CPAP- SCIENCE INDICATION AND
APPLICATION
PRESENTED BY
 DR NISHANT PRABHAKAR
 MD PEDIATRICS
DEFINITION
 Also called continuous distending pressure (CDP)
 Maintenance of an increased (positive)
transpulmonary pressure during the inspiratory &
expiratory phase of respiration in a spontaniously
breathing baby.
 By which alveoli are kept open which increase
the functional residual capacity(FRC) of the lungs
resulting in better gas exchange.
HISTORY
 Harrison 1968: described grunting in neonates as
naturally producing end expiratory pressure
 Gregory et al, 1971: introduced the clinical use of
distending pressure in neonates.(via endotrachial
tube or a head box)
 Kattwinkel reported successful use of nasal prongs in
neonates with RDS.
 Reports of significant lower incidence of chronic lung
disease from columbia university that used more
CPAP as compared to north american centre have led
to resurgence of interest in CPAP over the last 15
years.
PHYSIOLOGY OF CPAP
 Increases FRC
 Decreases V/Q mismatch
 Splints upper airway - airway resistance
 Increase tidal volume
 Decrease work of breathing
 Conserves surfectant
 Increase lung compliance
DISADVANTAGES OF
CONVENTIONAL VENTILATION
 High pressure- Barotrauma
 Tidal volumes- Volutrauma
 Atelectotrauma
 Inflammation & infection- Biotrauma
 CV ruptures the interalveolar septa thus
decrease the surface area of gas
exchange despite increasing lung volume.
Effect of Ventilator on Preterm
Lamb Lung
 At 0 hour
 24 hour after ventilation
 Limitations of premature lung
 1.Underdeveloped architect to
hold the lung open
 2.Thicker and few septa so less SA
for gas exchange
 Pinkerton KE, et al J Appl Physiol, 1994
Preterm Lambs at 72 Hours Distal
Airspace Wall Thickness -
 CV nCPAP
VENTILATOR INDUCED LUNG
INJURY (VILI)
 Biotrauma with tube
 atelectotrauma
VENTILATOR INDUCED LUNG
INJURY (VILI)
 Barotrauma
 Volutrauma
ADVANTAGE OF CPAP
1. Effect on lung growth- increased protein
and DNA
2- increased lung compliance
PVR Increases at Lung Volumes
Below and Above FRC
PVR
Lung Volume
FRC
HMD MAS
CPAP MAGIC
 Opens lung at FRC
 Keeps it open at minimum constant
pressure least atelecto & barotrauma
 Pulmonary arterial pressure are least
hence less V/Q mismatch less pressure
required.
 No ET tube- no biotrauma
Larger alveolus
r = 1.5
T = 3
P = (2 x 3) / 1.5
P = 4
Smaller alveolus
r = 1
T = 3
P = (2 x 3) / 1
P = 6
CPAP
Law of LaPlace : P = 2T/r
P : pressure T : surface tension r : radius
How does CPAP works
Stretches lung
pleura and upper
airway
 CPAP
Prevents collapse
of alveoli with
marginal stability
Stabilizes the
chest wall
Splints open
upper airway
Improves pH
Reduces airway
resistance
Recruitment of
alveoli
PaO2PaCO2
Improves V/Q mismatch and
reduces intrapulmonary shunt
Increased alveolar surface
area for gas exchange
Maintains lung at
FRC
Reduces work of
breathing
Reduces mixed and
central apnea
Reduces
obstructive apnea
Stimulates
stretch receptors
INDICATION OF CPAP
 COMMON
 Respiratory distress syndrome
 Apnea of prematurity (specially obstructive apnea)
 Post-extubation in preterm VLBW infants
 Transient tachypnea of newborn (TTNB)
 OTHER INDICATION
 Pneumonia
 Meconium aspiration/other aspiration syndrome
 Pulmonary edema/pulmonary hemorrhage
 Laryngomalacia/tracheomalacia/bronchomalacia
Early CPAP in RDS
 was proved to be more beneficial in the atelectatic
disease
 lower peak pressure required in infants treated with
CPAP
 enhance surfactant conservation
 reduce the need for IMV by 20%, except infants with
birth weight <1500 g.
 improve mortality and decrease the incidence of BPD
 prevent need for prolong intubation which reduce the
incidence of acquired subglottic stenosis
Failure of CPAP therapy in
RDS
 very low birth weight infant
 late application of CPAP
 severity of RDS
 associated disease e.g. sepsis,
hypotension
 infants with severe degree of
extrapulmonary shunt
 (Fox and coworkers, 1977)
CPAP in apnea of prematurity
 the application of low-level CPAP decrease the incidence
of apnea of prematurity (compared to other forms of
stimulation)
 improve oxygenation
 stimulation or inhibition of pulmonary reflexes
 alveolar stabilization
 mechanical splinting of airway; reduce supraglottic
resistance in both inspiration and expiration
 some investigators recommended the early use of CPAP
as a preventive measure of apnea of prematurity
CPAP IN INFANTS WITH MAS
 pathology of meconium aspiration
 atelectasis
 large airway obstruction
 V/Q abnormalities
 application of low-to moderate
level CPAP
 resolution of atelectasis
 stabilization of terminal airway
 incidence of pneumothorax: not
increased
 precautions in case with PPHN
C0NTRAINDICATION OF CPAP
 Progressive respiratory failure with PaCO2 levels >60
mmhg and/or inability to maintain oxygenation
(PaCO2 <50 mmHg)
 Certain congenital malformation of the airway
(choanal atresia, cleft palate, tracheoesophageal
fistula, congenital diaphragmatic hernia,etc)
 Severe cardiovascular instability (hypotension)
 Poor respiratory drive (frequent apnea and
bradycardia) that is not improved by CPAP.
ESSENTIALS OF CPAP
CPAP MACHINES
 An ideal CPAP delivery system consists of:
 A continuous supply of warm, humidified, blended
gases at a flow rate of 2-3 times the infant minute
ventilation.
 A device to connect CPAP circuit to infants airway.
 Means of creating a positive pressure in CPAP circuit.
CPAP DELIVERY SYSTEM
 Ventilator : ideal system to
provide CPAP but very costly
 CPAP system : should have
1. End expiratory pressure of 0-15
cm of water.
2. Humidification of upto 100%
3. Gas flow 5-8 L/min
4. Warming of gases to 34-37˚c
5. Blending oxygen-air mixture FiO2
0.21-1.0
6. Low noise compressor
7. Compatibility to run days & weeks
8. Reasonable cost
PATIENT INTERFACE FOR
PROVIDING CPAP
 Nasal prongs
1-Fisher & paykel 2-Hudson 3-Argyl nasal prong
CPAP DEVICES
SETTING PRESSURE, FLOW &
FiO2
1. Pressure- regulated by depth of immersion of
expiratory limb(water level being constant).start
with 5 cm water in case of RDS or pneumonia and 4
cm water for apnea management.(range- 4-8)
2. Flow- it should be minimal to produce bubbling in
the bubble chamber(2-5 L/min is sufficient)
3. FiO2- start with a FiO2 of 40 to 50% and after
adjusting the pressure,titrate FiO2 to maintain SpO2
between 89% to 94%.
MONITORING ADEQUACY AND
COMPLICATION OF CPAP
MONITORING THE INFANT
CONDITION
 Recommended monitoring:
• Respiratory status (RR, work of
breathing)
• Pre ductal oxygen saturation
• Cardiovascular status (HR, BP,
perfusion)
• GI status (abdominal distention, bowel
sounds)
• Neurological state (tone, activity,
responsiveness)
• Thermoregulation (temp)
WEANING FROM CPAP
 It is considered when clinical condition
for which CPAP was indicated is
passive.
 e.g. in case of RDS we have to see for
improvement in Silverman Anderson
score i.e. if score is less than 4 we can
try weaning
SILVERMAN ANDERSON SCORE
 Score >7 –respiratory failure
 Score 4-7 –respiratory distress
WEANING FROM CPAP
 CPAP for apnea may be removed after 24 -48 hrs of apnea
free interval.
 If the baby is stable on CPAP,first wean off the oxygen in
steps of 5% and then wean PEEP to minimum of 4cm in
step of 1cm/change.
 When baby is in FiO2<30%, PEEP 4cm, with normal
saturation and minimal retraction CPAP can be removed.
CPAP FAILURE
 CPAP failure is considered if FiO2 required is
>60% and PEEP required is > 7cm of water.
 If baby is continuing to have retraction,
grunting and apnea is considered fo
mechanical ventilation.
 If PaO2<50%, SpO2<85%, and PaCO2>60%
on CPAP with FiO2>60% and PEEP >7cm of
water is also considered for mechanical
ventilation.
BEFORE CONSIDERING CPAP
FAILURE ENSURE THE FOLLOWING
Maintaining Optimal
Airway Care: Humidification
• Maintain adequate
humidification of the
circuit to prevent
drying of secretions.
• Adjust settings to
maintain gas
humidification at or
close to 100%.
• Set the humidifier
temperature to 36.8-
37.3o C.
Complications associated with
bubble nasal CPAP
 ● Pneumothorax / PIE
- more in the acute phase
- not a contraindication for continuing CPAP
 ● Nasal obstruction
- Remove secretions and check for proper positioning
of the prongs
 ● Nasal septal erosion or necrosis
- Keep prongs away from the septum
 ● Gastric distension
 Intermittent or continuous aspiration of the stomach
 ● Feeding intolerance
Preventing Complications:
Gastric Distention
• NCPAP is not a
contraindication to
enteric feeding.
• Infants may
experience mild
abdominal
distention during
NCPAP delivery
from swallowing air.
CPAP BELLY
Preventing Complications:
Gastric Distention
 To prevent gastric
distention:
• Assess the infant’s
abdomen regularly
• Pass an oro-gastric tube
to aspirate excess air
before feeds q 2-4 hr
• An 8 Fr oro-gastric tube
may be left indwelling
to allow for continuous
air removal
CASE STUDY
 Antenatal steroid
 Arrangement before birth
Exercise
Cpap
Cpap
Cpap
Cpap
Cpap
Cpap

Contenu connexe

Tendances

Non-invasive ventilation - BiPAP
Non-invasive ventilation - BiPAPNon-invasive ventilation - BiPAP
Non-invasive ventilation - BiPAP
meducationdotnet
 

Tendances (20)

Oxygen therapy
Oxygen therapyOxygen therapy
Oxygen therapy
 
Non invasive ventilation
Non invasive ventilationNon invasive ventilation
Non invasive ventilation
 
Mechanical ventilation.ppt
Mechanical ventilation.pptMechanical ventilation.ppt
Mechanical ventilation.ppt
 
Noninvasive ventilation in neonates
Noninvasive ventilation in neonatesNoninvasive ventilation in neonates
Noninvasive ventilation in neonates
 
Oxygen delivery devices
Oxygen delivery devicesOxygen delivery devices
Oxygen delivery devices
 
Non-invasive ventilation - BiPAP
Non-invasive ventilation - BiPAPNon-invasive ventilation - BiPAP
Non-invasive ventilation - BiPAP
 
Basic modes of mechanical ventilation
Basic modes of mechanical ventilationBasic modes of mechanical ventilation
Basic modes of mechanical ventilation
 
Mechanical Ventilator by AJ
Mechanical Ventilator by AJMechanical Ventilator by AJ
Mechanical Ventilator by AJ
 
seminar on CPAP care
seminar on CPAP careseminar on CPAP care
seminar on CPAP care
 
Mechanical ventilation in neonates
Mechanical ventilation in neonatesMechanical ventilation in neonates
Mechanical ventilation in neonates
 
Pneumonectomy
PneumonectomyPneumonectomy
Pneumonectomy
 
Basic mechanical ventilation settings
Basic mechanical ventilation settingsBasic mechanical ventilation settings
Basic mechanical ventilation settings
 
Pediatric ARDS
Pediatric ARDSPediatric ARDS
Pediatric ARDS
 
BUBBLE CPAP: NEW BORN
BUBBLE CPAP: NEW BORNBUBBLE CPAP: NEW BORN
BUBBLE CPAP: NEW BORN
 
Basics of Neonatal Mechanical ventillation
Basics of Neonatal Mechanical ventillation Basics of Neonatal Mechanical ventillation
Basics of Neonatal Mechanical ventillation
 
Non Invasive Ventilation
Non Invasive VentilationNon Invasive Ventilation
Non Invasive Ventilation
 
NON INVASIVE VENTILATION IN NEONATES-PART 1
NON INVASIVE VENTILATION IN NEONATES-PART 1NON INVASIVE VENTILATION IN NEONATES-PART 1
NON INVASIVE VENTILATION IN NEONATES-PART 1
 
High flow nasal cannula (hfnc) by Tarfa Al-Sabhan
High flow nasal cannula (hfnc) by Tarfa Al-Sabhan High flow nasal cannula (hfnc) by Tarfa Al-Sabhan
High flow nasal cannula (hfnc) by Tarfa Al-Sabhan
 
Mechanical ventilation in neonates
Mechanical ventilation in neonatesMechanical ventilation in neonates
Mechanical ventilation in neonates
 
Weaning from mechanical ventilation
Weaning from mechanical ventilationWeaning from mechanical ventilation
Weaning from mechanical ventilation
 

En vedette

CPAP NASAL
CPAP NASALCPAP NASAL
CPAP NASAL
neoucin
 
VENTILACION MECANICA EN NEONATOLOGIA
VENTILACION MECANICA EN NEONATOLOGIAVENTILACION MECANICA EN NEONATOLOGIA
VENTILACION MECANICA EN NEONATOLOGIA
Monica Solis
 
Módulo 3. ventilación mecánica neonatal
Módulo 3. ventilación mecánica neonatalMódulo 3. ventilación mecánica neonatal
Módulo 3. ventilación mecánica neonatal
neonatologiaonline
 
Bubble cpap copy
Bubble cpap   copyBubble cpap   copy
Bubble cpap copy
Nilu Panch
 

En vedette (20)

Cpap nasal en el recien nacido
Cpap nasal en el recien nacidoCpap nasal en el recien nacido
Cpap nasal en el recien nacido
 
CPAP NASAL
CPAP NASALCPAP NASAL
CPAP NASAL
 
Cpap
CpapCpap
Cpap
 
Cpap slides
Cpap slidesCpap slides
Cpap slides
 
Cpap – nasal exp
Cpap – nasal expCpap – nasal exp
Cpap – nasal exp
 
CPAP and EMS
CPAP and EMSCPAP and EMS
CPAP and EMS
 
35. cpap
35. cpap35. cpap
35. cpap
 
VENTILACION MECANICA EN NEONATOLOGIA
VENTILACION MECANICA EN NEONATOLOGIAVENTILACION MECANICA EN NEONATOLOGIA
VENTILACION MECANICA EN NEONATOLOGIA
 
EMS CPAP Training
EMS CPAP TrainingEMS CPAP Training
EMS CPAP Training
 
Módulo 3. ventilación mecánica neonatal
Módulo 3. ventilación mecánica neonatalMódulo 3. ventilación mecánica neonatal
Módulo 3. ventilación mecánica neonatal
 
CPAP artesanal
CPAP artesanalCPAP artesanal
CPAP artesanal
 
Cpap nasal
Cpap nasalCpap nasal
Cpap nasal
 
Mechanical ventilation ppt
Mechanical ventilation pptMechanical ventilation ppt
Mechanical ventilation ppt
 
Bubble cpap copy
Bubble cpap   copyBubble cpap   copy
Bubble cpap copy
 
Bubble cpap
Bubble cpapBubble cpap
Bubble cpap
 
Cpap uso y causas
Cpap uso y causasCpap uso y causas
Cpap uso y causas
 
Cpap dispositivos aplicacion
Cpap  dispositivos aplicacionCpap  dispositivos aplicacion
Cpap dispositivos aplicacion
 
PROTOCOLS FOR NEONATES
PROTOCOLS FOR NEONATESPROTOCOLS FOR NEONATES
PROTOCOLS FOR NEONATES
 
Surfactante pulmonar
Surfactante pulmonarSurfactante pulmonar
Surfactante pulmonar
 
Oxigenoterapia
OxigenoterapiaOxigenoterapia
Oxigenoterapia
 

Similaire à Cpap

Ventilatory support in special situations balamugesh
Ventilatory support in special situations   balamugeshVentilatory support in special situations   balamugesh
Ventilatory support in special situations balamugesh
Dang Thanh Tuan
 
Mechanical Ventilation in ARDS vs COPD
Mechanical Ventilation in ARDS vs COPDMechanical Ventilation in ARDS vs COPD
Mechanical Ventilation in ARDS vs COPD
cairo1957
 

Similaire à Cpap (20)

Continuous Positive Airway Pressure.pptx
Continuous Positive Airway Pressure.pptxContinuous Positive Airway Pressure.pptx
Continuous Positive Airway Pressure.pptx
 
Ventillation 2
Ventillation 2Ventillation 2
Ventillation 2
 
Ventilatory support in special situations balamugesh
Ventilatory support in special situations   balamugeshVentilatory support in special situations   balamugesh
Ventilatory support in special situations balamugesh
 
NIV updated
NIV updatedNIV updated
NIV updated
 
Neonatal Ventilation
Neonatal VentilationNeonatal Ventilation
Neonatal Ventilation
 
Asthma-Non-invasive ventilation critical care seminar.pptx
Asthma-Non-invasive ventilation critical care seminar.pptxAsthma-Non-invasive ventilation critical care seminar.pptx
Asthma-Non-invasive ventilation critical care seminar.pptx
 
seminar on cpap dr poonam
seminar on cpap dr poonamseminar on cpap dr poonam
seminar on cpap dr poonam
 
Continuous Positive Airway Pressure in Neonates.pdf
Continuous  Positive Airway Pressure in Neonates.pdfContinuous  Positive Airway Pressure in Neonates.pdf
Continuous Positive Airway Pressure in Neonates.pdf
 
Non invasive ventilation
Non invasive ventilationNon invasive ventilation
Non invasive ventilation
 
Ards m ibrahim
Ards m ibrahimArds m ibrahim
Ards m ibrahim
 
Venti
VentiVenti
Venti
 
APRV
APRVAPRV
APRV
 
Nippv
NippvNippv
Nippv
 
Mechanical Ventilation in ARDS vs COPD
Mechanical Ventilation in ARDS vs COPDMechanical Ventilation in ARDS vs COPD
Mechanical Ventilation in ARDS vs COPD
 
Niv(non invasive ventilation) aiims ppt
Niv(non invasive ventilation) aiims pptNiv(non invasive ventilation) aiims ppt
Niv(non invasive ventilation) aiims ppt
 
CPAP seminar final
CPAP seminar finalCPAP seminar final
CPAP seminar final
 
acute respiratory distress syndrome
acute respiratory distress syndromeacute respiratory distress syndrome
acute respiratory distress syndrome
 
SCBU CPAP DR STEVE WONG
SCBU CPAP DR STEVE WONGSCBU CPAP DR STEVE WONG
SCBU CPAP DR STEVE WONG
 
Nippv
NippvNippv
Nippv
 
Weaning from ventilator
Weaning from ventilatorWeaning from ventilator
Weaning from ventilator
 

Dernier

Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
adilkhan87451
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
chetankumar9855
 

Dernier (20)

Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
 
Coimbatore Call Girls in Coimbatore 7427069034 genuine Escort Service Girl 10...
Coimbatore Call Girls in Coimbatore 7427069034 genuine Escort Service Girl 10...Coimbatore Call Girls in Coimbatore 7427069034 genuine Escort Service Girl 10...
Coimbatore Call Girls in Coimbatore 7427069034 genuine Escort Service Girl 10...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
 
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
 
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
 
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 

Cpap

  • 1. CPAP- SCIENCE INDICATION AND APPLICATION PRESENTED BY  DR NISHANT PRABHAKAR  MD PEDIATRICS
  • 2. DEFINITION  Also called continuous distending pressure (CDP)  Maintenance of an increased (positive) transpulmonary pressure during the inspiratory & expiratory phase of respiration in a spontaniously breathing baby.  By which alveoli are kept open which increase the functional residual capacity(FRC) of the lungs resulting in better gas exchange.
  • 3. HISTORY  Harrison 1968: described grunting in neonates as naturally producing end expiratory pressure  Gregory et al, 1971: introduced the clinical use of distending pressure in neonates.(via endotrachial tube or a head box)  Kattwinkel reported successful use of nasal prongs in neonates with RDS.  Reports of significant lower incidence of chronic lung disease from columbia university that used more CPAP as compared to north american centre have led to resurgence of interest in CPAP over the last 15 years.
  • 4. PHYSIOLOGY OF CPAP  Increases FRC  Decreases V/Q mismatch  Splints upper airway - airway resistance  Increase tidal volume  Decrease work of breathing  Conserves surfectant  Increase lung compliance
  • 5. DISADVANTAGES OF CONVENTIONAL VENTILATION  High pressure- Barotrauma  Tidal volumes- Volutrauma  Atelectotrauma  Inflammation & infection- Biotrauma  CV ruptures the interalveolar septa thus decrease the surface area of gas exchange despite increasing lung volume.
  • 6. Effect of Ventilator on Preterm Lamb Lung  At 0 hour  24 hour after ventilation  Limitations of premature lung  1.Underdeveloped architect to hold the lung open  2.Thicker and few septa so less SA for gas exchange  Pinkerton KE, et al J Appl Physiol, 1994
  • 7. Preterm Lambs at 72 Hours Distal Airspace Wall Thickness -  CV nCPAP
  • 8. VENTILATOR INDUCED LUNG INJURY (VILI)  Biotrauma with tube  atelectotrauma
  • 9. VENTILATOR INDUCED LUNG INJURY (VILI)  Barotrauma  Volutrauma
  • 10.
  • 11. ADVANTAGE OF CPAP 1. Effect on lung growth- increased protein and DNA
  • 12. 2- increased lung compliance
  • 13. PVR Increases at Lung Volumes Below and Above FRC PVR Lung Volume FRC HMD MAS
  • 14. CPAP MAGIC  Opens lung at FRC  Keeps it open at minimum constant pressure least atelecto & barotrauma  Pulmonary arterial pressure are least hence less V/Q mismatch less pressure required.  No ET tube- no biotrauma
  • 15. Larger alveolus r = 1.5 T = 3 P = (2 x 3) / 1.5 P = 4 Smaller alveolus r = 1 T = 3 P = (2 x 3) / 1 P = 6 CPAP Law of LaPlace : P = 2T/r P : pressure T : surface tension r : radius
  • 16. How does CPAP works
  • 17. Stretches lung pleura and upper airway  CPAP Prevents collapse of alveoli with marginal stability Stabilizes the chest wall Splints open upper airway Improves pH Reduces airway resistance Recruitment of alveoli PaO2PaCO2 Improves V/Q mismatch and reduces intrapulmonary shunt Increased alveolar surface area for gas exchange Maintains lung at FRC Reduces work of breathing Reduces mixed and central apnea Reduces obstructive apnea Stimulates stretch receptors
  • 18. INDICATION OF CPAP  COMMON  Respiratory distress syndrome  Apnea of prematurity (specially obstructive apnea)  Post-extubation in preterm VLBW infants  Transient tachypnea of newborn (TTNB)  OTHER INDICATION  Pneumonia  Meconium aspiration/other aspiration syndrome  Pulmonary edema/pulmonary hemorrhage  Laryngomalacia/tracheomalacia/bronchomalacia
  • 19. Early CPAP in RDS  was proved to be more beneficial in the atelectatic disease  lower peak pressure required in infants treated with CPAP  enhance surfactant conservation  reduce the need for IMV by 20%, except infants with birth weight <1500 g.  improve mortality and decrease the incidence of BPD  prevent need for prolong intubation which reduce the incidence of acquired subglottic stenosis
  • 20. Failure of CPAP therapy in RDS  very low birth weight infant  late application of CPAP  severity of RDS  associated disease e.g. sepsis, hypotension  infants with severe degree of extrapulmonary shunt  (Fox and coworkers, 1977)
  • 21. CPAP in apnea of prematurity  the application of low-level CPAP decrease the incidence of apnea of prematurity (compared to other forms of stimulation)  improve oxygenation  stimulation or inhibition of pulmonary reflexes  alveolar stabilization  mechanical splinting of airway; reduce supraglottic resistance in both inspiration and expiration  some investigators recommended the early use of CPAP as a preventive measure of apnea of prematurity
  • 22. CPAP IN INFANTS WITH MAS  pathology of meconium aspiration  atelectasis  large airway obstruction  V/Q abnormalities  application of low-to moderate level CPAP  resolution of atelectasis  stabilization of terminal airway  incidence of pneumothorax: not increased  precautions in case with PPHN
  • 23. C0NTRAINDICATION OF CPAP  Progressive respiratory failure with PaCO2 levels >60 mmhg and/or inability to maintain oxygenation (PaCO2 <50 mmHg)  Certain congenital malformation of the airway (choanal atresia, cleft palate, tracheoesophageal fistula, congenital diaphragmatic hernia,etc)  Severe cardiovascular instability (hypotension)  Poor respiratory drive (frequent apnea and bradycardia) that is not improved by CPAP.
  • 25. CPAP MACHINES  An ideal CPAP delivery system consists of:  A continuous supply of warm, humidified, blended gases at a flow rate of 2-3 times the infant minute ventilation.  A device to connect CPAP circuit to infants airway.  Means of creating a positive pressure in CPAP circuit.
  • 26.
  • 27. CPAP DELIVERY SYSTEM  Ventilator : ideal system to provide CPAP but very costly  CPAP system : should have 1. End expiratory pressure of 0-15 cm of water. 2. Humidification of upto 100% 3. Gas flow 5-8 L/min 4. Warming of gases to 34-37˚c 5. Blending oxygen-air mixture FiO2 0.21-1.0 6. Low noise compressor 7. Compatibility to run days & weeks 8. Reasonable cost
  • 28. PATIENT INTERFACE FOR PROVIDING CPAP  Nasal prongs 1-Fisher & paykel 2-Hudson 3-Argyl nasal prong
  • 30. SETTING PRESSURE, FLOW & FiO2 1. Pressure- regulated by depth of immersion of expiratory limb(water level being constant).start with 5 cm water in case of RDS or pneumonia and 4 cm water for apnea management.(range- 4-8) 2. Flow- it should be minimal to produce bubbling in the bubble chamber(2-5 L/min is sufficient) 3. FiO2- start with a FiO2 of 40 to 50% and after adjusting the pressure,titrate FiO2 to maintain SpO2 between 89% to 94%.
  • 31.
  • 33. MONITORING THE INFANT CONDITION  Recommended monitoring: • Respiratory status (RR, work of breathing) • Pre ductal oxygen saturation • Cardiovascular status (HR, BP, perfusion) • GI status (abdominal distention, bowel sounds) • Neurological state (tone, activity, responsiveness) • Thermoregulation (temp)
  • 34. WEANING FROM CPAP  It is considered when clinical condition for which CPAP was indicated is passive.  e.g. in case of RDS we have to see for improvement in Silverman Anderson score i.e. if score is less than 4 we can try weaning
  • 35. SILVERMAN ANDERSON SCORE  Score >7 –respiratory failure  Score 4-7 –respiratory distress
  • 36. WEANING FROM CPAP  CPAP for apnea may be removed after 24 -48 hrs of apnea free interval.  If the baby is stable on CPAP,first wean off the oxygen in steps of 5% and then wean PEEP to minimum of 4cm in step of 1cm/change.  When baby is in FiO2<30%, PEEP 4cm, with normal saturation and minimal retraction CPAP can be removed.
  • 37. CPAP FAILURE  CPAP failure is considered if FiO2 required is >60% and PEEP required is > 7cm of water.  If baby is continuing to have retraction, grunting and apnea is considered fo mechanical ventilation.  If PaO2<50%, SpO2<85%, and PaCO2>60% on CPAP with FiO2>60% and PEEP >7cm of water is also considered for mechanical ventilation.
  • 38. BEFORE CONSIDERING CPAP FAILURE ENSURE THE FOLLOWING
  • 39. Maintaining Optimal Airway Care: Humidification • Maintain adequate humidification of the circuit to prevent drying of secretions. • Adjust settings to maintain gas humidification at or close to 100%. • Set the humidifier temperature to 36.8- 37.3o C.
  • 40. Complications associated with bubble nasal CPAP  ● Pneumothorax / PIE - more in the acute phase - not a contraindication for continuing CPAP  ● Nasal obstruction - Remove secretions and check for proper positioning of the prongs  ● Nasal septal erosion or necrosis - Keep prongs away from the septum  ● Gastric distension  Intermittent or continuous aspiration of the stomach  ● Feeding intolerance
  • 41. Preventing Complications: Gastric Distention • NCPAP is not a contraindication to enteric feeding. • Infants may experience mild abdominal distention during NCPAP delivery from swallowing air.
  • 43. Preventing Complications: Gastric Distention  To prevent gastric distention: • Assess the infant’s abdomen regularly • Pass an oro-gastric tube to aspirate excess air before feeds q 2-4 hr • An 8 Fr oro-gastric tube may be left indwelling to allow for continuous air removal
  • 44. CASE STUDY  Antenatal steroid  Arrangement before birth
  • 45.
  • 46.
  • 47.
  • 48.

Notes de l'éditeur

  1. Let us understand the relationship between Lung volume – X axis with Pulmonary vascular resistance – y Axis . PVR is least once the lung is open at the FRC , at this blood flow is maximum with Best ventilation perfusion matching and gas exchange. You can see on Right- baby with MAS has higher lung volume while on left HMD has lower lung volume both RESULT in high PVR . In clinical practice , for a baby on CPAP with higher pressure the lung will over distend while low CPAP will cause reduced FRC – in both PVR will increase causing Rt to left shunt impairing pulmonary blood flow .Thus we must ensure lung is opened at FRC for best blood flow and better gas exchange .
  2. Imagine two alveoli are connected to same terminal airway – the smaller alveoli will have tendency to collapse and empty into larger due to higher collapsing pressure Notice Rt smaller alveoli has pressure 6 cm while left bigger one 4 cm as the radius small r is 1 and 1.5 respectively .