SlideShare une entreprise Scribd logo
1  sur  113
Calibration and Quality controls of
Automated Hematology Analyzer
Dr. Pranav N Shirbhate..
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
 VIVA Ques at the End
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
HISTORY
Antonie van
Leeuwenhoek
Invented Microscope
Wallace Coulter :First
automated analyzer for
counting and sizing cells
and presented it in 1956
The world of cells was
colorless until Paul
Ehrlich stained blood cells
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
AUTOMATIONS IN HEMATOLOGY
Cell counts (Automated hematologyanalyzers)
Diagnosis of hemoglobinopathies (HPLC)
Immunophenotyping (Flow cytometry)
Coagulation (Coagulometers)
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
 Hemogram is a Backbone of any lab evaluation.
 The Core Laboratory report well over 140,000 CBCs and
48,000 peripheral smears per year at our Dept of Pathology, MGIMS Sevagram.
 It is a routine investigation for-
Anaemia,Polycythemia, Infection, Inflammation, allergy, malignancy,
bleeding tendency etc
All automated cell counters are screening devices. Abnormalities must be verified
by a blood film, staining and scanning by an expert observer.
But testing each and every sample out of theses 1.4 lac samples a year by manual
method is very laborious, cumbersome, so comes the need of automation.
NEED OF AUTOMATION
AUTOMATED TECHNIQUES OF BLOOD COUNTING
• Semi-automated instruments
- Require some steps, as dilution of blood samples .
- Often measure only a small number of variables
• Fully automated instruments
- Require only that an appropriate blood sample is presented to the instrument.
- They can measure 8-20 variables including some new parameters which do not
have any equivalent in manual methods.
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
INTRODUCTION
• Quality Assuarance (QA) is the sum of all those activities in which the laboratory is
engaged to ensure that information generated by laboratory is correct.
• QA is not restricted to the development and retention of quality control charts but
rather includes all aspects of laboratory activities that affects the results produced,
from the choice of methods, to the education of personnel, to the handling of
specimens and reporting results.
• The real purpose of QA activities is to determine how correct or incorrect the results
emanating from the lab are, and to allow those managing the lab to determine whether
or not the lab is fulfilling its functions satisfactorily.
INTRODUCTION
3 major activities of QA :
1 ) Preventive –
2 ) Assesment –
3 ) Corrective
1 ) Preventive –
those activities that are done prior to the examination of the specimen or sample and that
are intended to establish systems conducive to accuracy in analytical testing
 Preventive maintenance and calibration of instruments,
 Testing of media,
 Orientation and training of personnel
INTRODUCTION
3 major activities of QA :
2 ) Assessment –
those activities that are done during testing to determine whether the test systems are
performing correctly
 The use of standard and controls,
 Maintenance of control charts
3 ) Corrective – those activities that are done, when error is detected, to correct the
system
 equipment troubleshooting,
 recalibration of instruments
INTRODUCTION
QA in Hematology Laboratory
QA in Hematology lab is intended to ensure the reliability of the lab tests. The objective is
to achieve precision and accuracy;
So components of QA programme :
 Standardization/ Calibration
 Internal Quality Control ( IQC )
 External Quality Control ( EQC )/ Proficiency testing
DEFINITIONS IN TOTAL QUALITY MANAGEMENT
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
QC & CALIBRATION TERMINOLOGIES
Accuracy, I) A measure of agreement between the estimate of a value and a "true"
value; quantifiable in terms of departure from accuracy, expressed as
systematic error or bias;
II) Accuracy of an analytic process, n - expressed as the difference between
the average result obtainable by a method under specified conditions
and the result accepted as true or standard.
III) Accuracy of a result or control result, - Expressed as the difference
between a result and the "true" value.
QC & CALIBRATION TERMINOLOGIES
Bias/ systematic error –
• Systematic, signed deviation of the test result from the accepted reference value.
• (Defined as the difference between the expectation of the test results and an
accepted reference value.)
Calibration, n –
• The process of testing and adjustment of an instrument, kit, or test system, to
provide a known relationship between the measurement response and the value of
the substance being measured by the test procedure.
QC & CALIBRATION TERMINOLOGIES
Calibrator, n - A (reference) material (e.g., solution, suspension) or
device of known quantitative/qualitative characteristics (e.g.,
concentration, activity, intensity, reactivity) used to calibrate,
graduate, or adjust a measurement procedure or to compare the
response obtained with the response of a test specimen/sample.
NOTE: The quantities of the analytes of interest in the calibrator are
known within limits ascertained during its preparation and are used
to establish the relationship of an analytical method's response to
the characteristic measured for all methods or restricted to some.
The calibrator must be traceable to a national or international
reference preparation or reference material when these are
available. (like eg NCCLS/CLSI)
ORGANIZATIONS INVOLVED IN STANDARDIZATION AND QUALITY
MANAGEMENT SYSTEMS
ORGANIZATIONS INVOLVED IN STANDARDIZATION AND QUALITY
MANAGEMENT SYSTEMS
WHAT THESE ORGANIZATIONS ARE? FOR e.g. CLSI
CLSI/ NCCLS
The Clinical and Laboratory Standards Institute (CLSI) is a
volunteer-driven, membership-supported, not-for-
profit, standards development organization. CLSI promotes
the development and use of voluntary laboratory consensus
standards and guidelines within the health care community.
The organization’s mission is to develop clinical and
laboratory practices and promote their use worldwide.
The CLSI started as the National Committee on Clinical
Laboratory Standards (NCCLS) and changed to its current
name (CLSI) at the beginning of the year 2005.
NCCLS/ CLSI STANDARDS AND GUIDELINES
CLSI publishes voluntary consensus standards and guidelines, reports, and a number of companion
products.
Standards are documents developed through the consensus process that clearly identify specific,
essential requirements for materials, methods, or practices for use in an unmodified form. Standards
may, in addition, contain discretionary elements, which are clearly identified.
Guidelines are documents developed through the consensus process describing criteria for a general
operating practice, procedure, or material for voluntary use.
Guidelines may be used as written or modified by the user to fit specific needs.
Reports are documents that have not been subjected to consensus review and are released by the
appropriate consensus committee.
Companion products include any item provided or sold that is intended to be used in conjunction
with a CLSI standard or guideline. Examples include, but are not limited to, Quick Guides, Wall
Charts, software, and templates.
QC & CALIBRATION TERMINOLOGIES
(Quality) control material, n - A device, solution, or lyophilized
preparation intended for use in the quality
control process;
NOTE: It should be similar to, and is analyzed along with, patient
specimens. If different, it should have a recognized, defined response
to analytic measurements.
Control materials
may or may not have known analyte concentrations (i.e., assigned
values) within specified limits (e.g. target value ± standard deviation).
Control materials are not used for calibration purposes.
QC & CALIBRATION TERMINOLOGIES
Derived red cell indices, n - Quantities (see MCH, MCHC, MCV) that may be calculated from the
measurement of hemoglobin (Hb) concentration, packed (red) cell volume (PCV), and erythrocyte
(RBC) concentration.
QC & CALIBRATION TERMINOLOGIES
Linearity, n - The ability (within a given range) to provide results that are directly proportional to
the concentration {amount} of the analyte in the test sample.
QC & CALIBRATION TERMINOLOGIES
Precision, n - The closeness of agreement between
independent test results obtained under prescribed
{stipulated} conditions;
NOTE: Precision is not typically represented as a
numerical value but is expressed quantitatively in terms
of imprecision—the SD or the CV of the results in a set of
replicate Measurements.
QC & CALIBRATION TERMINOLOGIES
Precision, n - The closeness of agreement between
independent test results obtained under prescribed
{stipulated} conditions;
NOTE: Precision is not typically represented as a
numerical value but is expressed quantitatively in terms
of imprecision—the SD or the CV of the results in a set of
replicate Measurements.
For practical purpose
• Accuracy: Refers to closeness to the true value
• Precision: Refers to reproducibility of test
QC & CALIBRATION TERMINOLOGIES
Drift, -
QC & CALIBRATION TERMINOLOGIES
Reference material, RM, n –
A material or substance, one or more of whose property values are sufficiently homogeneous
and well established to be used for the calibration of an apparatus, the assessment of a
measurement method, or for assigning values to materials;
Certified reference material, CRM, n - A reference material that has one or more values
certified by a technically valid procedure and is accompanied by, or is traceable to, a
certificate or other document that is issued by a certifying body;
NOTES: a) defines a CRM as a “reference material, accompanied by a certificate,
one or more of whose certified value is accompanied by an uncertainty at a stated level of
confidence”;
b) The term "Standard Reference Material (SRM)" is the name of a CRM, which is
the trademark name of a certified reference material that has been certified and is distributed by
the National Institute of Standards and Technology (NIST), a U.S. Government agency formerly
known as the National Bureau of Standards (NBS).
QC & CALIBRATION TERMINOLOGIES
Sensitivity – It is the smallest quantity of analyte that can be reproducibly distinguished from
background noise in a given assay system.
Specificity - For quantitative tests, the ability of an analytical method to determine only the
component it purports to measure or the extent to which the assay responds only to all
subsets of a specified analyte and not to other substances present in the sample.
QC & CALIBRATION TERMINOLOGIES
Set point - The level (i.e., analyte concentration) at which an analyzer is calibrated.
Standard (measurement), n - An authoritative “document” setting forth criteria for
performance and characteristics.
• Standard, primary, n - A standard that is designated or widely acknowledged as having
the highest metro logical qualities and whose value is accepted without reference to other
standards of the same quantity.
• Standard, secondary, n - A standard whose value is assigned by comparison with a
primary standard of the same quantity.
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
GENERAL REQUIREMENTS FOR QC & CALIBRATION
1) Quality Control Manual
• All quality control procedures should be described in detail in a quality manual. Clinical
Laboratory Technical Procedure Manuals for more details.
• The quality manual should also include a general statement of the laboratory quality policy
and an acknowledgment that the policy is understood by all laboratory workers.
• The persons with authority to design and modify all laboratory procedures and the persons
with authority to approve such designs and changes should be listed.
2) Record Keeping
• Comparisons with previous performance should be provided before analyzing control
materials.
• Electronic verification and background measurement recorded.
• CBC/differential files should include control set-up, control run, and control review/report
files.
GENERAL REQUIREMENTS FOR QC & CALIBRATION
3) Retention of Records
There are three major files of information. Storage of this data base is usually made HIS.
• Laboratory order entry: The laboratory order entry data is maintained for a user-defined
number of days and is then purged.
• Workload data: Workload data is kept in active files for six months and is then archived.
These archives are updated every month and stored in a safe place with timely access to
permit review as the need arises.
• Laboratory data files: The duration of QC record storage depends on local hospital or clinic
policies or accreditation requirements, or should be a minimum of two years or more,
whichever is greater.
Proficiency testing results, noncommercial controls, or calibrators, if used,
must have complete documentation of standardization and performance characteristics.
GENERAL REQUIREMENTS FOR QC & CALIBRATION
4) Training and Qualifications
• Laboratory workers who are assigned to carry out QC procedures must have been trained
and records must indicate this.
• As per CSLS guidelines, personnel who are given responsibility for interpreting QC results
should not be the same personnel doing the routine work.
5) Environment
Maintain environment that reflects routine working conditions, ambient temperature.
A record of temperature be maintained.
GENERAL REQUIREMENTS FOR QC & CALIBRATION
6) Analyzer Identification labels
GENERAL REQUIREMENTS FOR QC & CALIBRATION
7) Preanalytic Variables
Uncontrolled preanalytic variables can exert a misleading effect , therefore a
checklist of possible error sources will help ensure that they will not be overlooked.
Analyte Instability
• Least likely to be seen in stabilized blood preparations and most likely to be seen in recently
collected blood specimens.
• Over-vigorous mixing may also play a role.
• Viral disease may cause reduced lymphocyte count due to cell fragility. Cold agglutination
may affect platelets as well as red cells.
(Detection of analyte instability depends on the demonstration of a difference between the mean
value of the first half of an iterative assay sequence and the mean value of the second half. To be
significant the difference between means should exceed standard error of the first mean at 95%
confidence.)
GENERAL REQUIREMENTS FOR QC & CALIBRATION
Analyte Nonuniformity
Caused by inadequate mixing. Fresh blood is less susceptible to this.
Mixing a specimen by rapid
shaking or vortexing or any method that causes frothing has be avoided.
Two components in nonuniformity.
1. nonuniform dispersion of the total cell mass throughout the plasma.
2. nonuniform distribution of the different cell types with respect to each other.
To achieve uniformity of a whole blood suspension, - 10 complete
inversions of the specimen container.
(A completely filled container gives poor mixing; an air bubble must be present. If the
specimen has been stored for 2 to 5 minutes, three additional inversions should precede the
assay.)
GENERAL REQUIREMENTS FOR QC & CALIBRATION
Autoagglutination
• Autoagglutination is a cause of unstable suspensions.
• If not visible to the eye, autoagglutination or rouleaux should be suspected if
1) reduced RBC together with elevated MCHC are encountered;
2) The MCV may be slightly to markedly increased;
3) The rule of thumb “3 x Hb (g/dL) = Hct (%) ±3" is not met.
• Its presence can usually be confirmed by microscopy, but be aware that a warm slide may
cause autoagglutinates to dissociate.
• Specimens with autoagglutination should not be used in quality control procedures.
• Refrigerated specimens should be brought to room temperature before analysis.
GENERAL REQUIREMENTS FOR QC & CALIBRATION
Anticoagulant Effect
• Insufficient blood collected into an evacuated container may cause shrinkage of all cell
types, but red cells usually become restored to the size dictated by their osmotic
equilibrium with the cell-counting diluent.
• Leukocytes, erythrocytes, and platelets (e.g., EDTA-induced aggregation) may be irreversibly
changed and cause errors in the differential count and platelet count and MCV.
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
CALIBRATION
• Calibration fine tunes your Hematology analyzer and provides the
most accurate results possible.
• In the normal process of tracking data for an extended period of time,
your laboratory can make a specific decision to recalibrate a given
parameter. Never adjust to a specific value for an individual sample.
• For best performance, calibrate all the CBC parameters.
The WBC differential is calibrated at the factory. They do not require
calibration in the laboratory.
CALIBRATION
We should calibrate our instrument:
*At installation
*After the replacement of any component that involves dilution
characteristics or the primary measurements (such as the apertures)
*When advised to do so by your service representative
BASIC OF CALIBRATION OF NEW/REPAIRED INSTRUMENT
Establish Baseline Parameters
• The analytical baseline data provide reference levels for judging the performance of the
analyzer.
• Baselines should be established or verified when a new analyzer is installed and when a
modification or repair is made that could significantly affect the analyzer performance.
• The product reference manual should provide instruction for frequency of recalibration
under normal operating conditions.
• Calibration or recalibration must be preceded by verification of the major analytical
variables, which are general conditions of imprecision and accuracy and the analyzer.
• Reportable range should be verified for a new installation and after major repair.
• To avoid the possibility of analyzer defect causing errors in the baseline procedures,
preventive maintenance should be performed shortly before calibration.
ESTABLISH BASELINE PARAMETERS
Imprecision Baseline
• Baseline imprecision is described by the standard deviation of replicate assays of the same
specimen.
Measure Baseline Imprecision
• Make n consecutive assays of the same well-mixed stabilized blood. The preferred number
of iterations is n=31.
• Fewer repeated assays (for example, n = 10) may be used, but if this is done, convert
standard deviation (SD) to confidence limits (CL) as Evaluate Baseline Imprecision.
ESTABLISH BASELINE PARAMETERS
• Record the results of each analyte
assay in a table.
• Note the need to record the
elapsed time of the measurement.
ESTABLISH BASELINE PARAMETERS
ESTABLISH BASELINE PARAMETERS
Evaluate Baseline Imprecision
• Imprecision may be stated as SD or CL . (95% confidence limit).
• If comparison is to be made 95% between standard deviations from different sources, e.g.,
between the laboratory estimate and a manufacturer's labeled value, be sure that each
measurement has been made with the same number of iterations (n).
ESTABLISH BASELINE PARAMETERS
Evaluate Baseline Imprecision
• If different values of n have been used, convert SD
to CL95% as follows:
• Multiply each SD by the value of t for (n-1) given in
this Appendix.
For e.g.,
• the laboratory used n = 31, but
the manufacturer's labeled SD is based on n = 16.
• Therefore, multiply the laboratory SD by 2.0423 and
the manufacturer's SD by 2.1315.
• This will convert both SD's to CL , making it possible
to compare them.
ESTABLISH BASELINE PARAMETERS
Evaluate Baseline Imprecision
• If different values of n have been used, convert SD
to CL95% as follows:
• Multiply each SD by the value of t for (n-1) given in
this Appendix.
For e.g.,
• the laboratory used n = 31, but
the manufacturer's labeled SD is based on n = 16.
• Therefore, multiply the laboratory SD by 2.0423 and
the manufacturer's SD by 2.1315.
• This will convert both SD's to CL , making it possible
to compare them.
ESTABLISH BASELINE PARAMETERS
Accuracy Baseline
• The accuracy baseline is defined as the starting point against which future bias control
information will be compared.
• Verification of accuracy must be made immediately (less than 15 minutes) after calibration or
as directed by the manufacturer. If delayed, the data from the baseline verification process may
be contaminated by drift or other analyzer error.
• The term "calibration" means adjusting the analyzer to achieve zero bias, or verifying through a
similar process that zero bias already exists.
• The term “zero bias” means that any difference between the calibrator-assigned values and the
results of calibrator assay is due only to analytical imprecision.
• It is assumed that all automated hematology analyzers have a linear response over the
reportable range, that the limits of linearity are specified, and that zero analyte concentration
produces a zero analyzer response. These characteristics make it possible to calibrate analyzers
by a single-point method.
ESTABLISH BASELINE PARAMETERS
Reference-Assayed Whole Blood Calibration
Rarely used today in medical institutions. Not recommended.
Standardized Stabilized Blood Calibration Method
Frequency of calibration: - As per product reference manual.
- Usually calibration is recommended after a major repair or after a
significant change of base line accuracy (drift).
- In the latter case, the cause of excessive drift should be corrected
before proceeding with calibration.
Iteration of calibrator assays: The number of assays will influence the precision of the calibration
factor.
Greater assay iteration will result in a lower SEM.
ESTABLISH BASELINE PARAMETERS
• More replication will yield a more precise calibration set point and more accurate subsequent
patient assays but, because of the square root function, the rewards of increasing replication will
reduced.
• Furthermore, excessive mixing associated with increased replication may damage
cells and introduce other errors.
• It is suggested that 10 consecutive assays of the calibrator will give a satisfactory result.
For example,
the SD for WBC was found to be 0.2 x 109/L. The SEM of ten
calibrator assays will then be 0.2 ÷ /10 or 0.2 ÷ 3.16 = ±0.063 x 109/L.
This error, in combination with the analyzer SD will be propagated (included in) to patient assays. To
obtain the 95% assay bias range, use the root sum square variance method as..
RECORDING CALIBRATION RESULTS
ESTABLISH BASELINE PARAMETERS
Reportable Range
• Reportable range is the range of assay values over which accuracy is ensured; hence its
reportable to the patient’s record. (accuracy of values beyond reportable range is not ensured).
• It describes only the effect of analyte concentration and does not include possible error due to
interferences, such as agglutination or hemolysis.
• Upper and lower limits of reportable range for a newly installed analyzer can be verified by
recovery of the assigned values of abnormal high and abnormal low stabilized blood controls.
• Assays of sequential dilutions of blood between these limits do not contribute information that
is helpful in establishing extension of this reportable range.
• Reassurance of midrange linearity is given by the recovery of the assigned value of the normal
stabilized blood control and by the verification of the calibration set point.
For e.g., this happens a lot of time in leukemia cases where WBC values go beyond reportable
range. Analyzer counts the cells but certainty about the accuracy can be questioned.
CALIBRATION COMPLETION
Technician from authorized / standardized calibration party Runs the blacks cycle to test values
within range.
CALIBRATION COMPLETION
• He feeds all these (SD, CV, SEM and Base
line parameters) values to analyzer
software, and helps its synchronization
with HIS/ Hospital software.
• Calibration of the analyzer completes only
when we receive calibration certificate
clearly mentioning the CRM body like
CSLS/NCCLS.
CALIBRATION COMPLETION
Now our machine/ analyzer is ready to be used for routine runs.
DAILY MAINTENANCE
• Daily cleaning
• Electronic checks
• Compare open and closed mode sampling (using a normal patient sample)
• Run controls
ENSURE THE INSTRUMENT IS FUNCTIONING PROPERLY
• By Checking the reagent containers for:
o Sufficient quantity
o Not beyond expiration date
o No precipitates, turbidity, particulate matter, or unusual colour
o Proper connections between the instrument and the reagent containers
• Check the waste container for: *Sufficient capacity and * Proper connections
• Perform daily startup and Background counts
• In addition to verifying daily startup results, verify acceptable:
o Reproducibility
o Carryover
o Control Results
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
QUALITY CONTROL
Internal Quality Control –
The set of procedures undertaken in a laboratory for the continuous
assessment of work carried out in the laboratory and evaluation of tests to decide whether these
are reliable enough for their results to be released to the requesting physician.
NOTE: The main object is to ensure day-to-day consistency of measurements or observations, if
possible in agreement with an agreed-on indicator of truth, such as a control material with assigned
values.
QUALITY CONTROL
Internal Quality Control
• Once a performance baseline has been established, its control is based on methods that will
detect loss of accuracy or loss of precision, or both.
• Control methods must be sensitive enough to reveal loss of performance that could compromise
patient assay values, but they cannot be oversensitive to the point of signaling error when none
exists.
• As a minimum, a performance control system must provide the following information:
 Analytical imprecision has not deteriorated with time or use.
 Bias has remained within the limits established at the time of calibration.
INTERNAL QUALITY CONTROL
• The term “internal" indicates quality control procedures that take place within the laboratory
immediately before, during, or immediately after an analysis run.
• The procedures must be carried out as an integral part of patient assays.
• Control assays performed only once daily may not provide enough information to ensure
accuracy and consistency of results.
• When the analyzer is restarted after complete shut-down, precision and set point should be
verified to ensure that satisfactory analytical capability has not been lost.
• Internal quality control is distinguished from external quality control. The latter is administered
by a third party, such as an authorized proficiency testing body, based on blind assay of a
stabilized control material, or by a manufacturer offering statistical peer comparison of routine
stabilized blood control results.
• Analyzers of differing designs and operating principles will have differing propensities for failure
or for performance deterioration.
INTERNAL QUALITY CONTROL
INTERNAL QUALITY CONTROL
Quality Control Using Stabilized Blood
• Stabilized blood controls should be similar to calibrators,
except that assigned values, if provided, may be conferred by
calibrated analyzers using stabilized blood calibrators instead
of reference-assayed whole blood.
• Manufacturers' assigned values pose a temptation to use
controls as calibrators and to "unblind" some types of
internal and external performance reporting.
• Controls should have 2-3 levels of analyte concentration.
These levels should reflect performance at normal analyte
levels and at, or
close to, the low limits of the reportable range.
INTERNAL QUALITY CONTROL
Quality Control Using Stabilized Blood
• Stabilized blood controls must be labeled for use
with specified analyzers.
• The laboratory should document that the product
it uses is so labeled.
• We are using Bio-Rad 3 level of control LOW/ NORMAL/
HIGH.
INTERNAL QUALITY CONTROL
Frequency of Use
• We have to decide how many assays should separate controls.
This decision should be made on the basis of work load and
experience with the drift characteristics of the analyzer(s).
• If the results of daily control assays are within laboratory-
defined limits, such as ±2 SD of the assigned or conferred
value, it can be assumed that the results of patient assays
performed within a control-bracketed work segment are valid
and reportable.
• The fewer the control assays relative to patient specimens, the
greater will be the number of assumed invalid results in the
event of an out-of-control condition.
INTERNAL QUALITY CONTROL
Frequency of Use
• As a minimum, assays of normal and abnormal levels of
control
material should be made at the beginning of each run and one
assay of normal level control made at the end of the run to
ensure that patient results have been satisfactory.
We at our institution running these three control samples
at the start of the new day of analyzer.
INTERNAL QUALITY CONTROL
Charts are prepared with the values shown by analyzer for that day.
INTERNAL QUALITY CONTROL
Test Replication / Precision on patients sample
• The mean of replicate assays of control material is more
informative than a single measurement.
• Paired assays are recommended. Not only does the mean of
the pair give a more exact value, but the difference between
members of a pair also gives an estimate of imprecision.
INTERNAL QUALITY CONTROL
Test Replication / Precision on patients sample
Overlap is a special case of test replication. When a new lot of control
material is received, it should be tested in parallel with the currently used
lot for 4 days.
This will establish the ratio of the assigned values of the incoming
lot to the assigned values of the currently used material.
Differences in assigned values of the two lots should not be
taken as an indication to recalibrate.
We are doing precision bi-weekly ie on Tuesday and Friday at start of new
day for the analyzer with random fresh EDTA sample. (Overlap - not done)
INTERNAL QUALITY CONTROL
Correlation Check
Correlation check implies that any unexpected result of a test must be checked
to see whether it can be explained on clinical ground or whether it correlates
with other tests.
E.g. 1) Unexpectedly higher Hb needs explanation of blood transfusion or by
hemorrhage respectively.
2) Low MCHC confirmed by demonstrating hypochromic red cells on
Romanowsky stained blood film
3) High MCV must correlate with macrocytosis.
4)Marked leukocytosis, thrombocytopenia and thrombocytosis, needs
blood film examination
We do correlation check for each and every sample we receive in our laboratory.
INTERNAL QUALITY CONTROL
Checking of aberrant result
Recording blood count data on cumulative report for or charts is a good clinical
practice and provides an inbuilt quality control system by making it easy to
detect any aberrant result when compared with previously determined base
line.
E.g. Useful in detecting the occasional wild error caused by
1) incorrect labelling of specimen,
2) Inadequate suspension of blood before sampling,
3) Partial clotting of a blood sample,
4) Deterioration on storage.
We have HIS (Hospital Information System), which is helpful in providing previous
(cumulative) as well as other investigation done by patients which are very handy,
while maintaining internal quality of reporting.
INTERNAL QUALITY CONTROL
Quality Control Methodology
• The basis of the method is continued conformance of
control assay values to the assigned values of the stabilized
blood product. These values may be manufacturer-assigned
or assigned by the laboratory, using an analyzer that has
satisfied the requirements of base line accuracy.
• Recovery of assigned values within the confidence limits of
the analytical process, combined with the confidence limits
of the assigned values, will provide assurance that analyzer
accuracy and precision are acceptable.
INTERNAL QUALITY CONTROL
Quality Control Methodology
• Any disagreement that exceeds these expected range, ie
2SD, limits calls for further investigation to determine
whether it is a random event, loss of precision, loss of
accuracy, or deterioration of the control material.
• It is recommended that result recovery ranges be
expressed as ±2
SD, or preferably 95% of CL of the mean value.
INTERNAL QUALITY CONTROL
Interpretation of Results
An interpretive method for out-of-limits control result should be able to distinguish among the
following:
 Random event,
 Loss of precision, and
 Loss of accuracy.
INTERNAL QUALITY CONTROL
Random event versus imprecision:
• With control boundaries defined as ±2 SD, one control value
in exceed these limits as a random event.
• Exclude random event by two further assays of the same
type of control (normal, high, or low) that gave the out-of-
limit result.
• If the difference between these assay results is less than the
established ±2 SD limits, the out-of-limit control result was
probably a random event and requires no further action.
INTERNAL QUALITY CONTROL
Random event versus imprecision:
• If the difference is greater than ±2 SD, assume precision loss
and confirm this by two further assays.
• At least two of the total of five results will exceed ±2 SD limits
if precision has deteriorated.
• In this case, patients’ results that preceded the out-of-limit
condition should be reviewed in order to determine
reportability.
INTERNAL QUALITY CONTROL
Loss of accuracy:
• An out-of-limits control result due to loss of accuracy (bias change) can
be verified by the repeated assay method given above. If inaccuracy has
developed, the mean of the replicate results will differ by at least 1 SD
from the mean of previous control results.
• If the bias change affects only one of the three control levels, suspect a
deterioration of the control material.
• If the bias change affects the red cell indices, inspect the weighted
moving average values for supporting evidence of change in RBC, Hb, or
MCV.
• Several consecutive similar results on one side of the mean = calibration
fault causing consistent bias.
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
EXTERNAL QUALITY CONTROL (EQA) / PROFICIENCY TESTING
EXTERNAL QUALITY CONTROL (EQA) / PROFICIENCY TESTING
• EQA complements internal quality control and is a basic requirement
for clinical laboratory certification and accreditation.
• Certification is a procedure by which a third party gives written
assurance that a product, process or service conforms to specific
requirements.
• Accreditation is a procedure by which an authoritative body (NABL)
gives formal recognition that a body or person is competent to carry
out specific tasks.
EXTERNAL QUALITY CONTROL (EQA)
EXTERNAL QUALITY CONTROL (EQA)
• EQA has following important principles and criteria.
EXTERNAL QUALITY CONTROL (EQA)
• EQA has following important principles and criteria.
EXTERNAL QUALITY CONTROL (EQA)
EXTERNAL QUALITY CONTROL (EQA)
EXTERNAL QUALITY CONTROL (EQA)
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
LEVEY-JENNINGS CHART
SHEWHART OR LEVEY-JENNINGS CHART ie Internal Quality Control
• Dates of analysis plotted along
the X-axis and control values
plotted on Y- axis.
• Mean and one, two and three
standard deviation limits are
also marked on Y-axis.
• Distance from the mean is
measured in standard
deviations.
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
WESTGARD RULES TO BE FOLLOWED
13s rule
13s refers to a control rule that is commonly
used with a Levey-Jennings chart when the
control limits are set as the mean plus 3s and
the mean minus 3s.
A run is rejected when a single control
measurement exceeds the mean plus 3s or
the mean minus 3s control limit.
WESTGARD RULES TO BE FOLLOWED
12s rule
12s refers to the control rule that is commonly used
with a LJ chart when the control limits are set as the
mean plus/minus 2s.
In the original Westgard multirule QC procedure,
this rule is used as a warning rule to trigger careful
inspection of the control data by the following
rejection rules.
WESTGARD RULES TO BE FOLLOWED
22s rule
22s - reject when 2 consecutive control
measurements exceed the same mean plus 2s or
the same mean minus 2s control limit.
Warning
WESTGARD RULES TO BE FOLLOWED
R4s rule
R4s - reject when 1 control measurement in a group
exceeds the mean plus 2s and another exceeds the
mean minus 2s.
This rule should only be interpreted within-run, not
between-run. The graphic should really imply that
points 5 and 6 are within the same run.
WESTGARD RULES TO BE FOLLOWED
41s rule
41s - reject when 4 consecutive control
measurements exceed the same mean plus 1s
or the same mean minus 1s control limit.
WESTGARD RULES TO BE FOLLOWED
10x rule
10x - reject when 10 consecutive control
measurements fall on one side of the mean.
Note – There are plenty more Westgard’s rules. Can be
accessed at https://www.westgard.com/mltirule.htm
CONTENTS
 History
 Automations in Hematology
 Need of automation
 Introduction to Quality assurance
 QC & calibration terminologies
 Requirements for QC and calibration
 Calibration of New/ Repaired automated Hematology Analyzer
 Quality control in Hematology – Internal and External
 Levey–Jennings (LJ) chart
 Application of Westgard’s Rules
 NABL
NABL
National Accreditation Board for Testing & Calibration
Laboratories (NABL) is an autonomous society providing Accreditation
(Recognition) of Technical competence of a testing, calibration, medical
laboratory & Proficiency testing provider (PTP) & Reference Material
Producer (RMP) for a specific scope following
IEC/ISO17025:2005,
ISO 15189:2012,
IEC/ISO 17043:2010 &
IEC/ISO 17034:2016 Standards
It has Mutual Recognition Arrangements with
Asia Pacific Laboratory Accreditation Cooperation (APLAC),
Mutual Recognition Arrangement (MRA),
International Laboratory Accreditation Cooperation (ILAC).
NABL
• NABL provides accreditation in all major fields of Science and
Engineering such as Biological, Chemical, Electrical, Electronics,
Mechanical, Fluid-Flow and many more.
• NABL also provides accreditation for medical testing laboratories.
• In addition, NABL also offers accreditation for Proficiency testing
providers & Reference Material producers and is now signatory to
APLAC MRA for both.
NABL
Benefits of accreditation
• NABL is a Full member (ILAC MRA signatory) to ILAC as well as APLAC
Mutual Recognition Arrangements (MRA) ,based on mutual evaluation
and acceptance of other MRA Partner laboratory accreditation systems.
• This facilitate acceptance of test / calibration results between countries.
NABL
Benefits of accreditation
• This developing system of international mutual recognition agreements
between accreditation bodies has enabled accredited laboratories to
achieve a form of international recognition.
• In medical field NABL accredation works as VISA for the any report.
• Any report report reported in NABL accredited lab, makes it valid and
gives confidence to overseas treating physician/surgeon about reports
being standardized.
SO, ………………………. TO SUM-UP
WITHOUT UNDERSTANDING CALIBRATION AND QUALITY CONTROL
TAKE HOME MESSAGE
Automation in Hematology is like Vitamin
tablets.
It is a supplement and not
the substitute to Manual methods.
REGARDS..!!
REFERENCES
1. Standard P. Calibration and Quality Control of Automated Hematology Analyzers; Proposed Standard NCCLS.
Vol. 19. 2008.
2. Saiket M. World of Automation. In: Automated cell counter and quality control. 2014. p. 1–24.
3. Novak E, Guenther B. Automation in. In: Automation in Hematology. 2014. p. 369–94.
4. Anon, (2018). WestgardQC. [online] Available at: https://www.westgard.com/mltirule.htm [Accessed 2 Mar.
2018].
5. En.wikipedia.org. (2018). National Accreditation Board for Testing and Calibration Laboratories. [online]
Available at:
https://en.wikipedia.org/wiki/National_Accreditation_Board_for_Testing_and_Calibration_Laboratories
[Accessed 2 Mar. 2018].
6. Dacie, Lewis. Practical Haematology. Edition.12. 2012
VIVA
Controls Calibrators
Values ranges upto 2SD Values are fixed
Prepaired by various
companies, 3rd party
controls are also available
Calibrators are prepaired by analyser
manufacturer and is approved by CLSI
like authorities
After opening can be used
for 35 times within 25 days
After opening, can be used for 20times
within 1-2hr
Cant be used for calibrating
the machine Used to calibrate the machine
Used for Quality control
No expertise needed Needs handling by expertise
VIVA
What if the control values are continuously coming on one side (eg between 1sd and
2sd or >2SD) of mean???
Check for Sample error or Systemic Error
For sample error: 1) Check for the control sample ,whether it is viable/out of expiry date or not?
2) Can use another level of control for checking whether this problem is with
same level or with all levels?
If, values between 1sd and 2sd are acceptable, just has to calculate New Lab mean which will be
shifted a slight towards 1SD or the multiplying factor is derived to keep mean same.
For Systemic error: Values continuously >2SD are not acceptable, after checking all the sources of
error machine has to be recalibrated by service engineer.
VIVA
What is Within run and Between run?
Within run is
Values of the sample/ mean of the sample values is derived after running the sample
continuously (without a gap).
Eg If we running control sample at the start of day then run 5 times continuously without gap.
Between run is
Values of the sample/ mean of the sample values is derived after running the sample, for
specified times, not continuously, but with a gaps.
Eg If we have to run control sample for 5 times a day then, we run it 1 time after each hour gap or
between running other routine samples. (like stratified random sampling )

Contenu connexe

Tendances

Automation in blood banking
Automation in blood bankingAutomation in blood banking
Automation in blood bankingShreya D Prabhu
 
Demonstration of le cells
Demonstration of le cellsDemonstration of le cells
Demonstration of le cellsSHRUTHI VASAN
 
Automation in haematology bernard
Automation in haematology   bernardAutomation in haematology   bernard
Automation in haematology bernardBosco Mbonimpa
 
Troubleshooting QC Problems: Your QC has failed, what do you do next?
Troubleshooting QC Problems: Your QC has failed, what do you do next?Troubleshooting QC Problems: Your QC has failed, what do you do next?
Troubleshooting QC Problems: Your QC has failed, what do you do next?Randox
 
How often is Right for Laboratory Quality Control?
How often is Right for Laboratory Quality Control?How often is Right for Laboratory Quality Control?
How often is Right for Laboratory Quality Control?Randox
 
Quality assurance is necessary in blood banking
Quality assurance is necessary in blood bankingQuality assurance is necessary in blood banking
Quality assurance is necessary in blood bankingDr Shahida Baloch
 
Quality Control in Clinical Chemistry
Quality Control in Clinical ChemistryQuality Control in Clinical Chemistry
Quality Control in Clinical ChemistryDiganta Dey
 
Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Ola Elgaddar
 
Internal quality control
Internal quality controlInternal quality control
Internal quality controlSyed Basheer
 
Quality control in clinical laboratory
Quality control in clinical laboratoryQuality control in clinical laboratory
Quality control in clinical laboratorydrgomi basar
 
Haemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartHaemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartDr Rashmi Sood
 
Post analytical variables in Laboratory
Post analytical variables in LaboratoryPost analytical variables in Laboratory
Post analytical variables in Laboratorywww.jaailab.com
 
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...Adel Elazab Elged
 

Tendances (20)

Automation in blood banking
Automation in blood bankingAutomation in blood banking
Automation in blood banking
 
Hematology auto analyzer
Hematology auto analyzerHematology auto analyzer
Hematology auto analyzer
 
Demonstration of le cells
Demonstration of le cellsDemonstration of le cells
Demonstration of le cells
 
Automation in hematology part 1
Automation in hematology part 1Automation in hematology part 1
Automation in hematology part 1
 
Automation in haematology bernard
Automation in haematology   bernardAutomation in haematology   bernard
Automation in haematology bernard
 
Troubleshooting QC Problems: Your QC has failed, what do you do next?
Troubleshooting QC Problems: Your QC has failed, what do you do next?Troubleshooting QC Problems: Your QC has failed, what do you do next?
Troubleshooting QC Problems: Your QC has failed, what do you do next?
 
How often is Right for Laboratory Quality Control?
How often is Right for Laboratory Quality Control?How often is Right for Laboratory Quality Control?
How often is Right for Laboratory Quality Control?
 
WESTGARD RULES
WESTGARD RULESWESTGARD RULES
WESTGARD RULES
 
Quality assurance is necessary in blood banking
Quality assurance is necessary in blood bankingQuality assurance is necessary in blood banking
Quality assurance is necessary in blood banking
 
Quality Control in Clinical Chemistry
Quality Control in Clinical ChemistryQuality Control in Clinical Chemistry
Quality Control in Clinical Chemistry
 
Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017
 
Internal quality control
Internal quality controlInternal quality control
Internal quality control
 
Westgard rules
Westgard rulesWestgard rules
Westgard rules
 
Quality control in clinical laboratory
Quality control in clinical laboratoryQuality control in clinical laboratory
Quality control in clinical laboratory
 
Haemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartHaemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chart
 
quality control in pathology
quality control in pathologyquality control in pathology
quality control in pathology
 
Gel tech
Gel techGel tech
Gel tech
 
Post analytical variables in Laboratory
Post analytical variables in LaboratoryPost analytical variables in Laboratory
Post analytical variables in Laboratory
 
Automation in Hematology part 2
Automation in Hematology part 2Automation in Hematology part 2
Automation in Hematology part 2
 
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
 

Similaire à Calibration and Quality controls of automated hematology analyzer

Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .pptABRARAHMED767665
 
Clia final regulations
Clia final regulationsClia final regulations
Clia final regulationsMoustafa Rezk
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationTausif Momin
 
Pharmaceutical validation QA I Year.pptx
Pharmaceutical validation QA I Year.pptxPharmaceutical validation QA I Year.pptx
Pharmaceutical validation QA I Year.pptxmanoharkasde123
 
Pharmaceutical Instrument and Analytical Validation and Qualification (SHOPNI...
Pharmaceutical Instrument and Analytical Validation and Qualification (SHOPNI...Pharmaceutical Instrument and Analytical Validation and Qualification (SHOPNI...
Pharmaceutical Instrument and Analytical Validation and Qualification (SHOPNI...ShopnilAkash5
 
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...Md. Mizanur Rahman Miajee
 
Quality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryQuality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryDipesh Tamrakar
 
Total Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag YadavTotal Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag YadavDr Anurag Yadav
 
Ensuring the validity of results
Ensuring the validity of resultsEnsuring the validity of results
Ensuring the validity of resultsGH Yeoh
 
ICH Q2 Analytical Method Validation
ICH Q2  Analytical Method ValidationICH Q2  Analytical Method Validation
ICH Q2 Analytical Method ValidationNaila Kanwal
 
quality control in clinical pathology
quality control in clinical pathologyquality control in clinical pathology
quality control in clinical pathologyIshaque Vadakkethil
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stroundhydrologywebsite1
 
Quality Control In Clinical Laboratory
Quality Control In Clinical LaboratoryQuality Control In Clinical Laboratory
Quality Control In Clinical LaboratoryDr. Rajesh Bendre
 
Seminar by prakash on validation
Seminar by prakash on validationSeminar by prakash on validation
Seminar by prakash on validationprakash64742
 

Similaire à Calibration and Quality controls of automated hematology analyzer (20)

Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .ppt
 
Clia final regulations
Clia final regulationsClia final regulations
Clia final regulations
 
Quality control
Quality controlQuality control
Quality control
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Pharmaceutical validation QA I Year.pptx
Pharmaceutical validation QA I Year.pptxPharmaceutical validation QA I Year.pptx
Pharmaceutical validation QA I Year.pptx
 
Pharmaceutical Instrument and Analytical Validation and Qualification (SHOPNI...
Pharmaceutical Instrument and Analytical Validation and Qualification (SHOPNI...Pharmaceutical Instrument and Analytical Validation and Qualification (SHOPNI...
Pharmaceutical Instrument and Analytical Validation and Qualification (SHOPNI...
 
Quality control
Quality controlQuality control
Quality control
 
chem II 2019.ppt
chem II 2019.pptchem II 2019.ppt
chem II 2019.ppt
 
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
 
Quality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryQuality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistry
 
Total Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag YadavTotal Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag Yadav
 
Ensuring the validity of results
Ensuring the validity of resultsEnsuring the validity of results
Ensuring the validity of results
 
QA-QC-PSK.pdf
QA-QC-PSK.pdfQA-QC-PSK.pdf
QA-QC-PSK.pdf
 
Q2methodvalidation 161105173750
Q2methodvalidation 161105173750Q2methodvalidation 161105173750
Q2methodvalidation 161105173750
 
ICH Q2 Analytical Method Validation
ICH Q2  Analytical Method ValidationICH Q2  Analytical Method Validation
ICH Q2 Analytical Method Validation
 
quality control in clinical pathology
quality control in clinical pathologyquality control in clinical pathology
quality control in clinical pathology
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 
Quality Control In Clinical Laboratory
Quality Control In Clinical LaboratoryQuality Control In Clinical Laboratory
Quality Control In Clinical Laboratory
 
Quality management
Quality managementQuality management
Quality management
 
Seminar by prakash on validation
Seminar by prakash on validationSeminar by prakash on validation
Seminar by prakash on validation
 

Plus de Pranav S

ischemia and infarction.pdf
ischemia and infarction.pdfischemia and infarction.pdf
ischemia and infarction.pdfPranav S
 
Brain tumors in children with Updates- Pranav
Brain tumors in children with Updates- PranavBrain tumors in children with Updates- Pranav
Brain tumors in children with Updates- PranavPranav S
 
Polymerase chain reaction Pranav
Polymerase chain reaction PranavPolymerase chain reaction Pranav
Polymerase chain reaction PranavPranav S
 
Hematovigilance / Haemovigilance Pranav
Hematovigilance / Haemovigilance PranavHematovigilance / Haemovigilance Pranav
Hematovigilance / Haemovigilance PranavPranav S
 
Obesity Pathology
Obesity PathologyObesity Pathology
Obesity PathologyPranav S
 
Grossing of mandibulectomy specimen - Dr Pranav, MGIMS
Grossing of mandibulectomy specimen - Dr Pranav, MGIMSGrossing of mandibulectomy specimen - Dr Pranav, MGIMS
Grossing of mandibulectomy specimen - Dr Pranav, MGIMSPranav S
 
Systemic lupus erythematosus - Dr Pranav
Systemic lupus erythematosus - Dr PranavSystemic lupus erythematosus - Dr Pranav
Systemic lupus erythematosus - Dr PranavPranav S
 

Plus de Pranav S (7)

ischemia and infarction.pdf
ischemia and infarction.pdfischemia and infarction.pdf
ischemia and infarction.pdf
 
Brain tumors in children with Updates- Pranav
Brain tumors in children with Updates- PranavBrain tumors in children with Updates- Pranav
Brain tumors in children with Updates- Pranav
 
Polymerase chain reaction Pranav
Polymerase chain reaction PranavPolymerase chain reaction Pranav
Polymerase chain reaction Pranav
 
Hematovigilance / Haemovigilance Pranav
Hematovigilance / Haemovigilance PranavHematovigilance / Haemovigilance Pranav
Hematovigilance / Haemovigilance Pranav
 
Obesity Pathology
Obesity PathologyObesity Pathology
Obesity Pathology
 
Grossing of mandibulectomy specimen - Dr Pranav, MGIMS
Grossing of mandibulectomy specimen - Dr Pranav, MGIMSGrossing of mandibulectomy specimen - Dr Pranav, MGIMS
Grossing of mandibulectomy specimen - Dr Pranav, MGIMS
 
Systemic lupus erythematosus - Dr Pranav
Systemic lupus erythematosus - Dr PranavSystemic lupus erythematosus - Dr Pranav
Systemic lupus erythematosus - Dr Pranav
 

Dernier

Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...narwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptxDr.Nusrat Tariq
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingCall Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingNehru place Escorts
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000aliya bhat
 
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersBook Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersnarwatsonia7
 

Dernier (20)

Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptx
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingCall Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
 
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersBook Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
 

Calibration and Quality controls of automated hematology analyzer

  • 1. Calibration and Quality controls of Automated Hematology Analyzer Dr. Pranav N Shirbhate..
  • 2. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules  VIVA Ques at the End
  • 3. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 4. HISTORY Antonie van Leeuwenhoek Invented Microscope Wallace Coulter :First automated analyzer for counting and sizing cells and presented it in 1956 The world of cells was colorless until Paul Ehrlich stained blood cells
  • 5. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 6. AUTOMATIONS IN HEMATOLOGY Cell counts (Automated hematologyanalyzers) Diagnosis of hemoglobinopathies (HPLC) Immunophenotyping (Flow cytometry) Coagulation (Coagulometers)
  • 7. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 8.  Hemogram is a Backbone of any lab evaluation.  The Core Laboratory report well over 140,000 CBCs and 48,000 peripheral smears per year at our Dept of Pathology, MGIMS Sevagram.  It is a routine investigation for- Anaemia,Polycythemia, Infection, Inflammation, allergy, malignancy, bleeding tendency etc All automated cell counters are screening devices. Abnormalities must be verified by a blood film, staining and scanning by an expert observer. But testing each and every sample out of theses 1.4 lac samples a year by manual method is very laborious, cumbersome, so comes the need of automation. NEED OF AUTOMATION
  • 9. AUTOMATED TECHNIQUES OF BLOOD COUNTING • Semi-automated instruments - Require some steps, as dilution of blood samples . - Often measure only a small number of variables • Fully automated instruments - Require only that an appropriate blood sample is presented to the instrument. - They can measure 8-20 variables including some new parameters which do not have any equivalent in manual methods.
  • 10. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 11. INTRODUCTION • Quality Assuarance (QA) is the sum of all those activities in which the laboratory is engaged to ensure that information generated by laboratory is correct. • QA is not restricted to the development and retention of quality control charts but rather includes all aspects of laboratory activities that affects the results produced, from the choice of methods, to the education of personnel, to the handling of specimens and reporting results. • The real purpose of QA activities is to determine how correct or incorrect the results emanating from the lab are, and to allow those managing the lab to determine whether or not the lab is fulfilling its functions satisfactorily.
  • 12. INTRODUCTION 3 major activities of QA : 1 ) Preventive – 2 ) Assesment – 3 ) Corrective 1 ) Preventive – those activities that are done prior to the examination of the specimen or sample and that are intended to establish systems conducive to accuracy in analytical testing  Preventive maintenance and calibration of instruments,  Testing of media,  Orientation and training of personnel
  • 13. INTRODUCTION 3 major activities of QA : 2 ) Assessment – those activities that are done during testing to determine whether the test systems are performing correctly  The use of standard and controls,  Maintenance of control charts 3 ) Corrective – those activities that are done, when error is detected, to correct the system  equipment troubleshooting,  recalibration of instruments
  • 14. INTRODUCTION QA in Hematology Laboratory QA in Hematology lab is intended to ensure the reliability of the lab tests. The objective is to achieve precision and accuracy; So components of QA programme :  Standardization/ Calibration  Internal Quality Control ( IQC )  External Quality Control ( EQC )/ Proficiency testing
  • 15. DEFINITIONS IN TOTAL QUALITY MANAGEMENT
  • 16. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 17. QC & CALIBRATION TERMINOLOGIES Accuracy, I) A measure of agreement between the estimate of a value and a "true" value; quantifiable in terms of departure from accuracy, expressed as systematic error or bias; II) Accuracy of an analytic process, n - expressed as the difference between the average result obtainable by a method under specified conditions and the result accepted as true or standard. III) Accuracy of a result or control result, - Expressed as the difference between a result and the "true" value.
  • 18. QC & CALIBRATION TERMINOLOGIES Bias/ systematic error – • Systematic, signed deviation of the test result from the accepted reference value. • (Defined as the difference between the expectation of the test results and an accepted reference value.) Calibration, n – • The process of testing and adjustment of an instrument, kit, or test system, to provide a known relationship between the measurement response and the value of the substance being measured by the test procedure.
  • 19. QC & CALIBRATION TERMINOLOGIES Calibrator, n - A (reference) material (e.g., solution, suspension) or device of known quantitative/qualitative characteristics (e.g., concentration, activity, intensity, reactivity) used to calibrate, graduate, or adjust a measurement procedure or to compare the response obtained with the response of a test specimen/sample. NOTE: The quantities of the analytes of interest in the calibrator are known within limits ascertained during its preparation and are used to establish the relationship of an analytical method's response to the characteristic measured for all methods or restricted to some. The calibrator must be traceable to a national or international reference preparation or reference material when these are available. (like eg NCCLS/CLSI)
  • 20. ORGANIZATIONS INVOLVED IN STANDARDIZATION AND QUALITY MANAGEMENT SYSTEMS
  • 21. ORGANIZATIONS INVOLVED IN STANDARDIZATION AND QUALITY MANAGEMENT SYSTEMS
  • 22. WHAT THESE ORGANIZATIONS ARE? FOR e.g. CLSI CLSI/ NCCLS The Clinical and Laboratory Standards Institute (CLSI) is a volunteer-driven, membership-supported, not-for- profit, standards development organization. CLSI promotes the development and use of voluntary laboratory consensus standards and guidelines within the health care community. The organization’s mission is to develop clinical and laboratory practices and promote their use worldwide. The CLSI started as the National Committee on Clinical Laboratory Standards (NCCLS) and changed to its current name (CLSI) at the beginning of the year 2005.
  • 23. NCCLS/ CLSI STANDARDS AND GUIDELINES CLSI publishes voluntary consensus standards and guidelines, reports, and a number of companion products. Standards are documents developed through the consensus process that clearly identify specific, essential requirements for materials, methods, or practices for use in an unmodified form. Standards may, in addition, contain discretionary elements, which are clearly identified. Guidelines are documents developed through the consensus process describing criteria for a general operating practice, procedure, or material for voluntary use. Guidelines may be used as written or modified by the user to fit specific needs. Reports are documents that have not been subjected to consensus review and are released by the appropriate consensus committee. Companion products include any item provided or sold that is intended to be used in conjunction with a CLSI standard or guideline. Examples include, but are not limited to, Quick Guides, Wall Charts, software, and templates.
  • 24. QC & CALIBRATION TERMINOLOGIES (Quality) control material, n - A device, solution, or lyophilized preparation intended for use in the quality control process; NOTE: It should be similar to, and is analyzed along with, patient specimens. If different, it should have a recognized, defined response to analytic measurements. Control materials may or may not have known analyte concentrations (i.e., assigned values) within specified limits (e.g. target value ± standard deviation). Control materials are not used for calibration purposes.
  • 25. QC & CALIBRATION TERMINOLOGIES Derived red cell indices, n - Quantities (see MCH, MCHC, MCV) that may be calculated from the measurement of hemoglobin (Hb) concentration, packed (red) cell volume (PCV), and erythrocyte (RBC) concentration.
  • 26. QC & CALIBRATION TERMINOLOGIES Linearity, n - The ability (within a given range) to provide results that are directly proportional to the concentration {amount} of the analyte in the test sample.
  • 27. QC & CALIBRATION TERMINOLOGIES Precision, n - The closeness of agreement between independent test results obtained under prescribed {stipulated} conditions; NOTE: Precision is not typically represented as a numerical value but is expressed quantitatively in terms of imprecision—the SD or the CV of the results in a set of replicate Measurements.
  • 28. QC & CALIBRATION TERMINOLOGIES Precision, n - The closeness of agreement between independent test results obtained under prescribed {stipulated} conditions; NOTE: Precision is not typically represented as a numerical value but is expressed quantitatively in terms of imprecision—the SD or the CV of the results in a set of replicate Measurements. For practical purpose • Accuracy: Refers to closeness to the true value • Precision: Refers to reproducibility of test
  • 29. QC & CALIBRATION TERMINOLOGIES Drift, -
  • 30. QC & CALIBRATION TERMINOLOGIES Reference material, RM, n – A material or substance, one or more of whose property values are sufficiently homogeneous and well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to materials; Certified reference material, CRM, n - A reference material that has one or more values certified by a technically valid procedure and is accompanied by, or is traceable to, a certificate or other document that is issued by a certifying body; NOTES: a) defines a CRM as a “reference material, accompanied by a certificate, one or more of whose certified value is accompanied by an uncertainty at a stated level of confidence”; b) The term "Standard Reference Material (SRM)" is the name of a CRM, which is the trademark name of a certified reference material that has been certified and is distributed by the National Institute of Standards and Technology (NIST), a U.S. Government agency formerly known as the National Bureau of Standards (NBS).
  • 31. QC & CALIBRATION TERMINOLOGIES Sensitivity – It is the smallest quantity of analyte that can be reproducibly distinguished from background noise in a given assay system. Specificity - For quantitative tests, the ability of an analytical method to determine only the component it purports to measure or the extent to which the assay responds only to all subsets of a specified analyte and not to other substances present in the sample.
  • 32. QC & CALIBRATION TERMINOLOGIES Set point - The level (i.e., analyte concentration) at which an analyzer is calibrated. Standard (measurement), n - An authoritative “document” setting forth criteria for performance and characteristics. • Standard, primary, n - A standard that is designated or widely acknowledged as having the highest metro logical qualities and whose value is accepted without reference to other standards of the same quantity. • Standard, secondary, n - A standard whose value is assigned by comparison with a primary standard of the same quantity.
  • 33. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 34. GENERAL REQUIREMENTS FOR QC & CALIBRATION 1) Quality Control Manual • All quality control procedures should be described in detail in a quality manual. Clinical Laboratory Technical Procedure Manuals for more details. • The quality manual should also include a general statement of the laboratory quality policy and an acknowledgment that the policy is understood by all laboratory workers. • The persons with authority to design and modify all laboratory procedures and the persons with authority to approve such designs and changes should be listed. 2) Record Keeping • Comparisons with previous performance should be provided before analyzing control materials. • Electronic verification and background measurement recorded. • CBC/differential files should include control set-up, control run, and control review/report files.
  • 35. GENERAL REQUIREMENTS FOR QC & CALIBRATION 3) Retention of Records There are three major files of information. Storage of this data base is usually made HIS. • Laboratory order entry: The laboratory order entry data is maintained for a user-defined number of days and is then purged. • Workload data: Workload data is kept in active files for six months and is then archived. These archives are updated every month and stored in a safe place with timely access to permit review as the need arises. • Laboratory data files: The duration of QC record storage depends on local hospital or clinic policies or accreditation requirements, or should be a minimum of two years or more, whichever is greater. Proficiency testing results, noncommercial controls, or calibrators, if used, must have complete documentation of standardization and performance characteristics.
  • 36. GENERAL REQUIREMENTS FOR QC & CALIBRATION 4) Training and Qualifications • Laboratory workers who are assigned to carry out QC procedures must have been trained and records must indicate this. • As per CSLS guidelines, personnel who are given responsibility for interpreting QC results should not be the same personnel doing the routine work. 5) Environment Maintain environment that reflects routine working conditions, ambient temperature. A record of temperature be maintained.
  • 37. GENERAL REQUIREMENTS FOR QC & CALIBRATION 6) Analyzer Identification labels
  • 38. GENERAL REQUIREMENTS FOR QC & CALIBRATION 7) Preanalytic Variables Uncontrolled preanalytic variables can exert a misleading effect , therefore a checklist of possible error sources will help ensure that they will not be overlooked. Analyte Instability • Least likely to be seen in stabilized blood preparations and most likely to be seen in recently collected blood specimens. • Over-vigorous mixing may also play a role. • Viral disease may cause reduced lymphocyte count due to cell fragility. Cold agglutination may affect platelets as well as red cells. (Detection of analyte instability depends on the demonstration of a difference between the mean value of the first half of an iterative assay sequence and the mean value of the second half. To be significant the difference between means should exceed standard error of the first mean at 95% confidence.)
  • 39. GENERAL REQUIREMENTS FOR QC & CALIBRATION Analyte Nonuniformity Caused by inadequate mixing. Fresh blood is less susceptible to this. Mixing a specimen by rapid shaking or vortexing or any method that causes frothing has be avoided. Two components in nonuniformity. 1. nonuniform dispersion of the total cell mass throughout the plasma. 2. nonuniform distribution of the different cell types with respect to each other. To achieve uniformity of a whole blood suspension, - 10 complete inversions of the specimen container. (A completely filled container gives poor mixing; an air bubble must be present. If the specimen has been stored for 2 to 5 minutes, three additional inversions should precede the assay.)
  • 40. GENERAL REQUIREMENTS FOR QC & CALIBRATION Autoagglutination • Autoagglutination is a cause of unstable suspensions. • If not visible to the eye, autoagglutination or rouleaux should be suspected if 1) reduced RBC together with elevated MCHC are encountered; 2) The MCV may be slightly to markedly increased; 3) The rule of thumb “3 x Hb (g/dL) = Hct (%) ±3" is not met. • Its presence can usually be confirmed by microscopy, but be aware that a warm slide may cause autoagglutinates to dissociate. • Specimens with autoagglutination should not be used in quality control procedures. • Refrigerated specimens should be brought to room temperature before analysis.
  • 41. GENERAL REQUIREMENTS FOR QC & CALIBRATION Anticoagulant Effect • Insufficient blood collected into an evacuated container may cause shrinkage of all cell types, but red cells usually become restored to the size dictated by their osmotic equilibrium with the cell-counting diluent. • Leukocytes, erythrocytes, and platelets (e.g., EDTA-induced aggregation) may be irreversibly changed and cause errors in the differential count and platelet count and MCV.
  • 42. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 43. CALIBRATION • Calibration fine tunes your Hematology analyzer and provides the most accurate results possible. • In the normal process of tracking data for an extended period of time, your laboratory can make a specific decision to recalibrate a given parameter. Never adjust to a specific value for an individual sample. • For best performance, calibrate all the CBC parameters. The WBC differential is calibrated at the factory. They do not require calibration in the laboratory.
  • 44. CALIBRATION We should calibrate our instrument: *At installation *After the replacement of any component that involves dilution characteristics or the primary measurements (such as the apertures) *When advised to do so by your service representative
  • 45. BASIC OF CALIBRATION OF NEW/REPAIRED INSTRUMENT Establish Baseline Parameters • The analytical baseline data provide reference levels for judging the performance of the analyzer. • Baselines should be established or verified when a new analyzer is installed and when a modification or repair is made that could significantly affect the analyzer performance. • The product reference manual should provide instruction for frequency of recalibration under normal operating conditions. • Calibration or recalibration must be preceded by verification of the major analytical variables, which are general conditions of imprecision and accuracy and the analyzer. • Reportable range should be verified for a new installation and after major repair. • To avoid the possibility of analyzer defect causing errors in the baseline procedures, preventive maintenance should be performed shortly before calibration.
  • 46. ESTABLISH BASELINE PARAMETERS Imprecision Baseline • Baseline imprecision is described by the standard deviation of replicate assays of the same specimen. Measure Baseline Imprecision • Make n consecutive assays of the same well-mixed stabilized blood. The preferred number of iterations is n=31. • Fewer repeated assays (for example, n = 10) may be used, but if this is done, convert standard deviation (SD) to confidence limits (CL) as Evaluate Baseline Imprecision.
  • 47. ESTABLISH BASELINE PARAMETERS • Record the results of each analyte assay in a table. • Note the need to record the elapsed time of the measurement.
  • 49. ESTABLISH BASELINE PARAMETERS Evaluate Baseline Imprecision • Imprecision may be stated as SD or CL . (95% confidence limit). • If comparison is to be made 95% between standard deviations from different sources, e.g., between the laboratory estimate and a manufacturer's labeled value, be sure that each measurement has been made with the same number of iterations (n).
  • 50. ESTABLISH BASELINE PARAMETERS Evaluate Baseline Imprecision • If different values of n have been used, convert SD to CL95% as follows: • Multiply each SD by the value of t for (n-1) given in this Appendix. For e.g., • the laboratory used n = 31, but the manufacturer's labeled SD is based on n = 16. • Therefore, multiply the laboratory SD by 2.0423 and the manufacturer's SD by 2.1315. • This will convert both SD's to CL , making it possible to compare them.
  • 51. ESTABLISH BASELINE PARAMETERS Evaluate Baseline Imprecision • If different values of n have been used, convert SD to CL95% as follows: • Multiply each SD by the value of t for (n-1) given in this Appendix. For e.g., • the laboratory used n = 31, but the manufacturer's labeled SD is based on n = 16. • Therefore, multiply the laboratory SD by 2.0423 and the manufacturer's SD by 2.1315. • This will convert both SD's to CL , making it possible to compare them.
  • 52. ESTABLISH BASELINE PARAMETERS Accuracy Baseline • The accuracy baseline is defined as the starting point against which future bias control information will be compared. • Verification of accuracy must be made immediately (less than 15 minutes) after calibration or as directed by the manufacturer. If delayed, the data from the baseline verification process may be contaminated by drift or other analyzer error. • The term "calibration" means adjusting the analyzer to achieve zero bias, or verifying through a similar process that zero bias already exists. • The term “zero bias” means that any difference between the calibrator-assigned values and the results of calibrator assay is due only to analytical imprecision. • It is assumed that all automated hematology analyzers have a linear response over the reportable range, that the limits of linearity are specified, and that zero analyte concentration produces a zero analyzer response. These characteristics make it possible to calibrate analyzers by a single-point method.
  • 53. ESTABLISH BASELINE PARAMETERS Reference-Assayed Whole Blood Calibration Rarely used today in medical institutions. Not recommended. Standardized Stabilized Blood Calibration Method Frequency of calibration: - As per product reference manual. - Usually calibration is recommended after a major repair or after a significant change of base line accuracy (drift). - In the latter case, the cause of excessive drift should be corrected before proceeding with calibration. Iteration of calibrator assays: The number of assays will influence the precision of the calibration factor. Greater assay iteration will result in a lower SEM.
  • 54. ESTABLISH BASELINE PARAMETERS • More replication will yield a more precise calibration set point and more accurate subsequent patient assays but, because of the square root function, the rewards of increasing replication will reduced. • Furthermore, excessive mixing associated with increased replication may damage cells and introduce other errors. • It is suggested that 10 consecutive assays of the calibrator will give a satisfactory result. For example, the SD for WBC was found to be 0.2 x 109/L. The SEM of ten calibrator assays will then be 0.2 ÷ /10 or 0.2 ÷ 3.16 = ±0.063 x 109/L. This error, in combination with the analyzer SD will be propagated (included in) to patient assays. To obtain the 95% assay bias range, use the root sum square variance method as..
  • 56. ESTABLISH BASELINE PARAMETERS Reportable Range • Reportable range is the range of assay values over which accuracy is ensured; hence its reportable to the patient’s record. (accuracy of values beyond reportable range is not ensured). • It describes only the effect of analyte concentration and does not include possible error due to interferences, such as agglutination or hemolysis. • Upper and lower limits of reportable range for a newly installed analyzer can be verified by recovery of the assigned values of abnormal high and abnormal low stabilized blood controls. • Assays of sequential dilutions of blood between these limits do not contribute information that is helpful in establishing extension of this reportable range. • Reassurance of midrange linearity is given by the recovery of the assigned value of the normal stabilized blood control and by the verification of the calibration set point. For e.g., this happens a lot of time in leukemia cases where WBC values go beyond reportable range. Analyzer counts the cells but certainty about the accuracy can be questioned.
  • 57. CALIBRATION COMPLETION Technician from authorized / standardized calibration party Runs the blacks cycle to test values within range.
  • 58. CALIBRATION COMPLETION • He feeds all these (SD, CV, SEM and Base line parameters) values to analyzer software, and helps its synchronization with HIS/ Hospital software. • Calibration of the analyzer completes only when we receive calibration certificate clearly mentioning the CRM body like CSLS/NCCLS.
  • 59. CALIBRATION COMPLETION Now our machine/ analyzer is ready to be used for routine runs.
  • 60. DAILY MAINTENANCE • Daily cleaning • Electronic checks • Compare open and closed mode sampling (using a normal patient sample) • Run controls
  • 61. ENSURE THE INSTRUMENT IS FUNCTIONING PROPERLY • By Checking the reagent containers for: o Sufficient quantity o Not beyond expiration date o No precipitates, turbidity, particulate matter, or unusual colour o Proper connections between the instrument and the reagent containers • Check the waste container for: *Sufficient capacity and * Proper connections • Perform daily startup and Background counts • In addition to verifying daily startup results, verify acceptable: o Reproducibility o Carryover o Control Results
  • 62. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 63. QUALITY CONTROL Internal Quality Control – The set of procedures undertaken in a laboratory for the continuous assessment of work carried out in the laboratory and evaluation of tests to decide whether these are reliable enough for their results to be released to the requesting physician. NOTE: The main object is to ensure day-to-day consistency of measurements or observations, if possible in agreement with an agreed-on indicator of truth, such as a control material with assigned values.
  • 64. QUALITY CONTROL Internal Quality Control • Once a performance baseline has been established, its control is based on methods that will detect loss of accuracy or loss of precision, or both. • Control methods must be sensitive enough to reveal loss of performance that could compromise patient assay values, but they cannot be oversensitive to the point of signaling error when none exists. • As a minimum, a performance control system must provide the following information:  Analytical imprecision has not deteriorated with time or use.  Bias has remained within the limits established at the time of calibration.
  • 65. INTERNAL QUALITY CONTROL • The term “internal" indicates quality control procedures that take place within the laboratory immediately before, during, or immediately after an analysis run. • The procedures must be carried out as an integral part of patient assays. • Control assays performed only once daily may not provide enough information to ensure accuracy and consistency of results. • When the analyzer is restarted after complete shut-down, precision and set point should be verified to ensure that satisfactory analytical capability has not been lost. • Internal quality control is distinguished from external quality control. The latter is administered by a third party, such as an authorized proficiency testing body, based on blind assay of a stabilized control material, or by a manufacturer offering statistical peer comparison of routine stabilized blood control results. • Analyzers of differing designs and operating principles will have differing propensities for failure or for performance deterioration.
  • 67. INTERNAL QUALITY CONTROL Quality Control Using Stabilized Blood • Stabilized blood controls should be similar to calibrators, except that assigned values, if provided, may be conferred by calibrated analyzers using stabilized blood calibrators instead of reference-assayed whole blood. • Manufacturers' assigned values pose a temptation to use controls as calibrators and to "unblind" some types of internal and external performance reporting. • Controls should have 2-3 levels of analyte concentration. These levels should reflect performance at normal analyte levels and at, or close to, the low limits of the reportable range.
  • 68. INTERNAL QUALITY CONTROL Quality Control Using Stabilized Blood • Stabilized blood controls must be labeled for use with specified analyzers. • The laboratory should document that the product it uses is so labeled. • We are using Bio-Rad 3 level of control LOW/ NORMAL/ HIGH.
  • 69. INTERNAL QUALITY CONTROL Frequency of Use • We have to decide how many assays should separate controls. This decision should be made on the basis of work load and experience with the drift characteristics of the analyzer(s). • If the results of daily control assays are within laboratory- defined limits, such as ±2 SD of the assigned or conferred value, it can be assumed that the results of patient assays performed within a control-bracketed work segment are valid and reportable. • The fewer the control assays relative to patient specimens, the greater will be the number of assumed invalid results in the event of an out-of-control condition.
  • 70. INTERNAL QUALITY CONTROL Frequency of Use • As a minimum, assays of normal and abnormal levels of control material should be made at the beginning of each run and one assay of normal level control made at the end of the run to ensure that patient results have been satisfactory. We at our institution running these three control samples at the start of the new day of analyzer.
  • 71. INTERNAL QUALITY CONTROL Charts are prepared with the values shown by analyzer for that day.
  • 72. INTERNAL QUALITY CONTROL Test Replication / Precision on patients sample • The mean of replicate assays of control material is more informative than a single measurement. • Paired assays are recommended. Not only does the mean of the pair give a more exact value, but the difference between members of a pair also gives an estimate of imprecision.
  • 73. INTERNAL QUALITY CONTROL Test Replication / Precision on patients sample Overlap is a special case of test replication. When a new lot of control material is received, it should be tested in parallel with the currently used lot for 4 days. This will establish the ratio of the assigned values of the incoming lot to the assigned values of the currently used material. Differences in assigned values of the two lots should not be taken as an indication to recalibrate. We are doing precision bi-weekly ie on Tuesday and Friday at start of new day for the analyzer with random fresh EDTA sample. (Overlap - not done)
  • 74. INTERNAL QUALITY CONTROL Correlation Check Correlation check implies that any unexpected result of a test must be checked to see whether it can be explained on clinical ground or whether it correlates with other tests. E.g. 1) Unexpectedly higher Hb needs explanation of blood transfusion or by hemorrhage respectively. 2) Low MCHC confirmed by demonstrating hypochromic red cells on Romanowsky stained blood film 3) High MCV must correlate with macrocytosis. 4)Marked leukocytosis, thrombocytopenia and thrombocytosis, needs blood film examination We do correlation check for each and every sample we receive in our laboratory.
  • 75. INTERNAL QUALITY CONTROL Checking of aberrant result Recording blood count data on cumulative report for or charts is a good clinical practice and provides an inbuilt quality control system by making it easy to detect any aberrant result when compared with previously determined base line. E.g. Useful in detecting the occasional wild error caused by 1) incorrect labelling of specimen, 2) Inadequate suspension of blood before sampling, 3) Partial clotting of a blood sample, 4) Deterioration on storage. We have HIS (Hospital Information System), which is helpful in providing previous (cumulative) as well as other investigation done by patients which are very handy, while maintaining internal quality of reporting.
  • 76. INTERNAL QUALITY CONTROL Quality Control Methodology • The basis of the method is continued conformance of control assay values to the assigned values of the stabilized blood product. These values may be manufacturer-assigned or assigned by the laboratory, using an analyzer that has satisfied the requirements of base line accuracy. • Recovery of assigned values within the confidence limits of the analytical process, combined with the confidence limits of the assigned values, will provide assurance that analyzer accuracy and precision are acceptable.
  • 77. INTERNAL QUALITY CONTROL Quality Control Methodology • Any disagreement that exceeds these expected range, ie 2SD, limits calls for further investigation to determine whether it is a random event, loss of precision, loss of accuracy, or deterioration of the control material. • It is recommended that result recovery ranges be expressed as ±2 SD, or preferably 95% of CL of the mean value.
  • 78. INTERNAL QUALITY CONTROL Interpretation of Results An interpretive method for out-of-limits control result should be able to distinguish among the following:  Random event,  Loss of precision, and  Loss of accuracy.
  • 79. INTERNAL QUALITY CONTROL Random event versus imprecision: • With control boundaries defined as ±2 SD, one control value in exceed these limits as a random event. • Exclude random event by two further assays of the same type of control (normal, high, or low) that gave the out-of- limit result. • If the difference between these assay results is less than the established ±2 SD limits, the out-of-limit control result was probably a random event and requires no further action.
  • 80. INTERNAL QUALITY CONTROL Random event versus imprecision: • If the difference is greater than ±2 SD, assume precision loss and confirm this by two further assays. • At least two of the total of five results will exceed ±2 SD limits if precision has deteriorated. • In this case, patients’ results that preceded the out-of-limit condition should be reviewed in order to determine reportability.
  • 81. INTERNAL QUALITY CONTROL Loss of accuracy: • An out-of-limits control result due to loss of accuracy (bias change) can be verified by the repeated assay method given above. If inaccuracy has developed, the mean of the replicate results will differ by at least 1 SD from the mean of previous control results. • If the bias change affects only one of the three control levels, suspect a deterioration of the control material. • If the bias change affects the red cell indices, inspect the weighted moving average values for supporting evidence of change in RBC, Hb, or MCV. • Several consecutive similar results on one side of the mean = calibration fault causing consistent bias.
  • 82. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 83. EXTERNAL QUALITY CONTROL (EQA) / PROFICIENCY TESTING
  • 84. EXTERNAL QUALITY CONTROL (EQA) / PROFICIENCY TESTING • EQA complements internal quality control and is a basic requirement for clinical laboratory certification and accreditation. • Certification is a procedure by which a third party gives written assurance that a product, process or service conforms to specific requirements. • Accreditation is a procedure by which an authoritative body (NABL) gives formal recognition that a body or person is competent to carry out specific tasks.
  • 86. EXTERNAL QUALITY CONTROL (EQA) • EQA has following important principles and criteria.
  • 87. EXTERNAL QUALITY CONTROL (EQA) • EQA has following important principles and criteria.
  • 91. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 93. SHEWHART OR LEVEY-JENNINGS CHART ie Internal Quality Control • Dates of analysis plotted along the X-axis and control values plotted on Y- axis. • Mean and one, two and three standard deviation limits are also marked on Y-axis. • Distance from the mean is measured in standard deviations.
  • 94. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules
  • 95. WESTGARD RULES TO BE FOLLOWED 13s rule 13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.
  • 96. WESTGARD RULES TO BE FOLLOWED 12s rule 12s refers to the control rule that is commonly used with a LJ chart when the control limits are set as the mean plus/minus 2s. In the original Westgard multirule QC procedure, this rule is used as a warning rule to trigger careful inspection of the control data by the following rejection rules.
  • 97. WESTGARD RULES TO BE FOLLOWED 22s rule 22s - reject when 2 consecutive control measurements exceed the same mean plus 2s or the same mean minus 2s control limit. Warning
  • 98. WESTGARD RULES TO BE FOLLOWED R4s rule R4s - reject when 1 control measurement in a group exceeds the mean plus 2s and another exceeds the mean minus 2s. This rule should only be interpreted within-run, not between-run. The graphic should really imply that points 5 and 6 are within the same run.
  • 99. WESTGARD RULES TO BE FOLLOWED 41s rule 41s - reject when 4 consecutive control measurements exceed the same mean plus 1s or the same mean minus 1s control limit.
  • 100. WESTGARD RULES TO BE FOLLOWED 10x rule 10x - reject when 10 consecutive control measurements fall on one side of the mean. Note – There are plenty more Westgard’s rules. Can be accessed at https://www.westgard.com/mltirule.htm
  • 101. CONTENTS  History  Automations in Hematology  Need of automation  Introduction to Quality assurance  QC & calibration terminologies  Requirements for QC and calibration  Calibration of New/ Repaired automated Hematology Analyzer  Quality control in Hematology – Internal and External  Levey–Jennings (LJ) chart  Application of Westgard’s Rules  NABL
  • 102. NABL National Accreditation Board for Testing & Calibration Laboratories (NABL) is an autonomous society providing Accreditation (Recognition) of Technical competence of a testing, calibration, medical laboratory & Proficiency testing provider (PTP) & Reference Material Producer (RMP) for a specific scope following IEC/ISO17025:2005, ISO 15189:2012, IEC/ISO 17043:2010 & IEC/ISO 17034:2016 Standards It has Mutual Recognition Arrangements with Asia Pacific Laboratory Accreditation Cooperation (APLAC), Mutual Recognition Arrangement (MRA), International Laboratory Accreditation Cooperation (ILAC).
  • 103. NABL • NABL provides accreditation in all major fields of Science and Engineering such as Biological, Chemical, Electrical, Electronics, Mechanical, Fluid-Flow and many more. • NABL also provides accreditation for medical testing laboratories. • In addition, NABL also offers accreditation for Proficiency testing providers & Reference Material producers and is now signatory to APLAC MRA for both.
  • 104. NABL Benefits of accreditation • NABL is a Full member (ILAC MRA signatory) to ILAC as well as APLAC Mutual Recognition Arrangements (MRA) ,based on mutual evaluation and acceptance of other MRA Partner laboratory accreditation systems. • This facilitate acceptance of test / calibration results between countries.
  • 105. NABL Benefits of accreditation • This developing system of international mutual recognition agreements between accreditation bodies has enabled accredited laboratories to achieve a form of international recognition. • In medical field NABL accredation works as VISA for the any report. • Any report report reported in NABL accredited lab, makes it valid and gives confidence to overseas treating physician/surgeon about reports being standardized.
  • 107. WITHOUT UNDERSTANDING CALIBRATION AND QUALITY CONTROL
  • 108. TAKE HOME MESSAGE Automation in Hematology is like Vitamin tablets. It is a supplement and not the substitute to Manual methods.
  • 110. REFERENCES 1. Standard P. Calibration and Quality Control of Automated Hematology Analyzers; Proposed Standard NCCLS. Vol. 19. 2008. 2. Saiket M. World of Automation. In: Automated cell counter and quality control. 2014. p. 1–24. 3. Novak E, Guenther B. Automation in. In: Automation in Hematology. 2014. p. 369–94. 4. Anon, (2018). WestgardQC. [online] Available at: https://www.westgard.com/mltirule.htm [Accessed 2 Mar. 2018]. 5. En.wikipedia.org. (2018). National Accreditation Board for Testing and Calibration Laboratories. [online] Available at: https://en.wikipedia.org/wiki/National_Accreditation_Board_for_Testing_and_Calibration_Laboratories [Accessed 2 Mar. 2018]. 6. Dacie, Lewis. Practical Haematology. Edition.12. 2012
  • 111. VIVA Controls Calibrators Values ranges upto 2SD Values are fixed Prepaired by various companies, 3rd party controls are also available Calibrators are prepaired by analyser manufacturer and is approved by CLSI like authorities After opening can be used for 35 times within 25 days After opening, can be used for 20times within 1-2hr Cant be used for calibrating the machine Used to calibrate the machine Used for Quality control No expertise needed Needs handling by expertise
  • 112. VIVA What if the control values are continuously coming on one side (eg between 1sd and 2sd or >2SD) of mean??? Check for Sample error or Systemic Error For sample error: 1) Check for the control sample ,whether it is viable/out of expiry date or not? 2) Can use another level of control for checking whether this problem is with same level or with all levels? If, values between 1sd and 2sd are acceptable, just has to calculate New Lab mean which will be shifted a slight towards 1SD or the multiplying factor is derived to keep mean same. For Systemic error: Values continuously >2SD are not acceptable, after checking all the sources of error machine has to be recalibrated by service engineer.
  • 113. VIVA What is Within run and Between run? Within run is Values of the sample/ mean of the sample values is derived after running the sample continuously (without a gap). Eg If we running control sample at the start of day then run 5 times continuously without gap. Between run is Values of the sample/ mean of the sample values is derived after running the sample, for specified times, not continuously, but with a gaps. Eg If we have to run control sample for 5 times a day then, we run it 1 time after each hour gap or between running other routine samples. (like stratified random sampling )

Notes de l'éditeur

  1. Moving to the main topic of today’s discussion.
  2. Before directly considering the calibration and quality control we should clear with some of the terminologies.
  3. What we are interested today is about Standards.
  4. Again to continue with the terminologies-
  5. Span drift- Origin is stable. Zero drift- Origin is also shifted to one side of mean.
  6. This old couple who bought the expensive PC, but don’t even knw about the mouse her child talked about. Similarly without knowing quality control and calibration, our expensive autoanalysers are just a Waste.
  7. So to end with -
  8. So to end with -