagulation.ppt
Disseminated Intravascular Coagulation.ppt
 
LABORATORY PROCEDURES-ALTERATION IN THE ENDOCRINE SYSTEM
LABORATORY PROCEDURES-ALTERATION IN THE ENDOCRINE SYSTEMLABORATORY PROCEDURES-ALTERATION IN THE ENDOCRINE SYSTEM
LABORATORY PROCEDURES-ALTERATION IN THE ENDOCRINE SYSTEM
 
EHR Market Growth is The Boom Over - Jasper Colin
EHR Market Growth is The Boom Over - Jasper ColinEHR Market Growth is The Boom Over - Jasper Colin
EHR Market Growth is The Boom Over - Jasper Colin
 
Advance Directives and Advance Care Planning: Ensuring Patient Voices Are Heard
Advance Directives and Advance Care Planning: Ensuring Patient Voices Are HeardAdvance Directives and Advance Care Planning: Ensuring Patient Voices Are Heard
Advance Directives and Advance Care Planning: Ensuring Patient Voices Are Heard
 
Check Your own POSTURE & treat yourself.pptx
Check Your own POSTURE & treat yourself.pptxCheck Your own POSTURE & treat yourself.pptx
Check Your own POSTURE & treat yourself.pptx
 
Family Adoption Programme first year mbbs.pptx
Family Adoption Programme first year mbbs.pptxFamily Adoption Programme first year mbbs.pptx
Family Adoption Programme first year mbbs.pptx
 
DELIRIUM psychiatric delirium is a organic mental disorder
DELIRIUM  psychiatric  delirium is a organic mental disorderDELIRIUM  psychiatric  delirium is a organic mental disorder
DELIRIUM psychiatric delirium is a organic mental disorder
 
Text Neck Syndrome and its probable way out.pptx
Text Neck Syndrome and its probable way out.pptxText Neck Syndrome and its probable way out.pptx
Text Neck Syndrome and its probable way out.pptx
 
AI in Pediatrics: Taking Baby Steps in the Big World of Data
AI in Pediatrics: Taking Baby Steps in the Big World of DataAI in Pediatrics: Taking Baby Steps in the Big World of Data
AI in Pediatrics: Taking Baby Steps in the Big World of Data
 
Learn Tips for Managing Chemobrain or Mental Fogginess
Learn Tips for Managing Chemobrain or Mental FogginessLearn Tips for Managing Chemobrain or Mental Fogginess
Learn Tips for Managing Chemobrain or Mental Fogginess
 
Back care and back massage. powerpoint presentation
Back care and back massage. powerpoint presentationBack care and back massage. powerpoint presentation
Back care and back massage. powerpoint presentation
 
Incentive spirometry powerpoint presentation
Incentive spirometry powerpoint presentationIncentive spirometry powerpoint presentation
Incentive spirometry powerpoint presentation
 
Understanding Cholera: Epidemiology, Prevention, and Control.pdf
Understanding Cholera: Epidemiology, Prevention, and Control.pdfUnderstanding Cholera: Epidemiology, Prevention, and Control.pdf
Understanding Cholera: Epidemiology, Prevention, and Control.pdf
 

Medical Physics Imaging PET CT SPECT CT Lecture

  • 1. Lecture 14: Emission Tomography III Shahid Younas NUCLEAR IMAGING Emission Tomography III Single Photon Emission Computed Tomography (SPECT)
  • 2. Attenuation correction Lecture 14: Emission Tomography III X- or gamma rays that must traverse long paths through the patient produce fewer counts, due to attenuation, than those from activity closer to the near surface of the patient.
  • 3. Introduction-Attenuation correction Lecture 14: Emission Tomography III  Images acquired with SPECT has,  Poor spatial resolution  Apparent decrease in activity
  • 4. Introduction-Attenuation correction Lecture 14: Emission Tomography III  Transverse image slices of a phantom with a uniform activity distribution will show a gradual decrease in activity toward the center.
  • 5. Introduction-Attenuation correction Lecture 14: Emission Tomography III  The primary mechanism for attenuation in tissue is Compton Scattering.  This changes photon direction with loss of energy.  The change of direction results in missed count.
  • 6. Introduction-Attenuation correction Lecture 14: Emission Tomography III The effects of attenuation are more intense at lower energies but are still significant at the highest energy value.
  • 7. Introduction-Attenuation correction Lecture 14: Emission Tomography III Summing two planar projection images separated by 180.
  • 8. Introduction-Attenuation correction Lecture 14: Emission Tomography III The magnitude of attenuation effect depends on the tissue type.
  • 9. Attenuation correction Lecture 14: Emission Tomography III Thus, to accurately represent the activity distribution measured with SPECT, it is necessary to accurately correct for the effects of attenuation.
  • 10. Attenuation correction Techniques Lecture 14: Emission Tomography III  Approximate methods are available for attenuation correction.  Change Method, assumes a constant attenuation coefficient throughout the patient.  Over-undercompensate-as attenuation is not uniform
  • 11. Attenuation correction Techniques Lecture 14: Emission Tomography III Constant Attenuation Coefficient A1 A1 A1 A1 A1 A1 A1 A1 A1
  • 12. Attenuation correction Techniques Lecture 14: Emission Tomography III  Some SPECT cameras have radioactive sources to measure the attenuation through the patient,  After acquisition, the transmission projection data are reconstructed to provide maps of tissue attenuation characteristics across transverse sections of the patient, similar to x-ray CT images.
  • 13. Attenuation correction Techniques Lecture 14: Emission Tomography III  Some SPECT cameras have radioactive sources to measure the attenuation through the patient,  Finally these attenuation maps are used during SPECT image reconstruction to provide attenuation-corrected SPECT images.
  • 14. Attenuation correction Techniques Lecture 14: Emission Tomography III  Transmission sources are available in several configurations,  Scanning Collimated Line Sources  Fixed Line Sources
  • 15. Attenuation correction Techniques Lecture 14: Emission Tomography III  Transmission data usually acquired simultaneously with the acquisition of the emission projection data,  Performing the two separately poses significant problems in the spatial alignment of the two data sets.
  • 16. Attenuation correction Techniques Lecture 14: Emission Tomography III  Radionuclide used for transmission measurements is chosen to have primary gamma-ray emissions that differ significantly in energy from those of the radiopharmaceuticals.  Separate energy windows are used
  • 17. Attenuation correction Techniques Lecture 14: Emission Tomography III  Scattering of the higher energy photons in the patient and in the detector causes some cross-talk in the lower energy window.  AC using transmission sources is used Myocardial perfusion imaging.  AC using transmission sources is promising but it is still under development.
  • 18. SPECT Collimator Lecture 14: Emission Tomography III Most commonly used is the high-resolution parallel-hole collimator  Fan-beam collimators mainly used for brain SPECT  FOV decreases with distance from collimator
  • 19. Multihead SPECT Cameras Lecture 14: Emission Tomography III  Two or three scintillation camera heads reduce limitations imposed by collimation and limited time per view.  Y-offsets and X- and Y-magnification factors of all heads must be precisely matched throughout rotation.
  • 20. SPECT Performance Lecture 14: Emission Tomography III  Spatial resolution  X- and Y-magnification factors and multi-energy spatial registration  Alignment of projection images to axis-of-rotation  Uniformity  Camera head tilt
  • 21. SPECT Spatial resolution Lecture 14: Emission Tomography III  Can be measured by acquiring a SPECT study of a line source (capillary tube filled with a solution of Tc-99m, placed parallel to axis of rotation).  FWHM of the line sources are determined from the reconstructed transverse images (ramp filter).
  • 22. SPECT Spatial resolution Lecture 14: Emission Tomography III  National Electrical Manufacturers Association (NEMA) specifies a cylindrical plastic water-filled phantom, 22 cm in diameter, containing 3 line sources
  • 23. SPECT Spatial resolution Lecture 14: Emission Tomography III  NEMA spatial resolution measurements are primarily determined by the collimator used.  Tangential resolution 7 to 8 mm FWHM for LEHR  central resolution 9.5 to 12 mm  radial resolution 9.4 to 12 mm
  • 24. SPECT Spatial resolution Lecture 14: Emission Tomography III  NEMA measurements not necessarily representative of clinical performance  Studies can be acquired using longer imaging times and closer orbits than would be possible in a patient.
  • 25. SPECT Spatial resolution Lecture 14: Emission Tomography III  NEMA measurements not necessarily representative of clinical performance  Studies can be acquired using longer imaging times and closer orbits than would be possible in a patient.
  • 26. SPECT Spatial resolution Lecture 14: Emission Tomography III  NEMA measurements not necessarily representative of clinical performance  Studies can be acquired using longer imaging times and closer orbits than would be possible in a patient.  Filters used for clinical studies have lower spatial frequency cutoffs than the ramp filters used in NEMA measurements.
  • 27. Comparison with conventional planar scintillation camera imaging Lecture 14: Emission Tomography III  In theory, SPECT should produce spatial resolution similar to that of planar scintillation camera imaging.  In clinical imaging, its resolution is usually slightly worse.  Camera head is closer to patient in conventional planar imaging than in SPECT.
  • 28. Comparison with conventional planar scintillation camera imaging Lecture 14: Emission Tomography III  Short time per view of SPECT may mandate use of lower resolution collimator to obtain adequate number of counts.  In planar imaging, radioactivity in tissues in front of and behind an organ of interest causes a reduction in contrast.
  • 29. Comparison with conventional planar scintillation camera imaging Lecture 14: Emission Tomography III  Main advantage of SPECT is markedly improved contrast and reduced structural noise produced by eliminating the activity in overlapping structures.  SPECT also offers promise of partial correction for effects of attenuation and scattering of photons in the patient
  • 30. Magnification factors Lecture 14: Emission Tomography III The X- and Y-magnification factors, often called X and Y gains, related distances in the object being imaged, in the x and y directions, to the numbers of pixels between the corresponding points in the resultant image.
  • 31. Magnification factors Lecture 14: Emission Tomography III  Magnification factors determined from a digital image of two point sources placed against the camera’s collimator If X- and Y-magnification factors are unequal, the projection images will be distorted in shape, as will coronal, sagittal, and oblique images.
  • 32. COR calibration Lecture 14: Emission Tomography III  The axis of rotation (AOR) is an imaginary reference line about which the head or heads of a SPECT camera rotate.  If a radioactive line source were placed on the AOR, each projection image would depict a vertical straight line near the center of the image.
  • 33. COR calibration Lecture 14: Emission Tomography III  This projection of the AOR into the image is called the center of rotation (COR).  Ideally, the COR is aligned with the center, in the x-direction, of each projection image.
  • 34. COR calibration Lecture 14: Emission Tomography III  Misalignment may be mechanical or electronic.  Camera head may not be exactly centered in the gantry.
  • 35. COR calibration Lecture 14: Emission Tomography III COR Degradation and Sinogram
  • 36. COR calibration Lecture 14: Emission Tomography III  COR misalignment causes a loss of spatial resolution in the resultant transverse images.  Large misalignment cause a point source to appear as “doughnut”.  Doughnut are not centered in the image so can be distinguished from “ring” artifacts produced by non-uniformities.
  • 37. COR calibration Lecture 14: Emission Tomography III  COR alignment is assessed by placing a point source or line source in the camera field of view.  Projected imaged and or sinogram is analyzed by the camera’s computer.
  • 38. COR calibration Lecture 14: Emission Tomography III  Misalignment may be corrected by shifting each image in the x- direction by the proper number of pixels prior to filtered back- projection  If COR misalignment varies with camera head angle, it can only be corrected if computer permits angle-by-angle corrections.
  • 39. Uniformity Lecture 14: Emission Tomography III  Nonuniformities that are not apparent in low-count daily uniformity studies can cause significant artifacts in SPECT.  Artifact appears in transverse images as a ring centered about the AOR.
  • 40. Uniformity Lecture 14: Emission Tomography III Cylinder filled with a uniform radionuclide solution showing a ring artifact due to non- uniformity.
  • 41. Uniformity Lecture 14: Emission Tomography III  Primary intrinsic causes of non-uniformity are, a. Spatial non-linearities  stretch the image in some areas  reducing the local count density  compress other areas of the images  Increasing the count density a. Local variation in the light collection efficiency
  • 42. Uniformity Lecture 14: Emission Tomography III  Lookup table can not correct,  Local variations in detection efficiency such as dents or manufacturing defects in the collimators.
  • 43. Uniformity Lecture 14: Emission Tomography III  High-count uniformity images used to determine pixel correction factors,  At least 30 million counts for 64 x 64 images  At least 120 million counts for 128 x 128 images  Collected every 1 or 2 weeks; separate images for each camera head
  • 44. Camera head tilt Lecture 14: Emission Tomography III  Camera head or heads must be exactly parallel to the AOR.